US$369.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. GB 17342-2009: [GB/T 17342-2009] Ophthalmic instrument -- Trial case lenses Status: Valid GB 17342: Evolution and historical versions
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| GB 17342-2009 | English | 369 |
Add to Cart
|
3 days [Need to translate]
|
[GB/T 17342-2009] Ophthalmic instrument -- Trial case lenses
| Valid |
GB 17342-2009
|
| GB 17342-1998 | English | 399 |
Add to Cart
|
3 days [Need to translate]
|
Ophthalmic instrument--Trial case lenses
| Obsolete |
GB 17342-1998
|
PDF similar to GB 17342-2009
Basic data | Standard ID | GB 17342-2009 (GB17342-2009) | | Description (Translated English) | [GB/T 17342-2009] Ophthalmic instrument -- Trial case lenses | | Sector / Industry | National Standard | | Classification of Chinese Standard | C40 | | Classification of International Standard | 11.040.70 | | Word Count Estimation | 14,131 | | Date of Issue | 2009-06-12 | | Date of Implementation | 2010-07-01 | | Older Standard (superseded by this standard) | GB 17342-1998 | | Quoted Standard | GB/T 2828.1; GB/T 10050; JJG 580 | | Adopted Standard | ISO 9801-1997, MOD | | Regulation (derived from) | Announcement of Newly Approved National Standards No. 8, 2009 (No. 148 overall) | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China | | Summary | This Chinese standard relates to testing the refractive defects of the human eye optometry lenses with a variety of basic requirements. This standard applies to all types of full aperture and narrow aperture optometry lenses. |
GB 17342-2009: [GB/T 17342-2009] Ophthalmic instrument -- Trial case lenses---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Ophthalmic instrument.Trial case lenses
ICS 11.040.70
C40
National Standards of People's Republic of China
Replacing GB 17342-1998
Ophthalmic instruments optometry lenses
(ISO 9801.1997, MOD)
Posted 2009-06-12
2010-07-01 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China released
Foreword
All technical contents of this standard is mandatory.
This revised standard adopts ISO 9801.1997 "ophthalmic instruments refraction lens."
This standard and ISO 9801.1997 main technical differences are as follows.
--- The standard Chapter 4. Classification and use for the new section, the auxiliary lens description;
--- The standard 5.2. Added configuration requirements. the basic configuration of refraction of the lens, auxiliary lens configuration;
--- The standard 5.3. optical performance requirements. cylinder of ball lens tolerance; prism prism sheet, spherical degree, cylinder degree of tolerance;
The geometric center of the lens prism ball lens and the lens sub-file column and tolerance; cylinder axial binning and tolerance; prism sheet baseline
Binning and tolerance; make adjustments and provisions of the cross cylinder technical requirements;
--- The standard 5.4. Structural requirements. to adjust the thickness of refraction of the lens; auxiliary lens structure to provide for;
--- The standard 6.3. added auxiliary lens configuration of test methods;
--- The standard 7.1. Added requirements for ball lens, a lenticular lens, a prism sheet single pass rate and the rate of qualified FCL; made the wrong piece
Regulations;
--- The standard 8.1. double-sided lettering optometry lenses prescribed a variety of technical indicators should meet the optical performance requirements.
This standard replaces GB 17342-1998 "ophthalmic instruments refraction lens."
This standard and GB 17342-1998 "ophthalmic instruments optometry lenses," compared to the main changes are as follows.
--- The standard Chapter 4. Classification and use for the new section, the auxiliary lens description;
--- The standard 5.2. Added configuration requirements. the basic configuration of refraction of the lens, auxiliary lens configuration;
--- The standard 5.3. optical performance requirements. cylinder of ball lens tolerance; prism prism sheet, spherical degree, cylinder degree of tolerance;
Ball lenses and lenticular sheet lens geometric center of the prism binning and tolerance; cylinder axial binning and tolerance; cross-cylinder
Make adjustments and provisions of technical requirements;
--- The standard 5.4. Structural requirements. to make adjustments to the thickness of the refraction of the lens, the outer diameter of the frame; auxiliary lens structure made
Regulations;
--- The standard 6.3. added auxiliary lens configuration of test methods;
--- The standard 7.1. adjust the requirements of the ball lens, a lenticular lens, a prism sheet single pass rate and the rate of qualified FCL; made the wrong piece
Regulations;
--- The standard 7.3. Sampling checklist to make adjustments;
--- The standard 8.1. double-sided lettering optometry lenses prescribed a variety of technical indicators should meet the optical performance requirements.
