US$489.00 · In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1587-2016: Calibration Specification for Multimeters Status: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1587-2016 | English | 489 |
Add to Cart
|
4 days [Need to translate]
|
Calibration Specification for Multimeters
| Valid |
JJF 1587-2016
|
PDF similar to JJF 1587-2016
Basic data Standard ID | JJF 1587-2016 (JJF1587-2016) | Description (Translated English) | Calibration Specification for Multimeters | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Word Count Estimation | 21,220 | Date of Issue | 2016-11-25 | Date of Implementation | 2017-05-25 | Older Standard (superseded by this standard) | JJG 315-1983; JJG 598-1989; JJG 724-1991 | Regulation (derived from) | State Administration of Quality Supervision, Inspection and Quarantine Notice No.116 of 2016 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
JJF 1587-2016: Calibration Specification for Multimeters---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Multimeters
National Metrological Technical Code of the People's Republic of China
Digital Multimeter Calibration Specification
2016-11-25 release
2017-05-25 implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
Digital Multimeter Calibration Specification
Instead of JJG315-1983
JJG598-1989
JJG724-1991
Unit. National Electromagnetic Metrology Technical Committee
The main drafting unit. China Institute of Metrology
Participated in the drafting unit. Henan Institute of Metrology
Shandong Institute of Metrology
Liaoning Institute of Metrology
Shandong Electric Power Research Institute
This specification entrusts the National Electromagnetic Metrology Technical Committee to explain
The main drafters of this specification.
Liu Yue (China Institute of Metrology)
Participate in the drafters.
Zhao honggang (China Institute of Metrology)
Chen Qingping (Henan Institute of Metrology)
Ma Xuefeng (Shandong Institute of Metrology)
Liang Guoding (Liaoning Institute of Metrology)
Fan Qiaocheng (Shandong Electric Power Research Institute)
Zhao Ya-ting (China Institute of Metrology)
table of Contents
Introduction (II)
1 Scope (1)
2 References (1)
3 terms and units of measurement (1)
3.1 indication error (1)
3.2 Input Zero Current (1)
3.3 Input Impedance (1)
3.4 Voltage frequency product (1)
4 Overview (1)
5 Metrological characteristics (2)
5.1 Indication error (2)
5.2 Maximum permissible error (2)
6 Calibration conditions (2)
6.1 Environmental conditions (2)
6.2 Measurement standards and other equipment (3)
7 Calibration items and calibration methods (3)
7.1 Calibration items (3)
7.2 Calibration method (4)
8 Calibration results (9)
9 re-school time interval (9)
Appendix A Examples of Measurement Uncertainty Evaluation (10)
Appendix B Calibrate Original Record Format (12)
Appendix C Calibration Certificate Inside Page Format (14)
Introduction
This specification is based on JJF 1071-2010 "National Metrological Calibration Rules", JJF 1059.1-2012
"Measurement Uncertainty Assessment and Representation".
This specification replaces JJG315-1983 "DC digital voltmeter", JJG598-1989 "DC digital ammeter"
And JJG724-1991 "DC digital ohmmeter".
Compared with JJG315-1983, JJG598-1989 and JJG724-1991, the main changes are as follows.
- merged JJG315-1983, JJG598-1989 and JJG724-1991, and rewritten the number of more
The calibration contents of the DC voltage, DC current and DC resistance measurement functions of the table;
--- Added digital multimeter AC voltage and AC current measurement function of the calibration content.
Digital Multimeter Calibration Specification
1 Scope
This specification applies to DC voltage, DC current, DC resistance, AC voltage and AC current measurement function
The digital multimeter (hereinafter referred to as the digital table) is also suitable for applications with the above-mentioned single measurement function or combined measurement
Function of the instrument calibration.
The specification of AC and DC voltage measurement limit of 1000V, AC and DC current measurement limit of 100A, resistance measurement
The upper limit of 10GΩ, AC voltage frequency range of 10Hz ~ 1MHz, AC current frequency range of 10Hz ~
10kHz.
2 reference file
This specification refers to the following documents.
Digital Multimeter. GB/T 13978-2008
For dated references, only the dated edition applies to this specification; references that are not dated
, The latest version (including all modifications) applies to this specification.
3 terms and units of measurement
3.1 error error
The difference between the indication value of the digital table and the reference value (standard value) of the corresponding input amount.
3.2 Input zero current inputoffsetcurrent
When the input signal is zero, the input circuit in the digital table is caused by the internal current caused by the digital meter.
3.3 input impedance inputimpedance
The impedance of the input circuit measured between the input terminals of the instrument under operating conditions.
3.4 Voltage frequency product V · Hzvolt-herzproduct
The product of the square root value (V) of the AC voltage and its frequency (Hz).
4 Overview
The digital meter is used to measure the voltage (and/or current, resistance) and display the measured value in decimal
Multi-limit, multi-function measuring instruments.
DC voltage function is the basis of DC current and DC resistance function. The DC current is measured by DC current -
Voltage converter, the amount of current converted into the amount of voltage measurement to achieve. The DC resistance is measured by electricity
Resistance - voltage converter, the amount of resistance will be converted into voltage measurement method to achieve. AC voltage measurement is passed
AC-DC converter, the AC voltage will be converted to DC voltage to achieve. AC current measurement is done by crossing
Current-to-voltage converter, the amount of current converted into the amount of voltage measurement to achieve.
Digital table by the display bit is divided into three and a half, four and a half, five and a half, six and a half, seven and a half, eight and a half
The first show less than nine said half position).
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1587-2016_English be delivered?Answer: Upon your order, we will start to translate JJF 1587-2016_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1587-2016_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1587-2016_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|