US$349.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1582-2016: Calibration Specification for(Specific) Activity Rapid Detecting Instrument Status: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1582-2016 | English | 349 |
Add to Cart
|
3 days [Need to translate]
|
Calibration Specification for(Specific) Activity Rapid Detecting Instrument
| Valid |
JJF 1582-2016
|
PDF similar to JJF 1582-2016
Basic data Standard ID | JJF 1582-2016 (JJF1582-2016) | Description (Translated English) | Calibration Specification for(Specific) Activity Rapid Detecting Instrument | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A58 | Word Count Estimation | 15,144 | Date of Issue | 2016-11-25 | Date of Implementation | 2017-02-25 | Regulation (derived from) | State Administration of Quality Supervision, Inspection and Quarantine Notice No.116 of 2016 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
JJF 1582-2016: Calibration Specification for(Specific) Activity Rapid Detecting Instrument---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for (Specific) Activity Rapid Detecting Instrument
National Metrological Technical Code of the People's Republic of China
Radioactivity (ratio) activity of rapid detection instrument
Calibration specification
CalibrationSpecificationfor (Specific) ActivityRapid
2016-11-25 release
2017-02-25 Implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
Radioactivity (ratio) activity of rapid detection instrument
Calibration specification
CalibrationSpecificationfor (Specific)
Instrument
Unit. National Diameter Radiation Metrology Technical Committee
Drafting unit. Shanghai Institute of Metrology and Testing Technology
East China University of Technology
This specification entrusts the National Diameter Radiation Metrology Technical Committee to explain
The main drafters of this specification.
Tang Fangdong (Shanghai Institute of Metrology and Testing Technology)
Wang Renbo (Donghua University of Technology)
He Linfeng (Shanghai Institute of Metrology and Testing Technology)
Participate in the drafters.
Qu Jinhui (Donghua University of Technology)
Zhao Chao (Shanghai Institute of Metrology and Testing Technology)
Xu Yihe (Shanghai Institute of Metrology and Testing Technology)
table of Contents
Introduction (Ⅲ)
1 Scope (1)
2 References (1)
3 terms and units of measurement (1)
3.1 Terms (1)
3.2 Unit of measurement (1)
4 Overview (2)
5 Metrological characteristics (2)
5.1 background (2)
5.2 Measuring range (2)
6 Calibration conditions (2)
6.1 Environmental conditions (2)
6.2 Measurement standard (2)
7 Calibration items and calibration methods (2)
7.1 Background (2)
7.2 Detection efficiency or response (3)
7.3 Repeatability (3)
7.4 Relative inherent error (4)
8 Calibration results (4)
9 re-school time interval (4)
Appendix A Calibration Record Recommended Format (5)
Appendix B Calibration Certificate Page Content (6)
Appendix C Radioactivity (Specific) Activity Fast Detector Detection Efficiency Calibration Results Uncertainty Evaluation Example (7)
Appendix D Radiation (Bi) activity of the rapid detection of the lower limit of the detection method (9)
Introduction
This specification is based on JJF 1071-2010 "National Metrological Calibration Rules".
Radioactivity (specific) activity rapid detector using a total amount of γ radioactivity measurement methods, mainly used in drinking
Water, beverages and cereals, meat, vegetables and cosmetics and other radioactive (specific) live fast detection. This specification
Preparation of reference JJG417-2006 "γ spectrometer", JJG377-1998 "radioactive activity meter", and Japan with the bit
Sui Association in February.2012 promulgated the "radioactive cesium screening of food" guidance law and other technical information.
This specification is the first release.
Standard for Calibration of Radioactivity (Ratio) Activity Fast Detector
1 Scope
This specification applies to gamma radioactivity measurement range (40 ~ 104) Bq, for all-spectrum or a single specific nuclide γ
Radioactivity measurement of radioactivity (specific) activity Fast detector calibration, specific nuclides mainly include 137Cs,
131I and so on.
2 reference file
This specification refers to the following documents.
Radioactivity meter
JJG417-2006 γ spectrometer
Standard Terminology and Definitions for JJF 1001-2011
Nuclear science and technology terminology - Part 1. Nuclear physics and nuclear chemistry GB/T 4960.1-2010
Nuclear science and technology terminology - Part 6. Nuclear instrumentation GB/T 4960.6-2008
For dated references, only the dated edition applies to this specification; references that are not dated
, The latest version (including all modifications) applies to this specification.
3 terms and units of measurement
3.1 terminology
JJF 1001-2011, GB/T 4960.1-2010, GB/T 4960.6-2008 Defined and the following terms and
The definition applies to this specification.
3.1.1 background background
Readings of the instrument without test samples.
3.1.2 count rate countrate
Count of instruments per unit time.
3.1.3 Marin Cup marlincup
And cylindrical scintillator detector, commonly used in amorphous (such as fluid or powder) samples for radioactive
Activity measurement of a container, its unique geometry is conducive to reducing the sample from the absorption, improve detection efficiency.
3.1.4 reference source references source
Has a defined radioactivity, used to measure the efficiency of the instrument detection efficiency. Usually with known activity of radionuclides
The solution is quantitatively diluted and sealed in a measuring vessel that matches the instrument.
3.1.5 detection limit detectionlimitofradionuclideactivity
The minimum detectable value of the instrument for gamma radionuclide activity.
3.2 Units of measurement
3.2.1 [radioactive] activity. beige [lux]; symbol. Bq.
3.2.2 [radioactive] specific activity. shell per liter or per shell per kilogram; symbol. Bq · L-1 or Bq · kg-1.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1582-2016_English be delivered?Answer: Upon your order, we will start to translate JJF 1582-2016_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1582-2016_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1582-2016_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|