The standard proposed by China Institute of Metrology.
This standard by the National Standardization Technical Committee Optics and optical instruments - Optical glasses Technical Committee (SAC/TC103/SC3)
Centralized.
This standard is mainly drafted by. China Institute of Metrology, Shandong Institute of Metrology, Lianyungang days Connaught Optical Instrument Co.,
Division Shanghai Instrument Co., Ltd. ASE.
The main drafters of this standard. Liu Wenli, Chen Yan, Ren Hongwei, Yang Lei, Ning Lixin, chronicles post.
This standard was first published in 1998.
Ophthalmic instruments optometry lenses
1 Scope
This standard specifies the basic requirements for testing the human eye refractive defects of various refraction lens.
This standard applies to all types of full bore and reduced aperture trial case lenses.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent
Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research
Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 2828.1 Sampling procedures for inspection - Part 1. by acceptance quality limit (AQL) retrieval batch inspection sampling plan
(GB/T 2828.1-2003, ISO 2859-1.1999, IDT)
GB/T 10050 Optics and optical instruments - Reference wavelengths (eqvGB/T 10050-1988, ISO 7944. 1984)
JJG580 focimeter
3 Terms and Definitions
The following terms and definitions apply to this standard.
3.1
Lens in meters measured paraxial vertex length of the countdown. A lens contains two vertex.
3.2
After the lens near the axis in meters measured at the top of the reciprocal of the focal length. 1, from the rear of the lens to focus the vertex axis is called after nearly FIG.
Name diopters.
Under normal circumstances, the vertex of the lens mean a subsequent vertex. Refraction lens vertex power and wavelength. This standard specifies the use of wave
Long green mercury line, λe = 546.07nm, see GB/T 10050.
F --- the object focus;
F '--- the image side focal point;
H --- object side principal point;
H '--- the image side principal point;
Figure 1 a schematic top focal length lens
3.3
Measure the light deviated by a predetermined point on the lens produced (usually the center of the lens) later. Figure 2, the prism is in centimeters
Per meter (cm/m), the name of the unit prism diopters.
Figure 2 a schematic view of a prism
3.4
Mosaic in the box, for the determination of the refractive defects of the human eye lens.
3.5
With 1mm wall thickness of the protective frame and has a maximum limit aperture lenses.
3.6
Aperture is significantly smaller than the outer diameter of the frame thickness significantly thinner lenses.
3.7
Ball lens, a lenticular lens or ball - column joint optometry lenses chain. In its last time measured back vertex power equal in the frame of the
United have a nominal value of refraction of the lens in the meridian.
3.8
Paraxial parallel beam converge at a point of refraction lenses.
3.9
Paraxial parallel light beam to two separate, mutually orthogonal cross line contains two main vertex of the refraction lens.
Lenticular lens optometry wherein a vertex principal meridian zero. Lenticular refers to the degree of its top vertex is not zero and that a principal meridian
Power.
4 Classification and Uses
4.1 Purpose
In practice, a combination of all kinds of refraction of the lens is called a trial lens. Hospital of the trial lens, glasses shops for inspection
An ophthalmic measuring instruments to check the human eye's refractive state and strabismus, amblyopia and other visual features of the human eye.
Optometry lenses mainly composed of positive and negative refraction spherical lenses, positive and negative cylinder lenses optometry, optometry prism lenses, refraction lenses and other auxiliary components.
4.2 spherical lenses optometry
By the positive and negative refraction spherical lens elements. Positive spherical lenses for the human eye optometry hyperopia and presbyopia inspection; negative refraction spherical lenses for people
Myopic eye examination.
4.3 lenticular lens optometry
By the positive and negative refraction lens cylinder components. N lenticular lens of the human eye optometry hyperopia and astigmatism, presbyopia check for. negative cylinder lenses optometry
The human eye myopic astigmatism check.
4.4 prism refraction lens
Check for the human eye strabismus and heterophoria.
