US$209.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. GB/T 29254-2012: Laboratory autoclave controller Status: Valid
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| GB/T 29254-2012 | English | 209 |
Add to Cart
|
3 days [Need to translate]
|
Laboratory autoclave controller
| Valid |
GB/T 29254-2012
|
PDF similar to GB/T 29254-2012
Basic data | Standard ID | GB/T 29254-2012 (GB/T29254-2012) | | Description (Translated English) | Laboratory autoclave controller | | Sector / Industry | National Standard (Recommended) | | Classification of Chinese Standard | N61 | | Classification of International Standard | 71.120.10 | | Word Count Estimation | 9,922 | | Quoted Standard | GB/T 191-2008 | | Regulation (derived from) | National Standards Bulletin No. 41 of 2012 | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China | | Summary | This standard specifies the laboratory reactor conditions of use of the controller, requirements, test methods, inspection rules, marking, packaging, transportation and storage. This standard applies to the design pressure of not greater than 35 MPa, capa |
GB/T 29254-2012: Laboratory autoclave controller---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Laboratory autoclave controller
ICS 71.120.10
N61
National Standards of People's Republic of China
Laboratory reactor with controller
Issued on. 2012-12-31
2013-06-01 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China released
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
This standard is proposed and managed by China Machinery Industry Federation.
This standard is drafted by. Weihai Chemical Machinery Co., Ltd., Dalian Mastery reactor plant.
Participated in the drafting of this Standard. China Instrument Industry Association Branch of laboratory equipment, the special prosecutor hospital Weihai, Shandong Province Branch of Machinery Industry
Instrumentation Technology and Economy Institute, Yantai Keli Chemical Equipment Co., Ltd., Weihai Xintai Chemical Machinery Co., Ltd. of Yantai Songling
Equipment Co., Ltd.
The main drafters of this standard. Ma Wenqi, Wang bacon, Wang Jialong, Cai Weiwei, Wang Yumin, Li Gang Min, Eun end, Madian Qing, YUAN Jian-ming, Jin Lihui,
Following Xia Zhen, Deng Ai Qun, Wang Tianzhao, Lee Van.
Laboratory reactor with controller
1 Scope
This standard specifies the conditions of use with laboratory reactor controllers, requirements, test methods, inspection rules, marking, packaging, transport and
Storage.
This standard applies to the design pressure of not more than 35MPa, the volume of not more than 24L, stirring with a magnetic heating device to drive the reaction
Kettle controller (hereinafter referred to as the controller).
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 191-2008 Packaging - Pictorial signs
3 Conditions of Use
3.1 Environmental conditions
3.1.1 temperature (0 ~ 40) ℃.
3.1.2 relative humidity less than 85%.
3.1.3 atmospheric pressure is 80kPa ~ 106kPa.
3.1.4 around the non-flammable, explosive, corrosive gas and conductive dust.
3.1.5 mechanical vibration to its influence is negligible.
3.1.6 In addition to the geomagnetic off, other small external magnetic field to its influence is negligible.
3.2 Supply conditions
3.2.1 rated voltage of AC 220V, allows excursions range from -15% to 10%.
3.2.2 rated frequency of 50Hz, allow excursions range is ± 2%.
4 Requirements
4.1 Appearance
4.1.1 metal plating and chemical treatment of the surface should be uniform color, no exposed at the end, from the floor, blistering, scarring or abrasions and scratches obvious.
4.1.2 paint coating the surface should be smooth, uniform color, no blistering, peeling and obvious scratches.
4.1.3 The controller should not prevent the readout with defects.
4.1.4 knobs, switches, connectors and other properly installed, solid, flexible and comfortable to operate.
4.1.5 Nameplate clear and smooth.
4.2 Temperature Control
4.2.1 Temperature control error
At room temperature to a temperature control range of 700 ℃, temperature control error limit of ± 1 ℃.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB/T 29254-2012_English be delivered?Answer: Upon your order, we will start to translate GB/T 29254-2012_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB/T 29254-2012_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 29254-2012_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|