Path:
Home >
MISC >
Page71 > GA/T 2049-2023
|| Home >
Standard_List >
MISC >
Page71 > GA/T 2049-2023
| Std ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status |
| GA/T 2049-2023 | English | RFQ |
ASK
|
3 days [Need to translate]
|
(Forensic Science Detection of 5 Substances including Acetone in Suspected Precursor Chemicals Gas Chromatography-Mass Spectrometry)
| Valid |
Standard similar to GA/T 2049-2023
Basic data
| Standard ID | GA/T 2049-2023 (GA/T2049-2023) |
| Description (Translated English) | (Forensic Science Detection of 5 Substances including Acetone in Suspected Precursor Chemicals Gas Chromatography-Mass Spectrometry) |
| Sector / Industry | Public Security (Police) Industry Standard (Recommended) |
| Classification of International Standard | 13.310 |
| Date of Issue | 2023-03-01 |
| Date of Implementation | 2023-12-01 |
| Issuing agency(ies) | Ministry of Public Security |
GA/T 2049-2023: (Forensic Science Detection of 5 Substances including Acetone in Suspected Precursor Chemicals Gas Chromatography-Mass Spectrometry)
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
ICS 13.310
CCSA92
Public safety industry standards of the People's Republic of China
Forensic science suspects acetone, etc. in precursor chemicals
5 substances tested by gas chromatography-mass spectrometry
Published on 2023-03-01
Implemented on 2023-12-01
Released by the Ministry of Public Security of the People's Republic of China
Foreword
This document complies with the provisions of GB/T 1.1-2020 "Standardization Work Guidelines Part 1.Structure and Drafting Rules of Standardization Documents"
Drafting.
Please note that some content in this document may be subject to patents. The publisher of this document assumes no responsibility for identifying patents.
This document is proposed and submitted by the Toxic Analysis Sub-Technical Committee of the National Technical Committee for Standardization of Criminal Technology (SAC/TC179/SC1).
Return to home.
This document was drafted by. Anti-Narcotics Intelligence Technology Center of the Ministry of Public Security.
The main drafters of this document. Fan Yingfeng, Chen Jie, Liu Peipei, Meng Xin, and Yan Jin.
Forensic science suspects acetone, etc. in precursor chemicals
5 substances tested by gas chromatography-mass spectrometry
1 Scope
This document describes the gas chromatography of acetone, diethyl ether, methyl ethyl ketone, chloroform, and toluene in suspected precursor chemicals used in forensic science.
Spectrometry-mass spectrometry (GC-MS) qualitative testing method.
This document is applicable to the characterization of acetone, ether, methyl ethyl ketone, chloroform, and toluene in suspected precursor chemicals in the field of forensic science.
analyze.
2 Normative reference documents
The contents of the following documents constitute essential provisions of this document through normative references in the text. Among them, the dated quotations
For undated referenced documents, only the version corresponding to that date applies to this document; for undated referenced documents, the latest version (including all amendments) applies to
this document.
GA/T 122 Toxicological analysis terminology
3 Terms and definitions
The terms and definitions defined in GA/T 122 apply to this document.
4 Principles
Prompt the acetone, ether, methyl ethyl ketone, chloroform, and toluene in the suspected precursor chemicals (see Appendix A for compound information).
Pick. Gas chromatography-mass spectrometry was used for detection, and retention time, mass spectrum characteristic ions and ion abundance ratio were used as the basis for qualitative judgment.
5 Reagents and materials
5.1 Reagents
Unless otherwise stated, the reagents used in the analysis are of chromatographic grade and include.
a) Anhydrous sodium sulfate. analytical grade;
b) methanol;
c) Standard solution.
1) Standard material stock solutions. standard material stock solutions of acetone, diethyl ether, methyl ethyl ketone, chloroform, and toluene;
2) Standard working solution. Pipette an appropriate amount of standard material stock solution, dilute it with methanol, and prepare 0.1 mg/mL acetone and acetone respectively.
Standard working solution of diethyl ether, methyl ethyl ketone, chloroform and toluene, refrigerated at 0℃~4℃, valid for 1 month; also
Qualitative standard working solutions of other concentrations used in experiments can be prepared according to actual conditions.
5.2 Materials
Materials include.