US$279.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. GB/T 37182-2018: Gas analysis -- Gas chromatograph with plasma emission detector Status: Valid
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| GB/T 37182-2018 | English | 279 |
Add to Cart
|
3 days [Need to translate]
|
Gas analysis -- Gas chromatograph with plasma emission detector
| Valid |
GB/T 37182-2018
|
PDF similar to GB/T 37182-2018
Basic data | Standard ID | GB/T 37182-2018 (GB/T37182-2018) | | Description (Translated English) | Gas analysis -- Gas chromatograph with plasma emission detector | | Sector / Industry | National Standard (Recommended) | | Classification of Chinese Standard | G86 | | Classification of International Standard | 71.040.40 | | Word Count Estimation | 14,169 | | Date of Issue | 2018-12-28 | | Date of Implementation | 2019-11-01 | | Issuing agency(ies) | State Administration for Market Regulation, China National Standardization Administration |
GB/T 37182-2018: Gas analysis -- Gas chromatograph with plasma emission detector---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Gas analysis--Gas chromatograph with plasma emission detector
ICS 71.040.40
G86
National Standards of People's Republic of China
Gas analysis plasma emission gas chromatography
Published on.2018-12-28
2019-11-01 implementation
State market supervision and administration
China National Standardization Administration issued
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard was proposed by the China Petroleum and Chemical Industry Federation.
This standard is under the jurisdiction of the National Gas Standardization Technical Committee Gas Analysis Subcommittee (SAC/TC206/SC1).
This standard was drafted. Southwest Chemical Research and Design Institute Co., Ltd., Beijing Huayu Botai Technology Development Co., Ltd., China Test Technology
Institute of Chemistry, Institute of Metrology, Sichuan Zhongtuo Standards Technology Co., Ltd., China Institute of Atomic Energy, Sichuan
University, China Shipbuilding Industry Corporation, the 718th Research Institute, Hegang Co., Ltd., the branch gas plant, Shanghai Huaai chromatogram
Analysis Technology Co., Ltd.
The main drafters of this standard. Wang Shaonan, Sun Xiaoxuan, Zhao Shuide, Pan Yi, Wang Xing, Li Wenyu, Hu Shuguo, Deng Fanfeng, Xiao Yongbing, Zhang Penghui,
Xu Muhua, Yang Hongguang, Zhan Qin, Jiang Xiaoming, Wu Hao, Li Yao, Li Xiangyu, Yang Yikun, Wang Jianrang, Fang Hua.
Gas analysis plasma emission gas chromatography
1 Scope
This standard specifies a method for the determination of trace components in gases using a gas chromatograph equipped with a plasma emission detector.
This standard applies to high purity gas, industrial gas, standard gas and other gases such as hydrogen, oxygen, nitrogen, carbon monoxide, carbon dioxide, helium, argon, helium, neon and
Determination of hydrocarbon components such as alkane and benzene. The measurement range is 0.05×10-6~10×10-6 (volume fraction).
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this article.
Pieces. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 3723 General rules for sampling safety of industrial chemical products
GB/T 4842 argon
GB/T 4844 pure enamel, high purity 超 and ultra pure 氦
GB/T 8979 pure nitrogen, high purity nitrogen and ultra pure nitrogen
GB/T 30431 laboratory gas chromatograph
JJG700 gas chromatograph verification procedure
JJF1059.1 Measurement Uncertainty Evaluation and Representation
3 Terms and definitions
The following terms and definitions apply to this document.
3.1
Plasma emission detector plasmaemissiondetector
A device that converts an optical signal of a characteristic spectrum generated after a gas component to be tested forms a plasma into an electrical signal.
4 Principle
The gas component separated by the carrier gas through the column forms a plasma in the quartz cell by the action of an electromagnetic field, and the plasma of the gas component to be tested
The light emitted by the body is filtered and selected to convert the optical signal into an electrical signal by photoelectric conversion, and the content of the gas component to be tested is stronger than the electrical signal.
The degree has a proportional relationship, and the content of the gas component to be tested is calculated by comparing the standard sample signal with the signal of the component to be tested.
5 instruments
5.1 Overview
The basic composition and requirements of the instrument shall comply with the provisions of GB/T 30431.
The detection limit of the instrument should meet the requirements of the inspection task.
5.2 Common gas path of the instrument
5.2.1 Basic process
The basic procedure is mainly used for the determination of components in helium, see Figure A.1 in Appendix A.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB/T 37182-2018_English be delivered?Answer: Upon your order, we will start to translate GB/T 37182-2018_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB/T 37182-2018_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 37182-2018_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|