HJ/T 99-2003 PDF in English
HJ/T 99-2003 (HJ/T99-2003, HJT 99-2003, HJT99-2003)
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Name of Chinese Standard | Status |
HJ/T 99-2003 | English | 130 |
Add to Cart
|
0-9 seconds. Auto-delivery.
|
The techinical requirement for water quality automatic analyzer of dissolved oxygen(DO)
| Valid |
Standards related to (historical): HJ/T 99-2003
PDF Preview
HJ/T 99-2003: PDF in English (HJT 99-2003) HJ/T 99-2003
HJ
ENVIRONMENTAL PROTECTION INDUSTRY STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
The Technical Requirement for Water Quality
Automatic Analyzer of Dissolved Oxygen (DO)
ISSUED ON: MARCH 28, 2003
IMPLEMENTED ON: JULY 1, 2003
Issued by: Ministry of Environmental Protection of the People’s Republic
of China
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative References ... 4
3 Terms and Definitions ... 4
4 Type and Determination Range ... 5
5 Operating Voltage and Frequency ... 5
6 Performance Requirements ... 5
7 Composition of Instrument ... 6
8 Inspection Methods ... 7
9 Marking ... 10
10 Operating Instructions ... 11
11 Validation ... 11
The Technical Requirement for Water Quality
Automatic Analyzer of Dissolved Oxygen (DO)
1 Scope
This Standard specifies the technical performance requirements and performance test
methods of electrode method-based dissolved oxygen (DO) (0 ~ 35 °C) water quality
automatic analyzer for surface water, industrial sewage and municipal sewage.
2 Normative References
Through the reference in this Standard, the clauses of the following documents
become clauses of this Standard. In terms of references with a specified date, all the
subsequent modification sheets (excluding the corrected content) or revised versions
are not applicable to this Standard. However, the various parties that reach an
agreement in accordance with this Standard are encouraged to explore the possibility
of adopting the latest version of these documents. In terms of references without a
specified date, the latest version is applicable to this Standard.
GB 11913-89 Water Quality - Determination of Dissolved Oxygen - Electrochemical
Probe Method
3 Terms and Definitions
The following terms and definitions are applicable to this Standard.
3.1 Test Specimen
Test specimen refers to surface water, industrial sewage and municipal sewage
imported into the automatic analyzer.
3.2 Calibration Solution
Calibration solution is prepared to obtain the same indicated value as the DO
concentration of the specimen. There are several types of calibration solution as
follows.
3.2.1 Zero-point calibration solution.
3.2.2 Measuring range calibration solution.
or flickers, etc.
7.1.7 The words, symbols and markings describing the functions shall comply with the
stipulations of “9 Marking” of this Standard.
7.2 Composition: DO automatic analyzer is composed of detection unit, display and
recording unit, data processing and signal transmission unit, etc.
7.3 Sampling part: it shall have a complete and air-tight sampling system.
7.4 Measurement unit refers to the immersion of the electrode in the specimen and the
stable transmission of the generated signal to the display and recording unit. It is
composed of electrode supporting part and converter, etc.
7.4.1 Electrode is composed of anode, cathode, thermometer and electrolyte, etc. Use
an oxygen-permeable film (such as: fluororesin, polyethylene, silicone rubber, etc.) to
cover the electrode. It has a structure, in which, the specimen does not directly contact
with the anode and the cathode.
7.4.2 The electrode supporting part refers to electrode sleeves used for electrode
fixation. It is made of stainless steel, rigid polyvinyl chloride, polypropylene and other
materials that would not be corroded by the specimen.
7.4.3 Converter and display: they shall have a water-proof structure; the distance
between the electrode and the converter shall be as short as possible.
7.5 Display and recording unit: it shall have the function of displaying, recording and
printing dissolved oxygen concentration (mg/L) in equal division and the digital form.
7.6 Data transmission unit: it shall have a complete data collection and transmission
system.
