US$349.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 776-2014: Verification Regulation for Microwave Radiation and Leakage Energy Measuring Instruments Status: Valid JJG 776: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJG 776-2014 | English | 349 |
Add to Cart
|
3 days [Need to translate]
|
Verification Regulation for Microwave Radiation and Leakage Energy Measuring Instruments
| Valid |
JJG 776-2014
|
JJG 776-1992 | English | 239 |
Add to Cart
|
2 days [Need to translate]
|
Venification Regulation of Microwave Radiation and Leakage Energy Measuring Instrument
| Obsolete |
JJG 776-1992
|
PDF similar to JJG 776-2014
Basic data Standard ID | JJG 776-2014 (JJG776-2014) | Description (Translated English) | Verification Regulation for Microwave Radiation and Leakage Energy Measuring Instruments | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A58 | Classification of International Standard | 17.240 | Word Count Estimation | 15,157 | Date of Issue | 1/23/2014 | Date of Implementation | 7/23/2014 | Older Standard (superseded by this standard) | JJG 776-1992 | Adopted Standard | IEEE Std 1309-2005, NEQ | Regulation (derived from) | ?AQSIQ Announcement 2014 No. 10 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This Standard applies to the operating frequency range of 915MHz-18GHz microwave radiation measuring instrument for the first time with the leak test, follow-up testing and use of checks. |
JJG 776-2014: Verification Regulation for Microwave Radiation and Leakage Energy Measuring Instruments ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Verification Regulation for Microwave Radiation and Leakage Energy Measuring Instruments
State verification procedures People's Republic of China
Microwave radiation and leakage meter
Issued on. 2014-01-23
2014-07-23 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Microwave radiation and leakage meter
Verification Regulation
Instruments
Replacing JJG 776-92
Focal point. the National Radio Technical Commission metering
Drafting units. China Institute of Metrology
This procedure commissioned the National Radio Technical Commission responsible for the interpretation of measurement
The drafters of the regulations.
Lida Bo (China Institute of Metrology)
Tengjun Heng (China Institute of Metrology)
Bo Li (China Institute of Metrology)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2 Overview (1)
3 metering performance requirements (1)
4 General Requirements (1)
5 control of measuring instruments (1)
5.1 test conditions (1)
5.2 test items and test methods (3)
5.3 test results processing (4)
5.4 test cycle (4)
Appendix A test recording format (5)
Appendix B test certificates/test results in the notice page and format (6)
Introduction
The regulations refer to international standards IEEEStd1309-2005, IEEEStandardforCalibrationofElec-
tromagneticFieldSensorsandProbes, ExcludingAntennas, from9kHzto40GHz be
prepared by. The standard is based on the current major international calibration laboratories electromagnetic field probe and sensor calibration.
The procedures in place JJG 776-92 "Microwave radiation and leakage measurement instrument." The version with procedures and JJG 776-92
Than the power density of the device used in the device varies. In addition, the test frequency limit extended to 18GHz, for
Verification also in the basic range of minor changes to the (0 ~ 1) mW/cm2 as the basic range of the test, if the subject
Instrument range than 1mW/cm2, the increase (1 ~ 3) mW/cm2 power density of the measurement values.
The previous versions of the release JJG 776 case.
--- JJG 776-1992.
Microwave radiation and leakage meter
1 Scope
This procedure is applied for the first time in the operating frequency range of 915MHz ~ 18GHz microwave radiation leakage and measuring instrument
Test, test and subsequent use of checks.
2 Overview
Microwave radiation and leakage meter is used to measure radiation and microwave equipment microwave energy leaking to the space instruments.
Microwave probe of the electromagnetic field sensitive element, can be directly converted microwave signal into a DC signal, the indicator value.
3 metering performance requirements
Frequency range. 915MHz ~ 18GHz;
Power density measurement range. (0 ~ 3) mW/cm2;
Power density measurement maximum permissible error. ± 2.75dB.
4 General technical requirements
It should be the subject instrument manufacturer, instrument name, instrument model, serial number and power requirements information. Microwave radiation
And leakage measurement instrument control buttons, key switch, input and output ports should be clearly marked, and having a metering device
A manufacturing license logo.
There should be no prejudice to the subject instrument readings and affect the normal mechanical damage, control knobs and buttons should be able to work properly
For the display to show, all kinds of signs should be clear and complete; pointer instrument gauge needles should bend, mechanical meter
Zero can be adjusted freely.
5 control of measuring instruments
Controls include initial verification of measuring instruments, verification and subsequent use of checks.
5.1 verification Conditions
5.1.1 Environmental conditions
1) Ambient temperature. (23 ± 5) ℃.
2) Relative Humidity. 20% to 80%.
3) Power supply voltage and frequency. (220 ± 11) V, (50 ± 1) Hz.
4) The laboratory shall be no severe impact and vibration measurements of electromagnetic interference.
5.1.2 The test equipment
5.1.2.1 power density means the overall requirements of the standard
Frequency range. 915MHz ~ 18GHz;
The maximum power density value. ≥3mW/cm2;
Standard power density Uncertainty. ≤1dB.
|