Calibration Specification for Contact(Stylus)Instruments of Surface Roughness Measurement by Profile Method
| Valid |
JJF 1105-2018
|
JJF 1105-2003 | English | 799 |
Add to Cart
|
4 days [Need to translate]
|
Calibration specification for contact (stylus) instruments of surface roughness measurement by the profile method
| Obsolete |
JJF 1105-2003
|
Standard ID | JJF 1105-2018 (JJF1105-2018) | Description (Translated English) | Calibration Specification for Contact(Stylus)Instruments of Surface Roughness Measurement by Profile Method | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A52 | Word Count Estimation | 22,277 | Date of Issue | 2018-02-27 | Date of Implementation | 2018-08-27 | Older Standard (superseded by this standard) | JJF 1105-2003 | Drafting Organization | Shanghai Metrology and Testing Technology Research Institute, Shanghai Jiaotong University, China Institute of Metrology, Beijing Institute of Metrology and Measurement | Administrative Organization | National Geometric Quantity Engineering Measurement Technology Committee | Regulation (derived from) | AQSIQ Announcement No. 26 of 2018 | Issuing agency(ies) | State Administration of Quality Supervision, Inspection and Quarantine | Standard ID | JJF 1105-2003 (JJF1105-2003) | Description (Translated English) | Calibration specification for contact (stylus) instruments of surface roughness measurement by the profile method | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A52 | Word Count Estimation | 20,281 | Date of Issue | 2003-05-12 | Date of Implementation | 2003-11-12 | Older Standard (superseded by this standard) | JJG 301-1982 | Quoted Standard | GB/T 3505-2000; GB/T 10610-1998; GB/T 6062-2001; ISO 5436-1985; ISO 11562-1996; JJF 1059-1999; JJF 1001-1998 | Drafting Organization | Harbin Institute of Technology | Administrative Organization | National Engineering geometric parameters measured Technical Committee | Issuing agency(ies) | State Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to: stylus surface roughness measuring instrument the calibration. |
JJF 1105-2018
(Needle surface roughness measuring instrument calibration specification)
National Metrology Technical Specification of the People's Republic
Stylus surface roughness measuring instrument calibration specification
Published on.2018-02-27
2018-08-27 implementation
The General Administration of Quality Supervision, Inspection and Quarantine issued
Replace JJF 1105-2003
Focal Point. National Geometric Quantity Engineering Measurement Technology Committee
Main drafting unit. Shanghai Institute of Metrology and Testing Technology
Shanghai Jiaotong University
China Institute of Metrology
Beijing Institute of Metrology and Measurement Science
Participated in drafting unit. Taylor Hoopson Co., Ltd. China Branch
This specification entrusts the National Geometric Quantity Engineering Parametric Measurement Technical Committee to explain
The main drafters of this specification.
Wang Yongge (Shanghai Institute of Metrology and Testing Technology)
Yuan Yibao (Shanghai Jiaotong University)
Shi Yushu (China Institute of Metrology)
Wu Xun (Beijing Institute of Metrology and Measurement Science)
Participate in the drafters.
Cai Yuyu (Shanghai Institute of Metrology and Testing Technology)
Jiang Zhihua (Shanghai Institute of Metrology and Testing Technology)
Tang Haixiang (Taylor Hopson Co., Ltd. China Branch)
table of Contents
Introduction (II)
1 range (1)
2 References (1)
3 Overview (1)
4 Measurement characteristics (3)
4.1 Sensor stylus (3)
4.2 Straightness of sensor sliding track (3)
4.3 Residual contour (3)
4.4 indication error (3)
4.5 indication repeatability (3)
4.6 indication stability (3)
5 Calibration conditions (4)
5.1 Environmental conditions (4)
5.2 Calibration items and calibration standards and other equipment (4)
6 Calibration method (4)
6.1 Sensor stylus tip arc radius and angle (4)
6.2 Straightness of sensor sliding track (4)
6.3 Residual contour (4)
6.4 indication error (4)
6.5 indication repeatability (5)
6.6 indication stability (6)
7 Calibration result expression (6)
8 re-study interval (6)
Appendix A Evaluation of Uncertainty of Calibration Results of Indication Error of Stylus Roughness Tester (Example) (7)
Appendix B Calibration of Sensor Stylus Static Force and Measurement Characteristics of Sensor Head (11)
Appendix C Calibration of the stylus type surface roughness measuring instrument indicator and recorder metering characteristics (13)
Appendix D λc and λs band transmission characteristics (15)
Appendix E Information and Format of Calibration Certificate (17)
introduction
JJF 1071-2010 "National Metrology Calibration Specification Writing Rules", JJF 1001-2011 "General Measurement Terms and
Definitions, JJF 1059.1-2012 "Measurement and Expression of Measurement Uncertainty" together constitute the basis for the revision of this Code.