7.7 Auxiliary devices: in accordance with the demands, the automatic analyzer may be
equipped with the following auxiliary devices.
7.7.1 Electrode cleaning device: it refers to a cleaning device, which uses fluids like
water and air, to clean the electrode.
7.7.2 Automatic water collection device: it refers to a device that automatically collects
specimens and transmits it to the electrode at a certain flow rate.
8 Inspection Methods
8.1 Test Conditions
8.1.1 Ambient temperature: between 10 ~ 40 °C; during the test, the temperature
variation is within ± 5 °C/d.
specified by the manufacturer.
8.2.3.3 Adjustment: alternately perform zero-point calibration and measuring range
calibration operations; adjust the analyzer, until the difference between the determined
value of the calibration solution and the displayed value is within ± 0.25 mg/L.
8.3 Performance Test Methods
8.3.1 Repeatability error: under the test conditions of 7.1, immerse the electrode in the
measuring range calibration solution; while using a magnetic stirrer to stir it,
continuously determine it for 6 times. Record each determined value; calculate the
relative standard deviation.
8.3.2 Zero drift: adopt the zero-point calibration solution to continuously determine for
24 h. Use the initial zero value (the average value of the first 3 determined values)
within this period of time to calculate the percentage of the difference between the
maximum variation amplitude and the initial zero value relative to the measuring range
value.
8.3.3 Measuring range drift: adopt the measuring range calibration solution; before and
after the zero drift test, while using a magnetic stirrer to stir it, respectively determine
3 times; calculate the average value. Through the variation amplitude after subtracting
the zero drift component, calculate the percentage relative to the measuring range
value.
8.3.4 Response time: transfer the electrode from the measuring range calibration
solution to the zero-point calibration solution; determine the time required when the
displayed value reaches 1 mg/L.
8.3.5 Temperature compensation accuracy: respectively at (20 ± 0.5) °C and (30 ±
0.5) °C, prepare saturated dissolved oxygen solution. Respectively immerse the
electrode in the above-mentioned solutions; while using a magnetic stirrer to stir it,
read the respective indicated values (mg/L). Respectively determine the temperature
of the above-mentioned solutions (accurate to ± 1 °C). In accordance with the
determined results, obtain the difference of saturated dissolved oxygen concentrations
in Attached Table 1.
8.3.6 Mean time between failures: use actual water samples; continuously run for 2
months; record the total running time (h) and the number of failures (number of times);
calculate the mean time between failures (MTBF) ≥ 720 h/number of times (this
indicator may be evaluated on site).
8.3.7 Actual water sample comparison test: select 5 (or more) actual water samples;
respectively conduct comparison tests of three concentration levels (high, medium and
low) of each water sample through the automatic analyzer and the method in the
national standard (GB 11894-89). The number of the comparison tests for each water
10 Operating Instructions
In the operating instructions, at least the following related matters must be explained.
10.1 Selection of installation location.
10.2 Flow rate of specimen.
10.3 Piping and wiring.
10.4 Warm-up time.
10.5 Method of application.
10.5.1 Determination preparation and calibration method.
10.5.2 Calibration solution preparation method.
10.5.3 Method of determination operation.
10.5.4 Disposal when the determination is suspended.
10.6 Maintenance and inspection.
10.6.1 Daily inspection method.
10.6.2 Regular inspection method.
10.6.3 Cleaning of electrode system.
10.6.4 Countermeasures in case of failure.
10.7 Other matters needing attention during use.
11 Validation
11.1 Daily validation: the calibration cycle of reproducibility, drift and response time is
on-site validation at least once a month, which may be conducted automatically or
manually.
11.2 Supervisory validation: the installed continuous monitoring system must be
periodically validated; the results of the periodic validation shall be reported and
submitted to the corresponding administrative department on environmental protection.
The periodic validation shall be undertaken by a monitoring institution with
corresponding qualifications.
Periodic validation mainly includes: on-site comparison tests conducted in accordance
...... Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.
|