4.5 Supplemental trial lenses
4.5.1 Cross-cylinder lens
Cross-cylinder lens is a special lenticular lens, on its two mutually perpendicular orientations, respectively, they have the same value but opposite sign
Positive and negative cylinder vertex. Used to check the cylinder axial and cylinder degrees.
4.5.2 Ma pole piece
Ma pole piece consists of a row of equal diameter, smooth translucent cylindrical components. Check for strabismus and phoria.
4.5.3 black sheet
Also known as black sheet covering sheet, it is a totally opaque cover sheet. For covering is not checked monocular.
4.5.4 pinhole film
Pinhole film is at the center of a black piece of a through aperture. Mainly used to identify the vision loss caused by refractive error was cited by the eye disease
Onwards. Light through the hole should be round and smooth, and the rest opaque.
4.5.5 fractured pieces
There is a light transmissive slit on the light shielding sheet for checking astigmatism.
4.5.6 scrub films
Frosted translucent sheet, for children or outdoor instead of black film.
4.5.7 plain sheet
Plano lens is opaque to check simulated blindness and so on.
4.5.8 Cross sheet
On plain piece engraved with "ten" shaped groove, it is mainly used to determine the center of the eye and the eye position determination when checking strabismus.
4.5.9 Filter
It should be plain lenses. Usually red, green lens, check for color vision.
Red and green sheet with the use of tablets for binocular stereo vision inspection, refractive media opacity visual function; red-chip can also be used for amblyopia
Treatment, color vision examination; there are brown films such as photophobia eye for visual inspection of lenses.
4.5.10 polarizer
So that natural light into a plane polarized lens for examining the implicit vision, strabismus, unequal vision, stereoscopic vision and other visual functions.
5 Requirements
5.1 General
Refraction lens FCL factory should meet the requirements of 5.2,5.3,5.4 and 5.5, according to the requirements of Chapter 6 of the inspection.
FCL 5.2 Configuration Requirements
5.2.1 refraction spherical lens
The same size balls trial case lenses should be about two.
a) must include a + lens measuring range 0.12m-1 and -0.12m-1, at least in -12.00m-1 ~ + 12.00m-1;
b) In -0.25m-1 ~ -4.00m-1 and + 0.25m-1 lens values within the range of ~ + 4.00m-1 intervals not greater than
0.25m-1;
c) In the < -4.00m-1 ~ -8.00m-1 and> + 4.00m-1 ~ + 8.00m-1 lens value within the range of intervals not greater
At 0.50m-1;
d) In the < -8.00m-1 ~ -12.00m-1 and> + 8.00m-1 lens values within ~ + 12.00m-1, the interval can not be
Greater than 1.00m-1.
5.2.2 lenticular lens optometry
The same specifications lenticular lens optometry should be about two.
a) must include a + lens measuring range 0.12m-1 and -0.12m-1, at least in -4.00m-1 ~ + 4.00m-1;
b) In -0.25m-1 ~ -3.00m-1 and + 0.25m-1 lens values within the range of ~ + 3.00m-1 intervals not greater than
0.25m-1;
c) In the < -3.00m-1 ~ -4.00m-1 and> + 3.00m-1 ~ + 4.00m-1 lens value within the range of intervals not greater
At 0.50m-1.
5.2.3 prism refraction lens
Prismatic refraction lens should meet the following requirements.
a) measuring prism refraction lens range to at least 8.0cm/m, where 0.5cm/m, 1.0cm/m, 2.0cm/m shall have two;
b) within 3.0cm/m ~ 8.0cm/m range lenses magnitude intervals not greater than 1.0cm/m.
5.2.4 Auxiliary lenses optometry
Auxiliary lens should include cross-cylinder lenses, maddox rod lenses, pinhole film, black film, fractured pieces, plain film, scrub films, red and green filters and the like.
5.3 Optical Performance Requirements
5.3.1 plano lens vertex tolerance
Vertex plain lens tolerance should be consistent with the requirements of Table 1.
Table 1 plano lens vertex tolerance
Vertex nominal value
m-1
Tolerance (MPE)
Spherical degree
m-1
Lenticular degree
m-1
Prism
cm/m
0 ± 0.03 ± 0.03 0.06
5.3.2 refraction spherical lens vertex tolerance
Refraction spherical lens vertex tolerance shall comply with the requirements of Table 2.
