US$179.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. YY/T 0449-2018: Ultrasonic Doppler fetal monitor Status: Valid YY/T 0449: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
YY/T 0449-2018 | English | 179 |
Add to Cart
|
3 days [Need to translate]
|
Ultrasonic Doppler fetal monitor
| Valid |
YY/T 0449-2018
|
YY 0449-2009 | English | 359 |
Add to Cart
|
3 days [Need to translate]
|
Ultrasonic Doppler fetal monitor
| Obsolete |
YY 0449-2009
|
YY 0449-2003 | English | 719 |
Add to Cart
|
5 days [Need to translate]
|
Ultrasonic Doppler foetal monitor
| Obsolete |
YY 0449-2003
|
PDF similar to YY/T 0449-2018
Basic data Standard ID | YY/T 0449-2018 (YY/T0449-2018) | Description (Translated English) | Ultrasonic Doppler fetal monitor | Sector / Industry | Medical Device & Pharmaceutical Industry Standard (Recommended) | Classification of Chinese Standard | C41 | Classification of International Standard | 11.040.50 | Word Count Estimation | 9,937 | Date of Issue | 2018-11-07 | Date of Implementation | 2019-11-01 | Older Standard (superseded by this standard) | YY 0449-2009 | Quoted Standard | GB 9706.1; GB 9706.9; GB/T 14710; YY 0709; YY/T 1142 | Regulation (derived from) | Announcement No. 87 of 2018 by the State Drug Administration | Issuing agency(ies) | State Drug Administration | Summary | This standard specifies the terms and definitions, requirements, test methods and inspection rules for ultrasound Doppler fetal monitors. This standard applies to the ultrasound Doppler fetal monitor (hereinafter referred to as "instrument"), the product uses continuous wave or pulse wave ultrasound Doppler principle, continuous monitoring of the fetus during the perinatal period, and timely provided in case of abnormality Alarm information. |
YY/T 0449-2018: Ultrasonic Doppler fetal monitor---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Ultrasonic Doppler fetal monitor
ICS 11.040.50
C41
People's Republic of China Pharmaceutical Industry Standard
Replacing YY 0449-2009
Ultrasound Doppler fetal monitor
Published on.2018-11-07
2019-11-01 implementation
State Drug Administration issued
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces YY 0449-2009 "Ultrasonic Doppler fetal monitor", compared with YY 0449-2009, except for editorial changes
The main technical changes are as follows.
--- Modified the definition of ultrasound Doppler fetal monitor (see 3.1, 3.1 of the.2009 edition);
--- Removed the content of the product category (see Chapter 4 of the.2009 edition);
--- Increased the temperature drift indicator of the contraction pressure value (see 4.6);
--- Modified the requirements for the storage record function (see 4.7, see 5.6 of the.2009 version).
The documents of our country that have a consistent correspondence with the international documents referenced by the standard norms are as follows.
GB-9706.1-2007 Medical electrical equipment - Part 1. General requirements for safety (IEC 60601-1.1988, IDT);
--- GB 9706.9-2008 Medical electrical equipment - Part 2-37. Particular requirements for safety of ultrasonic diagnostic and monitoring equipment
(IEC 60601-2-37.2001, IDT);
--- YY 0709-2009 Medical electrical equipment - Part 1-8. General requirements for safety - Parallel standard. General requirements, medical
Testing and guidance for alarm systems in gas and medical electrical systems (IEC 60601-1-8.2003, IDT).
Some elements of this standard may involve patents. The issuing organization of this standard does not assume responsibility for identifying these patents.
This standard was proposed by the State Drug Administration.
This standard is under the jurisdiction of the National Medical Electrical Equipment Standardization Technical Committee Medical Ultrasound Equipment Subcommittee (SAC/TC10/SC2).
This standard was drafted. Hubei Medical Device Quality Supervision and Inspection Institute.
The main drafters of this standard. Wang Zhiwei, Jiang Shilin.
The previous versions of the standards replaced by this standard are.
---YY 0164-1994;
---YY 0449-2003;
---YY 0449-2009.
Ultrasound Doppler fetal monitor
1 Scope
This standard specifies the terms and definitions, requirements, test methods and inspection rules for ultrasound Doppler fetal monitors.
This standard applies to ultrasound Doppler fetal monitor (hereinafter referred to as "instrument"), which uses continuous wave or pulse wave ultrasound Doppler
Principle, continuous monitoring of the fetus during the perinatal period, and timely provide alarm information when an abnormality occurs.
2 Normative references
The following documents are indispensable for the application of this document. For dated references, only the dated version applies to this article.
Pieces. For undated references, the latest edition (including all amendments) applies to this document.
GB 9706.1 Medical electrical equipment - Part 1. General requirements for safety
GB 9706.9 Medical electrical equipment - Part 2-37. Particular requirements for safety of medical ultrasound diagnostic and monitoring equipment
GB/T 14710 Medical electrical requirements and test methods
YY 0709 Medical electrical equipment - Part 1-8. Safety general requirements. Parallel standard. General requirements, medical electrical equipment and medical
Testing and guidance for alarm systems in electrical systems
YY/T 1142 Medical Ultrasound Equipment and Probe Frequency Characteristics Test Method
3 Terms and definitions
The following terms and definitions apply to this document.
3.1
A device that continuously monitors the fetus during the perinatal period and provides alarm information in the event of an abnormality. The host, the ultrasound probe, and the palace
Compressed pressure sensor and other sensors connected to it, using ultrasonic Doppler principle to detect fetal heart rate, with monitoring and storage
Fetal heart rate, contraction pressure, and other parameters.
4 requirements
4.1 Ultrasonic operating frequency
The deviation of the ultrasonic operating frequency fW from the nominal frequency should be no more than ±10%.
4.2 fetal heart rate measurement and display range
The measurement and display range of fetal heart rate should be no narrower than 65 times/min~210 times/min.
4.3 fetal heart rate measurement error
The error in fetal heart rate measurement should be no more than ± 2 times/min.
4.4 Alarm function
The fetal heart rate alarm should meet the requirements specified in YY 0709, and the alarm state delay should not exceed 30s.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of YY/T 0449-2018_English be delivered?Answer: Upon your order, we will start to translate YY/T 0449-2018_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of YY/T 0449-2018_English with my colleagues?Answer: Yes. The purchased PDF of YY/T 0449-2018_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version YY/T 0449-2018?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version YY/T 0449-2018 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|