JJG 1101-2014

Chinese Standard: 'JJG 1101-2014'
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)Related StandardStatusGoogle Book
JJG 1101-2014English399 Add to Cart Days<=3 Verification Regulation of Medical Diagnostic X-ray Source for Dental Panorama JJG 1101-2014 Valid JJG 1101-2014
JJG 1101-2014Chinese17 Add to Cart <=1-day [PDF from Chinese Authority, or Standard Committee, or Publishing House]

   

BASIC DATA
Standard ID JJG 1101-2014 (JJG1101-2014)
Description (Translated English) Verification Regulation of Medical Diagnostic X-ray Source for Dental Panorama
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A58
Classification of International Standard 17.240
Word Count Estimation 17,193
Date of Issue 2014/8/25
Date of Implementation 2014/11/25
Quoted Standard JJG 744-2004; GB 9706.1-2007; GB/T 19042.4-2005
Drafting Organization China Testing Technology Research Institute
Administrative Organization National Technical Committee ionizing radiation metrology
Regulation (derived from) People's Republic of China The State Administration of Quality Supervision, Inspection and Quarantine Notice 2014 (No. 95); AQSIQ Notice 2014 No. 95
Issuing agency(ies) State Administration of Quality Supervision, Inspection and Quarantine
Summary This Standard applies to medical diagnosis panoramic dental X-ray radiation in the initial verification, verification and subsequent use of checks. This Standard does not apply to intra-oral X-ray imaging device calibration of radiation sources.

JJG 1101-2014
Verification Regulation of Medical Diagnostic X-ray Source for Dental Panorama
State verification procedures People's Republic of China
Medical Diagnostic panoramic dental X-ray radiation source
Issued on. 2014-08-25
2014-11-25 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Medical Diagnostic panoramic dental X-ray radiation source
Verification Regulation
Focal point. the National Technical Committee of ionizing radiation metrology
The main drafting units. China Testing Technology Research Institute
Fujian Institute of Metrology
Participated in the drafting unit. Shanghai Institute of Measurement and Testing
Xinjiang Institute of Metrology
Ankang City, Shaanxi measurement test
This procedure commissioned the National Technical Committee is responsible for ionizing radiation measurement interpretation
The main drafters of this procedure.
Yang Gan (China Testing Technology Research Institute)
Li Jie (Fujian Institute of Metrology)
Yang (China Testing Technology Research Institute)
Drafters participate.
Lu Xun (Shanghai Institute of Measurement and Testing)
Feng Xuefeng (Xinjiang Institute of Measurement and Testing)
Wang Bin (Shaanxi Ankang City metering test)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2. References (1)
3 terminology and units of measurement (1)
3.1 The term (1)
3.2 units of measurement (2)
4 Overview (2)
5 metering performance requirements (2)
5.1 air kerma rate (2)
Repeatability 5.2 radiation output (2)
5.3 Quality radiation output (2)
5.4 voltage (2)
5.5 Spatial resolution (2)
5.6 low contrast resolution (2)
5.7 Exposure Time (2)
6 General Requirements (2)
6.1 Appearance and flags (2)
6.2 electrical, mechanical and barrier properties (3)
7 control of measuring instruments (3)
7.1 test conditions (3)
7.2 Verification Project (3)
7.3 Verification Methods (4)
7.4 Verification Results Processing (5)
7.5 test cycle (5)
Appendix A test original records (recommended) formatting styles (6)
Appendix B test certificates/test results in the notice page (recommended) formatting styles (7)
Appendix C measuring layout and dental mold body (10)
introduction
This procedure based on JJF1002-2010 "national verification procedures writing rules" and JJF1001-2011 "Tong
Metering term and definition of "establishment.
Metrological characteristics and test methods panoramic dental X-ray radiation source with reference to the GB/T 19042.4-2005 /
IEC 61223-3-4.2000 "evaluation of medical imaging departments and routine testing - Part 3-4. dental X-ray set
Imaging equipment performance acceptance test "and JJG 744-2004" Medical Diagnostic X-ray radiation "set forth in the respective
Provisions.
This procedure was first formulated.
Medical Diagnostic panoramic dental X-ray radiation source verification procedures
1 Scope
The order applies to medical diagnostic panoramic dental X-ray radiation source initial verification, verification and subsequent use
an examination.
This procedure does not apply to intraoral X-ray imaging apparatus of the radiation source verification.
2. References
This procedure references the following documents.
JJG 744-2004 diagnostic X-ray radiation
GB 9706.1-2007/IEC 60601-1. 1988 Medical electrical equipment - Part 1. General requirements for safety
GB/T 19042.4-2005/IEC 61223-3-4.2000 Evaluation and routine medical imaging departments test
Part 3-4. Dental X-ray imaging equipment performance acceptance tests
For dated references, only the dated edition applies to this procedure; undated reference documents
Member, the latest edition (including any amendments) applies to this procedure.
3 terminology and units of measurement
3.1 Terms
3.1.1 Air Kerma airkerma
The sum of the initial kinetic energy of the ionized particles in the uncharged mass dm of air released from all of the charged particles
dEtr divided dm, symbol K.
K =
dEtr
dm
3.1.2 air kerma rate airkermarate
Dt increase the amount of time in the air kerma dK divided by dt, symbol K
dK
dt
3.1.3 Spatial resolution spatialresolution
In the images under high contrast conditions can distinguish between two adjacent objects ability.
3.1.4 Low contrast resolution lowcontrastresolution
And the ability to distinguish between homogeneous mass of low contrast objects.
3.1.5 half-value layer half-valuelayer
Will reduce the amount of radiation particles unidirectional flow to decrease the thickness of the initial value of half time, the symbol for the HVL.
3.1.6 tube voltage tubevoltage
Is applied between the X-ray tube anode and cathode potential difference.
Related standard: JJF 1101-2019    JJG 1100-2014