Basic series of measurement technology regulations.
This specification is for JJF 1105-2003 "Standard Specification for Stylus Surface Roughness Measuring Instrument" (hereinafter referred to as the original specification)
Fan) revision. In addition to the editorial changes, the main changes are as follows.
--- Added introduction.
--- Replace the reference file with the currently active version.
--- Increased structure of desktop stylus surface roughness measuring instrument and portable stylus surface roughness measuring instrument
Schematic, modified typical block diagram of the stylus instrument.
--- Modified the requirements for the metering characteristics of the sensor stylus. According to GB/T 6062-2009 "Product Geometry
Technical specification (GPS) surface structure profiling method (nominal characteristics of contact (stylus) type instrument", the original specification "sensing
The stylus static measurement force is modified to “sensor stylus static force measurement”, and its measurement characteristics are stated in the appendix.
Applicable standard and calibration methods; deleted the relevant calibration content of the sensor "static force change rate" item.
--- Modified the measurement characteristics requirements and calibration methods for the straightness of the sensor's sliding track.
--- Modified the "sensor head" related measurement characteristics, expressed in Appendix B.
--- Modified the measurement characteristics requirements and expression methods of the indication error.
---Modified the measurement characteristics requirements and expression methods of the instrument "exhibition repeatability" and "display value stability" in the text
And calibration methods. Revised the expression of the repeatability of the vertical magnification and horizontal magnification of the recorder in Appendix C,
The requirements for metering characteristics have also been modified.
--- Re-evaluated the stylus surface according to JJF 1059.1-2012 "Measurement Uncertainty Evaluation and Representation"
The uncertainty of the measurement result of the roughness error meter.
The previous versions of this specification were released.
---JJF 1105-2003;
---JJG301-1982.
Stylus surface roughness measuring instrument calibration specification
1 Scope
This specification applies to the calibration of the stylus type surface roughness measuring instrument.
2 reference files
This specification refers to the following documents.
GB/T 3505-2009 Product Geometrical Specification (GPS) Surface structure profiling terminology, definition
Surface structure parameters
GB/T 6062-2009 Product Geometrical Specification (GPS) Surface Structure Profile Method Contact
Nominal characteristics of the needle-type instrument
GB/T 10610-2009 Product Geometrical Specification (GPS) Surface Structure Profile Method for Surface Evaluation
Structure rules and methods
GB/T 18777-2009 Product Geometry Specification (GPS) Surface Structure Contour Method Phase Correction
Filtering characteristics of the filter
GB/T 19067.1-2003 Product geometry specification (GPS) Surface structure profile method
Standard Part 1. Physical measurement standards
GB/T 19600-2004 Product geometry specification (GPS) Surface structure profile method contact
Calibration of (stylus) instruments
For dated references, only dated versions apply to this specification; undated quotations
The latest version (including all amendments) applies to this specification.
3 Overview
The stylus type surface roughness measuring instrument (hereinafter referred to as the stylus type roughness meter) generally consists of a sensor, a driver, and a battery.
A sub-signal processing device, a computer, a printer, and the like. Partial analog stylus roughness meter with indicator and
Recorder. The working principle is as follows. the driver of the instrument drives the sensor to slide along the surface to be measured at a constant speed, and the sensor passes
The sharp stylus senses the geometric change of the surface being measured and converts it into an electrical signal. The signal is amplified and processed, then turned
The digital signal is stored in the memory of the computer system. The computer digitally filters this total contour and separates it out
Surface roughness profile and calculate its parameters. The measurement results can be output from the monitor or output from the printer. Analog
The stylus type roughness meter can read the measurement result from the indicator table, and the output surface profile of the recorder. Stylus roughness test
The block diagram is shown in Figure 1.
According to different principles of sensors, stylus type roughness meter can be divided into inductive, piezoelectric, photoelectric, laser
And grating type. Sensors can also be divided into lead-type sensors and non-head-type sensors. Guided sensors are only suitable
Used to measure surface roughness, but no lead sensor can be used to measure waves in addition to surface roughness.
......
JJF 1105-2003
Calibration specification for contact (stylus) instruments of surface roughness measurement by the profile method
National Metrology Technical Specification of the People's Republic
Stylus surface roughness measuring instrument calibration specification
Released on.2003-05-12
Implementation of.2003-11-12
The General Administration of Quality Supervision, Inspection and Quarantine issued
Stylus surface roughness measuring instrument
Calibration specification
Measurementby
Replace JJG301-1982
This specification was approved by the General Administration of Quality Supervision, Inspection and Quarantine on May 12,.2003, and
Implemented on November 12,.2003.
