JJF 1101-2019English399 Add to Cart Days<=4 Calibration Specification for Environmental Testing Equipment for Temperature and Humidity Parameters Valid JJF 1101-2019
JJF 1101-2019Chinese22 Add to Cart <=1-day [PDF from Chinese Authority, or Standard Committee, or Publishing House]  

   

BASIC DATA
Standard ID JJF 1101-2019 (JJF1101-2019)
Description (Translated English) Calibration Specification for Environmental Testing Equipment for Temperature and Humidity Parameters
Sector / Industry Metrology & Measurement Industry Standard
Word Count Estimation 20,211
Date of Issue 2019-09-27
Date of Implementation 2020-03-27
Older Standard (superseded by this standard) JJF 1101-2003
Drafting Organization China Academy of Metrology, Hebei Institute of Metrology Supervision and Inspection, Fujian Academy of Metrology
Administrative Organization National Technical Committee for Temperature Measurement
Regulation (derived from) State Administration of Market Supervision and Administration Announcement No. 40 of 2019
Issuing agency(ies) State Administration of Market Supervision and Administration

JJF 1101-2019
Calibration Specification for Environmental Testing Equipment for Temperature and Humidity Parameters
National Measurement Technical Specification of the People's Republic of China
Specification for temperature and humidity parameter calibration of environmental test equipment
2019-09-27 release
2020-03-27 Implementation
Issued by the State Administration of Market Supervision and Administration
Environmental test equipment temperature,
Humidity parameter calibration specification
Replace JJF 1101-2003
Centralized unit. National Temperature Measurement Technical Committee
Main drafting unit. China Academy of Metrology
Hebei Metrology Supervision and Inspection Institute
Fujian Academy of Metrology
Participated in the drafting unit. Yunnan Institute of Metrology and Testing Technology
Zhejiang Academy of Metrology
Beijing Lindian Weiye Electronic Technology Co., Ltd.
This specification entrusts the National Temperature Measurement Technical Committee to interpret
The main drafters of this specification.
Jin Zhijun (Chinese Academy of Metrology)
Geng Rongqin (Hebei Metrology Supervision and Inspection Institute)
Lin Jun (Fujian Institute of Metrology)
Participating drafters.
Yang Ning (Yunnan Institute of Metrology and Testing Technology)
Cui Chao (Zhejiang Institute of Metrology)
Liu Hongyan (Hebei Metrology Supervision and Inspection Institute)
Huang Wei (Beijing Lindian Weiye Electronic Technology Co., Ltd.)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2 References (1)
3 Terminology (1)
4 Overview (2)
5 Metrological characteristics (2)
6 Calibration conditions (3)
6.1 Environmental conditions (3)
6.2 Load conditions (3)
6.3 Measurement standards and other equipment (3)
7 Calibration items and calibration methods (3)
7.1 Calibration items (3)
7.2 Calibration method (4)
7.3 Data processing (5)
8 Expression of calibration results (7)
9 Recalibration interval (7)
Appendix A Reference Format for Calibration Records of Environmental Test Equipment (9)
Appendix B. Reference Format for the Internal Page of the Calibration Certificate of Environmental Test Equipment (11)
Appendix C Examples of Uncertainty Evaluation of Temperature and Relative Humidity Deviations of Environmental Test Equipment (12)
introduction
This specification is based on JJF 1071-2010 "Rules for Compiling National Measurement and Calibration Specifications", JJF 1001-2011
Use Measurement Terms and Definitions "and JJF 1059.1-2012" Measurement Uncertainty Evaluation and Expression "as the basic series of regulations
Fan to write.
This specification adopts GB/T 5170.1-2016 "General Rules for Inspection Methods of Environmental Test Equipment for Electrical and Electronic Products"
Definition of relevant terms and technical content.
This specification is a revision of JJF 1101-2003 "Code for Calibration of Temperature and Humidity of Environmental Test Equipment", and
Compared with JJF 1101-2003, in addition to editorial changes, the main technical changes in this specification are as follows.
--- The name of the specification was changed to "Environmental Test Equipment Temperature and Humidity Parameter Calibration Specification", the specification guidance and reference
The goals are clearer and more specific;
--- Environmental testing equipment, working space, steady state, temperature fluctuation, humidity fluctuation,
Temperature uniformity, humidity uniformity, etc., modified the calculation method of temperature and humidity deviation and temperature and humidity fluctuation,
Consistent with the relevant content of GB/T 5170.1-2016;
--- Revised the relevant content of the technical requirements of environmental testing equipment, and changed the requirements of the equipment to temperature and humidity
Parameter requirements;
--- Standards generally use multi-channel temperature (temperature and humidity) display instruments or multi-channel temperature (temperature and humidity) measurement
Device;
--- Increasing the suggestion of measuring point layout information of environmental test equipment with volume less than 0.05m3;
--- It is recommended to change the recalibration time interval to 1 year;
--- Deleted the "Dry and Wet Bulb Method for Measuring Relative Humidity" appendix;
--- Revised the uncertainty analysis of temperature and humidity deviation calibration results.
The previous releases of this specification are.
--- JJF 1101-2003 "Code for Calibration of Temperature and Humidity of Environmental Test Equipment".
Specification for temperature and humidity parameter calibration of environmental test equipment
1 Scope
This specification is applicable to drying ovens with temperature range -80 ℃ ~ 300 ℃ and humidity range 10% RH ~ 100% RH
Incubator, climate aging box, mold test box, salt spray test box, corrosive gas test box, high and low temperature test box, delivery
Calibration of temperature and humidity parameters of environmental test equipment such as humid heat test chamber, constant temperature and humidity chamber.
The temperature and humidity parameters of environmental test equipment in other ranges can also be calibrated with reference to this specification.
2 Reference documents
This specification refers to the following documents.
JJF 1001-2011 General Measurement Terms and Definitions
GB/T 5170.1-2016 General rules for inspection methods of environmental test equipment for electrical and electronic products
GB/T 5170.2-2017 Environmental test equipment inspection method Part 2.Temperature test equipment
GB/T 5170.5-2016 Electrical and electronic products environmental test equipment inspection method damp heat test equipment
For dated reference documents, only the dated version applies to this specification; for those without date references
The latest version (including all amendments) is applicable to this specification.
3 Terminology
The terms and definitions defined in JJF 1001-2011, GB/T 5170.1-2016 and the following apply to this specification.
3.1 Environmental testing equipment environmental testing equipment
Equipment that simulates one or more environmental conditions and performs environmental tests on products.
3.2 working space
The part of the environmental test equipment that can maintain the specified temperature and relative humidity performance within the specified deviation range
space.
3.3 steady state
The variation of temperature and relative humidity at any point in the working space of the environmental test equipment reaches the performance index of the equipment itself
Time-seeking state.
3.4 Temperature deviation
When the environmental test equipment is in a stable state, the highest temperature and the lowest temperature of each measurement point in the working space are measured within the specified time
The upper and lower deviations between the temperature and the set temperature. Temperature deviation includes temperature deviation and temperature deviation.
[GB/T 5170.1-2016, definition 3.2.4]
3.5 Relative humidity deviation relativehumiditydeviation
When the environmental test equipment is in a stable state, each measuring point of the working space actually measures the highest relative humidity and the
The upper and lower deviations between the low relative humidity and the set relative humidity. Relative humidity deviation includes relative humidity deviation and relative humidity
Lower deviation.
[GB/T 5170.1-2016, definition 3.2.5]
Related standard: JJG 1101-2014    JJF 1248-2020