Table 2 refraction spherical lens vertex tolerance in meters -1
Vertex nominal value (absolute value)
Tolerance (MPE)
Spherical degree of cylinder
0.12 ± 0.03 ± 0.03
> 0.12 ~ 6.00 ± 0.06 ± 0.03
> 6.00 ~ 12.00 ± 0.09 ± 0.03
> 12.00 ± 0.12 ± 0.05
5.3.3 optometry lenticular lens vertex tolerance
Cylinder vertex optometry degree lenticular lens tolerance should be consistent with the requirements of Table 3; its spherical degree of tolerance are not allowed to exceed ± 0.06m-1, its solid
There are no larger than the prism 0.12cm/m.
Table 3 lenticular lens optometry lenticular vertex tolerance in meters -1
Cylinder of the nominal value (absolute value) Tolerance (MPE)
0.12 ± 0.03
> 0.12 ~ 1.00 ± 0.06
> 1.00 ~ 4.00 ± 0.09
> 4.00 ~ 6.00 ± 0.12
> 6.00 ± 0.18
5.3.4 prism refraction lens tolerance
Prismatic refraction prism lens tolerance should be consistent with the requirements of Table 4, in 0cm/m ~ prism within 5.0cm/m range optometry lenses
Spherical degree, cylinder degree shall not exceed ± 0.05m-1.
Table 4 prism sheet prism tolerance in centimeters per meter
Prism nominal value tolerance (MPE)
> 0 ~ 3.0 ± 0.10
> 3.0 ~ 6.0 ± 0.12
> 6.0 ± 0.25
5.3.5 optical center displacement tolerance
Spherical lenses optometry, optometry optical center displacement lenticular lens is represented by the prism lens at the geometric center of the circle, which allows the displacement of the optical center
The difference shall be in accordance with Table 5.
Table 5 optical center displacement tolerance
Vertex nominal value (absolute value)
m-1
Prism tolerance (MPE)
cm/m
≥0.12 ~ 1.00 ± 0.12
> 1.00 ~ 4.00 ± 0.25
> 4.00 ~ 7.00 ± 0.35
> 7.00 ~ 10.00 ± 0.50
> 10.00 ~ 12.00 ± 0.60
> 12.00 ± 0.80
5.3.6 lenticular lens optometry axial tolerance
Lenticular lens optometry axial line is defined as 0 ° ~ 180 ° direction, marked with the axial deviation of the lens diameter at both ends with the angle between the
He said its value should be in accordance with Table 6.
Table 6 cylinder lens axial tolerance
Cylinder of the nominal value (absolute value)
m-1
Axial tolerance (MPE)
≥0.12 ~ 0.50 ± 3 °
> 0.50 ± 2 °
5.3.7 prism sheet baseline tolerance
Axial positioning prism sheet substrate by the prism baseline represented, the deviation should be in accordance with Table 7.
Table 7 prism baseline tolerance
Prism nominal value
cm/m
Tolerance (MPE)
> 0 ~ 0.50 ± 5 °
> 0.50 ~ 1.00 ± 4 °
> 1.00 ~ 2.00 ± 3 °
> 2.00 ~ 10.00 ± 2 °
> 10.00 ± 1 °
5.3.8 Cross-cylinder lens optical performance requirements
Cross-cylinder spherical degree of tolerance shown in Table 2, the cylinder of the tolerance shown in Table 3, the displacement of the optical center of the tolerance shown in Table 5, Table 6 axial tolerance.
Note. The axial cross-cylinder lens negative cylinder applications marked in red, positive cylinder axis bit application blue (or black) mark. Mark in two different colors, different
Measurement mode, the optical performance requirements should meet those requirements.
5.4 Structural requirements
This requirement applies to all frames and lenses installed.
5.4.1 Size
Optometry lenses with some frame edges are rounded, the radius of the arc is not greater than 1.4mm. Installed outside diameter of the lens frame
To 38 0-0.2mm.