Focal Point. National Geometric Quantity Engineering Measurement Technology Committee
Main drafting unit. Harbin Institute of Technology
Participated in drafting unit. Times Group Corporation
Harbin Measuring and Cutting Tool Factory
This specification is interpreted by the focal point
The main drafters of this specification.
Yuan Yibao (Harbin Institute of Technology)
Participate in the drafters.
Wang Zhongbin (Time Group Corporation)
Lang Yanmei (Harbin Measuring and Cutting Tool Factory)
Tang Wenyan (Harbin Institute of Technology)
table of Contents
1 range (1)
2 Citations (1)
3 Overview (1)
4 Measurement characteristics (3)
4.1 Sensor stylus (3)
4.2 Sensor Head (3)
4.3 Straightness of driving sensor sliding motion (3)
4.4 Residual contour (3)
4.5 indication error (4)
4.6 indication repeatability (4)
4.7 indication stability (4)
5 Calibration conditions (4)
5.1 Environmental conditions (4)
5.2 Calibration items and standards and other equipment (4)
6 Calibration method (4)
6.1 Sensor stylus tip arc radius and angle (5)
6.2 Sensor stylus static measurement force and its rate of change (5)
6.3 Sensor Head Pressure (5)
6.4 Sensor head working surface roughness (5)
6.5 Sensor Head End Radius (5)
6.6 Straightness of driving sensor sliding motion (5)
6.7 Residual contour (5)
6.8 indication error (5)
6.9 indication repeatability (6)
6.10 indication stability (6)
7 Calibration result expression (6)
8 re-study interval (6)
Appendix A λc and λs band transmission characteristics (7)
Appendix B Calibration of Indicators and Loggers (9)
Appendix C Uncertainty Analysis of Calibration Results of Stylus Instrument Indication Error (11)
Appendix D Calibration Certificate Contents (15)
Stylus surface roughness measuring instrument calibration specification
1 Scope
This specification applies to the calibration of the stylus type surface roughness measuring instrument.
2 Citations
GB/T 3505-2000 Geometrical Product Specifications - Surface Structure. Profile Method - Terms
Meaning and parameters
GB/T 10610-1998 Geometrical Product Specifications - Surface Structures. Profiles
method
GB/T 6062-2001 Geometrical Product Specifications - Surface Structure. Profile Method - Contact (Tactile) Instrument
Nominal characteristic
ISO 5436.1985 CalibrationSpecimens-Stylusinstruments-Types,calibrationanduseof
Specimens calibration template - stylus instrument - type, calibration and use of the specimen
ISO 11562.1996 GeometricalProductSpecifications(GPS)-SurfaceTexture.Profile
method-MetrologicalcharacteristicsofphasecorrectfiltersProduct Geometry Specifications - Surface Structure. Wheel
Measured characteristics of the profile-phase correction filter
JJF 1059-1999 Measurement Uncertainty Evaluation and Representation
JJF 1001-1998 General measurement terms and definitions
Use of this specification should be done with the current valid version of the above cited documents.
3 Overview
The stylus type surface roughness measuring instrument (hereinafter referred to as the stylus type instrument) generally consists of a sensor, a driver, and an electronic letter.
No. Processing device, computer, printer, etc. Its working principle is. the driver of the instrument drives the sensor along the measured
The surface slides at a constant speed, and the sensor senses the geometric change of the surface to be measured through the sharp stylus and converts it into telecommunications.
number. The signal is amplified and processed and converted to digital signals for storage in a memory of the computer system. Computer pair
This original contour is digitally filtered to separate the surface roughness and calculate its parameters. Measurement results can be changed by the display
It can also be output by the printer.
The stylus instrument produced in the 1950s and early 1980s has an indicator and a recorder. This analog instrument
The measurement result can be read by the indicator table, and the recorder outputs the surface contour.
According to the different principles of sensors, stylus instruments can be divided into inductive, piezoelectric, photoelectric, laser and light.
Grid type, etc., can also be divided into lead type and no lead type. Guide head instruments are only used to measure surface roughness without guidance
In addition to measuring surface roughness, head instruments can also be used to measure surface waviness and surface geometry.
According to the structure, shape, weight and usage of the instrument, the stylus instrument is divided into two types. desktop and portable.
The block diagram is shown in Figure 1. Figure 2 is a schematic diagram of the composition of a stylus instrument.
......
|