Standard ID | YY 1024-2013 (YY1024-2013) | Description (Translated English) | Oviduct ligating hook | Sector / Industry | Medical Device & Pharmaceutical Industry Standard | Classification of Chinese Standard | C36 | Classification of International Standard | 11.040.30 | Word Count Estimation | 8,843 | Older Standard (superseded by this standard) | YY 91024-1999 | Quoted Standard | GB/T 1220; GB/T 2828.1; GB/T 2829; GB/T 10610; YY/T 0149-2006; YY/T 1052 | Issuing agency(ies) | State Food and Drug Administration | Summary | This standard specifies: tubal extraction hook (hereinafter referred to as the extraction hook) structure type and material requirements, test methods, inspection rules, marking, packaging, transport and storage. This standard applies: for gynecological s |
YY 1024-2013
Oviduct ligating hook
ICS 11.040.30
C36
People's Republic of China pharmaceutical industry standards
Replacing YY 91024-1999
Tubal extraction hook
Issued on. 2013-10-21
2014-10-01 implementation
China Food and Drug Administration released
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
This standard replaces YY 91024-1999 "tubal extract the hook."
This standard is in the original standard YY 91024-1999 main technical differences.
--- Cancel the reference standard's number, and the standard number of references to be updated;
--- Delete the original standard B Tubal extraction hook;
--- Remove the original standard icon Central Africa use parts of the size, delete the original standard shown in dimensional tolerances. Illustrating a main part
Size expressed as ranges;
--- Modify the surface roughness in the original standard hook inside and at the weld.
Please note that some of the content of this standard may involve patents, the issuing authority of this standard does not assume the responsibility to identify these patents.
This standard by the national family planning equipment Standardization Technical Committee (SAC/TC169) and focal points.
This standard was drafted. Shanghai Medical Devices (Group) Co., surgical instruments factory, the State Food and Drug Administration Shanghai medical devices
Mechanical Quality Supervision and Inspection Center.
The main drafters of this standard. Liu Wei group, Yao Tianping, Wengbing Hao.
The standard standard replaces the previous editions.
--- WS2-235-1977;
--- GB 8671-1988;
--- YY 91024-1999.
Tubal extraction hook
1 Scope
This standard specifies the type and structure of the material extracted tubal hook (hereinafter referred to as the extraction hook), requirements, test methods, inspection rules, marking
Chi, packaging, transport and storage.
This standard applies to time for the purposes of gynecological surgery for tubal ligation tubal extracted with fallopian tubes to extract the hook.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 1220 stainless steel rods
GB/T 2828.1 Sampling procedures for inspection - Part 1. by acceptance quality limit (AQL) retrieval batch inspection sampling plan
GB/T 2829 periodic inspection Sampling procedures and tables (suitable for process stability test)
GB/T 10610 Geometrical product specifications (GPS) Surface texture Profile method - Rules and procedures for evaluation of the surface structure
YY/T 0149-2006 stainless steel medical equipment corrosion test methods
YY/T 1052 surgical instruments logo
3 structure type and material
3.1 Structure type
Extraction hook Types and basic dimensions should comply with the requirements of Figure 1.
3.2 Material
Extraction hook can be used 12Cr13 or 12Cr18Ni9 materials GB/T 1220 stipulated.
Note. may also be proven to be evaluated as safe, other materials comply with the requirements of the standard.
......
|