YY 0302.2: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
YY 0302.2-2016 | English | RFQ |
ASK
|
6 days [Need to translate]
|
Dentistry - Rotary bur instruments - Part 2: Finishing burs
| Obsolete |
YY 0302.2-2016
|
YY/T 0302.2-2016 | English | 819 |
Add to Cart
|
6 days [Need to translate]
|
Dentistry-Rotary bur instruments--Part 2: Finishing burs
| Valid |
YY/T 0302.2-2016
|
PDF similar to YY 0302.2-2016
Basic data Standard ID | YY 0302.2-2016 (YY0302.2-2016) | Description (Translated English) | Dentistry - Rotary bur instruments - Part 2: Finishing burs | Sector / Industry | Medical Device & Pharmaceutical Industry Standard | Classification of Chinese Standard | C33 | Classification of International Standard | 11.060.20 | Word Count Estimation | 37,332 | Date of Issue | 2016-03-23 | Date of Implementation | 2018-01-01 | Quoted Standard | GB/T 6682-2008; YY/T 0149-2006; YY 0967.1-2015; YY/T 0873.1-2013; YY/T 0873.3-2014; YY/T 0874-2013 | Adopted Standard | ISO 3823-2-2003, MOD | Regulation (derived from) | Notice of the General Administration of Food and Drug Administration (No. 74 of 2016) | Issuing agency(ies) | State Food and Drug Administration |
YY 0302.2-2016: Dentistry - Rotary bur instruments - Part 2: Finishing burs---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Dentistry - Rotary bur instruments - Part 2. Finishing burs
ICS 11.060.20
C33
People's Republic of China Pharmaceutical Industry Standard
Dental Rotary Gear
Part 2. Needles for dressing
(ISO 3823-2..2003, MOD)
2016-03-23 release
2018-01-01 implementation
State Food and Drug Administration issued
Directory
Preface III
Introduction IV
1 Scope 1
2 normative reference document 1
Category 1
4 symbol 1
5 request 1
Test Method
Quality Control
8 mark 31
9 Packaging 31
Reference 32
Foreword
YY 0302 consists of the following two parts.
--- YY 0302.1 Dental Rotary Gear - Part 1. Steel and Carbide Needles;
YY 0302.2 Dental rotation gears - Part 2. Trimming needle.
This part is part 2 of YY 0302.
This part is drafted in accordance with the rules given in GB/T 1.1-2009.
This part of the use of re-drafting to amend the use of ISO 3823-2..2003 "dental rotary needle device Part 2. dressing with the needle"
And its amendment 1..2008. The main changes are as follows.
On the normative reference documents, this standard has made the adjustment of technical differences to meet the technical conditions of China, the adjustment of the situation focused
Reflected in Chapter 2, "Normative references", as follows.
GB/T 6682-2008 Analytical laboratory water specifications and test methods (ISO 3696. 1987, MOD)
YY/T 0149-2006 Test method for corrosion resistance of stainless steel medical equipment (ISO 13402..1995, MOD)
YY 0967.1-2015 Dental rotation tool bar Part 1. Metal rod (ISO 1797-1..1992, MOD)
YY/T 0873.1-2013 Digital coding systems for dental rotating instruments - Part 1. General characteristics (ISO 6360-1..2004,
IDT)
YY/T 0873.3-2014 Digital coding systems for dental rotating instruments - Part 3. Characteristics of needle and tool (ISO 6360-3.
2005, IDT)
YY/T 0874-2013 Test method for dental rotating instruments (ISO 8325..2004, IDT)
Please note that some of the contents of this document may involve patents. The issuer of this document does not assume responsibility for the identification of these patents.
This section is proposed by the State Food and Drug Administration.
This part of the National Dental Material and Equipment Standardization Technical Committee of Dental Equipment and Instruments Subcommittee (SAC/TC99
SC1).
This part of the drafting unit. Ningbo Xinyuan Dental Equipment Co., Ltd., the State Food and Drug Administration Guangzhou Medical Device Quality Supervision and Inspection
Test center.
This part of the main drafters. Fan Xiang, Wan Yi Yi, Wang Haijuan, Yuan Yilin, Ruan Zhi Meng, Yuan Bi Yu.
Introduction
This section is one of the criteria for dental rotating instruments.
This section contains modifications to the requirements for carbide finishing. The requirements for the needle for steel dressing remain the same.
The requirements for size and other requirements for the use of needles for car finishing and carbide finishing for car finishing as specified in this section are considered to be
It is important to ensure that these needles are interchangeably and safely used in dental clinics.
The nominal diameter of the working part listed in Tables 1 to 68 corresponds to the diameter specified in YY/T 1011-2014.
It should be noted that the YY/T 0783 standard gives a fifteen digital coding system for the division of all types of dental rotation
instrument.
Dental Rotary Gear
Part 2. Needles for dressing
1 Scope
YY 0302 this part of the 17 most commonly used shape of steel and carbide repair needle size and related requirements, the package
Including the quality control requirements and marking requirements of these devices.
2 normative reference documents
The following documents are indispensable for the application of this document. For dated references, only the dated edition applies to this article
Pieces. For undated references, the latest edition (including all modifications) applies to this document.
GB/T 6682-2008 Analytical laboratory water specifications and test methods (ISO 3696. 1987, MOD)
YY/T 0149-2006 Test method for corrosion resistance of stainless steel medical equipment (ISO 13402..1995, MOD)
YY 0967.1-2015 Dental rotation tool bar Part 1. Metal rod (ISO 1797-1..1992, MOD)
YY/T 0873.1-2013 Digital coding systems for dental rotating instruments - Part 1. General characteristics (ISO 6360-1..2004,
IDT)
YY/T 0873.3-2014 Digital coding systems for dental rotating instruments - Part 3. Characteristics of needle and tool (ISO 6360-3.
2005, IDT)
YY/T 0874-2013 Test method for dental rotating instruments (ISO 8325..2004, IDT)
3 classification
According to the working part of the material, steel and carbide repair needle is divided into the following two types.
--- 1 type. steel trim with needle;
--- 2 type. Carbide trim needle.
4 symbols
The following symbols are used in this section.
d1 working part diameter, head diameter;
d2 neck diameter;
d3 tip part of the diameter;
l1 working part length, head length;
l2 total length;
α working part angle.
5 request
5.1 Materials
5.1.1 Work section
The working part of the steel trimmed needle shall be made of steel, and the working part of the carcass trim shall be made of carbide. material
|