Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)StatusPDF
JJF 1226-2009English879 Add to Cart 4 days [Need to translate] Calibration Specification of the Clinical Electronic Thermometer Obsolete JJF 1226-2009
JJF 1226-1990English279 Add to Cart 3 days [Need to translate] (Technical Specification reference operating gas flow from 0. 3 to 3 cubic meters/time) Obsolete JJF 1226-1990

PDF similar to JJF 1226-2009


Standard similar to JJF 1226-2009

JJF 1366   JJF 1187   JJF 1107   JJF 2019   JJF 1909   JJF 1908   

Basic data

Standard ID JJF 1226-2009 (JJF1226-2009)
Description (Translated English) Calibration Specification of the Clinical Electronic Thermometer
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A54
Classification of International Standard 17.200
Word Count Estimation 22,271
Date of Issue 2009-07-10
Date of Implementation 2009-10-10
Quoted Standard JJG 111-2003; JJF 1007-2007; JJF 1071-2000; JJF 1171-2007; GB/T 21416-2008; OIML R115; EN 12470-3-2000; ITS 90-1990
Regulation (derived from) AQSIQ Announcement No. 78 of 2009
Issuing agency(ies) General Administration of Quality Supervision, Inspection and Quarantine
Summary This standard applies to internal batteries, intermittent measurements surplus human or animal armpit, mouth, rectum and other different parts of the body, in digital form displays the measured temperature value and functionality with the highest temperature is kept non-contact type of medical electronic thermometer top side (hereinafter referred to as the thermometer) calibration.

JJF 1226-2009: Calibration Specification of the Clinical Electronic Thermometer

---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification of the Clinical Electronic Thermometer People's Republic of China National Metrology Technical Specifications Medical electronic thermometer calibration specification Posted 2009-07-10 2009-10-10 implementation The State Administration of Quality Supervision, Inspection and Quarantine released Medical electronic thermometer calibration specification This specification by the State Administration of Quality Supervision, Inspection and Quarantine approved on July 10, 2009, and from From October 10, 2009 into effect. Focal point. the National Technical Committee of temperature measurement Main drafting unit. Zhejiang Institute of Metrology China Institute of Metrology Shanghai Measurement and Testing Technology Research Institute Participated in the drafting. Hangzhou Hua'an health care supplies Co., Ltd. Terumo Medical Products (Hangzhou) Co., Ltd. This specification temperature measured by the National Technical Committee is responsible for interpretation The main drafters of this specification. Shen Zhong only (Zhejiang Institute of Metrology) Zhou Lianqin (Zhejiang Institute of Metrology) Chen Yue (China Institute of Metrology) Yao Lifang (Shanghai Institute of Measurement and Testing) Drafters participate. Sun Wenyu (Hangzhou Hua'an health care supplies Co., Ltd.) Mao Jian (Terumo Medical Products (Hangzhou) Co., Ltd.) Shou Wenjie (Zhejiang Institute of Metrology)

table of Contents

1 Scope (1) 2. Referenced Documents (1) 3 Terms (1) 4 Overview (1) 5 Metrological characteristics (2) 5.1 Measurement range (2) 5.2 Resolution (2) 5.3 Indication tolerance limit (2) Functional checks 5.4 (2) 6 calibration conditions (3) Environmental conditions 6.1 (3) 6.2 standards and other measuring equipment (3) 7 calibration items and calibration methods (3) 7.1 calibration items (3) 7.2 Check before calibration (3) 7.3 Calibration method (4) 7.4 Data Processing (4) Expression 8 calibration results (5) 9 Recalibration interval (6) Appendix A calibration record format example (7) Appendix B calibration certificate reference page format (9) Appendix C standard platinum resistance thermometers (30 ~ 45) ℃ when the reference function table (10) Appendix D Examples of measurement uncertainty analysis (11) Medical electronic thermometer calibration specification

1 Scope

This specification applies to the internal battery, intermittent measurements in different parts of the human or animal armpit, oral, rectal, etc. The temperature in digital form displays the measured temperature value and the temperature was kept contact with the highest non-predictive function electronic medical Thermometer (hereinafter referred thermometer) calibration.

2 Citations

This specification references the following documents. JJG111-2003 "glass thermometer test procedures" JJF 1007-2007 "temperature measurement terminology and definitions" JJF 1071-2000 "National Calibration Specifications writing rules" JJF 1171-2007 "Temperature Data Loggers Calibration norms" GB/T 21416-2008 "Medical electronic thermometer" OIMLR115Clinicalelectricalthermometerswithmaximumdevice (with maximum display Medical electronic thermometer device) EN12470-3.2000Clinicalthermometers-Part 3. Performanceofcompactelectrical thermometers (non-predictiveandpredictive) with maximum device (medical thermometers - Part 3. Compact electrical device with maximum display Sub thermometers (non-predictive and predictive) performance) ITS-90 1990 International Temperature Scale When using this specification it should be noted that using the currently valid version of the documents cited.

3 Terms

3.1 temperature probe (temperatureprobe) One part of the thermometer, the thermometer temperature measurement site, by the temperature sensor, cover, plastic closures, connectors Wire connection plugs and other components. 3.2 display unit (indicatingunit) One of the components of the thermometer, the temperature sensor for processing the output signal, and displays the measured temperature. 3.3 Compact electronic thermometers (compactelectricalthermometers) Temperature probe, and a display unit permanently close integral electronic thermometer. 3.4 Separate electronic thermometer (separateelectricalthermometers) Temperature probe and a display unit separation, electronic thermometers connected by wires.

4 Overview

Thermometer mainly by the temperature probe, ASIC and a digital display unit, battery-powered, for measuring