Refraction lens together with its frame, including the overall thickness of not more than 2.8mm. Optometry lenses should fit into the interval for each piece of 3mm
Audition holder. Which is greater than and equal to the absolute value of the full aperture ball lens thickness 4.00m-1 allows more than 2.8mm, greater than and equal to
3cm/full aperture and narrow aperture prism sheet thickness close to the object side, the side of the m allowed to exceed 2.8mm.
5.4.2 effective aperture
Absolute value of the vertex is less than the nominal value of the refraction lens 12m-1 through effective aperture of not less than 18mm; the nominal value of the absolute vertex
Values greater than the refraction lens 12m-1 through the effective aperture of not less than 16mm.
5.4.3 Structure
5.4.3.1 frame structural requirements
Refraction lens frame surface should be smooth, without having any tip can cause injury to the patient or optometrist edges, sharp corners or rough surface, and
No significant deformation and cracks. Optometry lens and frame should be firmly fixed, no loose, mosaic should be neat. Frame identification shall be in accordance with Table 9
Claim.
5.4.3.2 prism sheet structure requirements
Prism sheet should be installed in the side of the eye so close to the surface of the frame parallel to the mounting surface. Standard nominal values on the vertical frame should be light
Incident on the near side eye when prism, and consistent with the measured values focimeter.
Ma pole piece structure 5.4.3.3 requirements
Difference in thickness between the pole pieces Markov semi-cylindrical ends of the axial direction of the plane should not exceed 0.08mm.
5.4.3.4 pinhole film structure requirements
Diameter through-aperture pinhole film should be within the range of 0.5mm ~ 2.0mm.
5.4.3.5 Structure requirements fractured pieces
Fractured pieces of the light transmissive slit width should be within the range of 0.5mm ~ 2.0mm.
5.5 Material and surface quality
5.5.1 Materials
Frame materials used should not contain any corrosive elements.
5.5.2 surface quality
Optometry in the effective lens aperture not have bubbles, defects, impurities, scratches and any visually observe an irregular surface
defect. Optometry translucent lens should be good, without mildew.
6 test method
6.1 optical performance test
6.1.1 Test Method vertex
When vertex inspection of the lens, the lens should be measured imaging centers coincides with the reticle crosshairs and prism Indication minimum or
Zero.
6.1.2 optical center displacement test methods
Inspection spherical lenses optometry, optometry lenticular lens, the optical center of the cross cylinder displacement may first focimeter prism measurement mode
Adjust to polar coordinate display (such as "P-B" mode, etc.), the geometric center of the test lens is located at the geometric center of the lens holder, measured at this time
Prismatic indication is obtained by displacement of the test optical center of the lens.
6.1.3 Axial marker test method
When axial test lenticular lens optometry mark, three points should be labeled with axial connection focimeter printing mechanism of the lens ring at both ends coincide, that is,
Readable axial deviation of the test lens.
6.1.4 prism test method
When the test prism refraction prism lens to the imaging of the horizontal axis and the reticle crosshair horizontal axis coincide. Obtained at this time
Prism lens under test, the degree spherical and cylinder readings.
6.1.5 The prism sheet baseline test methods
When the test prism refraction lens prism baseline, following the step 6.1.4, then fine rotating the subject lens, lens ring so that the baseline marks the end
And three connection focimeter printing mechanism coincide. Baseline readings were obtained at this time deviation detection lens.
6.1.6 Test Equipment
Optical performance testing should be used in accordance with the test procedures certified by JJG580 primary standard focimeter. In the subject of optometry lenses
Before test, a standard focimeter indoors holding time is not less than 2h, while maintaining a temperature of 20 ℃ ± 5 ℃, relative humidity less than
It tested 85% of environmental conditions.
When refraction lens inspection, should first be a standard focimeter indexing, select Settings 0.01m-1 division, the Abbe number 58 attached
Near the standard wavelength selection e line, while focusing lens mount meter inspection, detection can be carried out after removal of dust.
When refraction lens inspection shall be placed on the upper face of the test lens logo focimeter lens holder, if the lens needs double-sided lettering
Identification of the two sides were tested.
6.2 Material and surface quality inspection
Without the aid of the optical amplification device, be tested in the lens bright-field, dark background. Shows the recommended quality inspection system schematic 3
FIG. Illumination around the test apparatus is about 200lx. Recommended lighting using 15W more than 60W fluorescent or incandescent,
Illumination placement at the illustrated lenses should not be lower than 350lx.
