US$559.00 · In stock Delivery: <= 5 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1558-2016: Calibration Specification of Variable Frequency Electrical Quantities Transducer for Measuring Status: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1558-2016 | English | 559 |
Add to Cart
|
5 days [Need to translate]
|
Calibration Specification of Variable Frequency Electrical Quantities Transducer for Measuring
| Valid |
JJF 1558-2016
|
PDF similar to JJF 1558-2016
Basic data Standard ID | JJF 1558-2016 (JJF1558-2016) | Description (Translated English) | Calibration Specification of Variable Frequency Electrical Quantities Transducer for Measuring | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Word Count Estimation | 24,285 | Date of Issue | 2016-03-03 | Date of Implementation | 2016-06-03 | Quoted Standard | JJG 126-1995; GB/T 13850 | Regulation (derived from) | Announcement No.26 of the General Administration of Quality Supervision, Inspection and Quarantine | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
JJF 1558-2016: Calibration Specification of Variable Frequency Electrical Quantities Transducer for Measuring ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification of Variable Frequency Electrical Quantities Transducer for Measuring
National Metrological Technical Code of the People's Republic of China
Standard Specification for Variable Frequency Electricity Transmitter for Measurement
2016-03-03 release
2016-06-03 implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
Measurement of variable frequency power transmitter
Calibration specification
Unit. National Electromagnetic Metrology Technical Committee
The main drafting unit. Hunan Institute of Measurement and Testing
Zhejiang Institute of Metrology
Participated in the drafting unit. Hunan Galaxy Electric Co., Ltd.
This specification entrusts the National Electromagnetic Metrology Technical Committee to explain
The main drafters of this specification.
Li Yixuan (Hunan Institute of Measurement and Testing)
Wang Yougui (Hunan Institute of Measurement and Testing)
Lin Ting (Zhejiang Institute of Metrology)
Participate in the drafters.
Wu Shuanghe (Hunan Institute of Measurement and Testing)
Wang Ning (Hunan Galaxy Electric Co., Ltd.)
Liao Zhongchi (Hunan Galaxy Electric Co., Ltd.)
table of Contents
Introduction (Ⅲ)
1 Scope (1)
2 References (1)
3 Overview (1)
4 Metering characteristics (1)
4.1 Voltage (1)
4.2 Current (1)
5 Calibration conditions (2)
5.1 Environmental conditions (2)
5.2 Measurement standards and other equipment (2)
6 Calibration items and calibration methods (2)
6.1 Calibration items (2)
6.2 Calibration method (3)
7 Calibration results (13)
8 re-school time interval (13)
Appendix A Measurement Variable Frequency Electricity Transmitter Voltage Measurement Results Uncertainty Evaluation Example (14)
Appendix B Calibrate Original Record Format (16)
Appendix C Calibration Certificate Inside Page Format (Page 2) (17)
Appendix D Calibration Certificate Calibration Result Page Format (Page 3) (18)
Introduction
This specification is based on JJF 1071-2010 "national measurement calibration standard preparation rules", JJF 1001-2011 "pass
Metrological terms and definitions "and JJF 1059.1-2012" Measurement Uncertainty Assessment and Representation ".
In this specification, the measurement of variable frequency power transmitter in the practical application of the measurement object is usually a complex waveform of the electricity
Voltage, current signal, the output can be DC, AC and digital signals. Such transmitters have a wide band of measurements
Capability, mainly used for variable frequency drive in the power measurement.
This specification is the first release.
Standard Specification for Variable Frequency Electricity Transmitter for Measurement
1 Scope
This specification applies to the input for the AC voltage or current signal, the output is AC, DC analog signal or digital signal
Of the measured frequency of the voltage transmitter in the (5 ~ 1500) Hz at a specific frequency of voltage, current, delay phase shift of the school
quasi. This specification does not apply to calibrations of measuring transmitters operating only at high frequency or fixed frequency operation.
2 reference file
This specification refers to the following documents.
JJG126-1995 AC power conversion for the DC power electrical measurement transmitter
GB/T 13850 AC power conversion to analog or digital signal of the electrical measurement transmitter
For dated references, only the dated edition applies to this specification; references that are not dated
, The latest version (including all modifications) applies to this specification.
3 Overview
Measurement of variable frequency power transmitter is a certain frequency range and according to a specific law changes in the voltage or current letter
Is converted into an instrument that is measured as a function of the relationship signal, and its output is all the features or parts that contain the input power
Sub-characteristic information exchange, DC analog signal or digital signal.
Measuring variable frequency power transmitter is usually composed of sensors, converters and transmission units, etc., the input power through
The sensor is converted into a small power signal, converted by the converter through the transmission unit for secondary equipment, the principle shown in Figure 1
As shown. Measurement of variable frequency power transmitter (hereinafter referred to as the transmitter) in accordance with the input power can be divided into voltage transmitter, power
Flow transmitter; according to the output signal type can be divided into analog output transmitter and digital transmitter; analog output transmitter
The signal type can be divided into AC voltage output, AC current output, DC voltage output, DC current output.
figure 1
4 measurement characteristics
4.1 Voltage
Measuring range. 100mV ~ 10kV
Maximum permissible error. ± (0.05% ~ 5%)
Delay phase shift maximum allowable error. ± (0.01 ° ~ 5 °)
4.2 Current
Measuring range. 5mA ~ 1000A
Maximum permissible error. ± (0.05% ~ 5%)
Delay phase shift maximum allowable error. ± (0.01 ° ~ 5 °)
Note. The above indicators are not used for qualifying, for reference only.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1558-2016_English be delivered?Answer: Upon your order, we will start to translate JJF 1558-2016_English as soon as possible, and keep you informed of the progress. The lead time is typically 3 ~ 5 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1558-2016_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1558-2016_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|