Note. This observation method has certain subjectivity, requires inspectors have some practical experience.
1 --- non-reflective black background (150mm × 360mm);
2 --- light source;
3 --- visor (adjustable baffle);
4 --- movable refraction lens;
5 --- eye observation plane;
Rao --- adjustable visor distance (mm).
Refraction lens surface quality inspection apparatus of FIG. 3 schematic
6.3 Structural testing
6.3.1 Frame structure test method 6.2 visual reference to be.
6.3.2 Inspection Ma pole piece structure. Using the minimum scale of not more than 0.01mm micrometer, measuring Ma pole piece semi-cylindrical axial
The thickness difference between both ends of the planar direction.
6.3.3 Structural testing of fractured pieces and pinholes tablets. Using the minimum scale of 0.01mm of digital calipers.
7 Inspection rules
7.1 FCL refraction lens factory, such as the following occurs. refraction spherical lenses, lenticular lenses optometry, optometry cross between a prism lens of the three
Assembly phase error; spherical lenses optometry, optometry lenticular lens, the positive and negative error between the lens assembly; all kinds of refraction lens interrupted, missing pieces; spherical
Optometry degree spherical lenses, cylinder lenses optometry degree cylinder, prism index of refraction of the lens prism of more than 2 times the respective tolerance; that is deemed
"Wrong piece."
7.2 manufactured products according to the requirements of this standard, according to the case overall acceptance. Each box requires spherical refraction lenses, lenticular lenses and optometry
Single pass rate prism refraction lenses shall not be less than 90%; FCL pass rate of not less than 90%. FCL when the factory, and one appears a
More wrong piece, FCL is considered substandard.
7.3 batch manufactured products according to GB/T 2828.1 special inspection level of S-4, AQL 4.0 sampling requirements for the inspection, see
Table 8.
Table 8
Bulk products range
Sampling Sample size
Acceptance number
Ac
Failed to determine the number of
Re
2 ~ 90301
91 ~ 5001312
7.4 pairs of special specifications of the products according to the requirements of both supply and demand agreement separately.
8 signs, packaging, transportation, storage
8.1 mark
8.1.1 all kinds of vertex or prismatic refraction lens nominal value shall be marked on the frame that the opposite side of the eye, if the nominal value of the standard
Note on the frame of the double-sided, double-sided lens should ensure that the various technical indicators meet the tolerance requirements.
8.1.2 lenticular lens optometry optometry lenses and prisms should be marked axial or baseline, respectively, on the frame or lens. If it does not prevent the lenses
Measures box rotation should be marked cylinder axial or baseline prism directly on the lenses.
8.1.3 color flag of the frame helps to identify specific lens types and categories; and spherical degree, cylinder degree and prism can be directly
Read from the indication mark.
8.1.4 prism refraction lens mounting base surface should prism surface side optometry eye lens adjacent parallel. Marked on the name of the frame
Value usually refers to the incident light perpendicular to the prismatic refraction prism lens mounting base surface when.
8.1.5 various types of lenses in accordance with the requirements of Table 9 distinguish between different colors of the frame and (or) flags of different colors.
Table 9 Lens identification mark
Lens color Frame Type letters or symbols or identification mark
Ball column refraction lens vertex nominal value -
Positive + Black
Negative - Red
Δ refraction prism sheet white/gray
Table 9 (continued)
Lens color Frame Type letters or symbols or identification mark
Horse type lever piece MR -
I fractured pieces or SS -
Pinhole film ◎ PH or white/black/gray
● black sheet or BL -
Matte film FL -
Cross sheet
Or CL -
Red filter RF -
Green filter GF -
Polarizer PF -
8.2 Packaging
8.2.1 refraction lens category shall be packed in boxes.
8.2.2 Instructions shall be attached to the box, packing slips and box configuration specification must be consistent with the actual configuration, the description should include the following
content.
a) the manufacturer's name and address;
b) intro...
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB 17342-2009_English be delivered?Answer: Upon your order, we will start to translate GB 17342-2009_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB 17342-2009_English with my colleagues?Answer: Yes. The purchased PDF of GB 17342-2009_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version GB 17342-2009?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB 17342-2009 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|