US$419.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 823-2014: Ion Chromatographs Status: Valid JJG 823: Evolution and historical versions
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| JJG 823-2014 | English | 419 |
Add to Cart
|
3 days [Need to translate]
|
Ion Chromatographs
| Valid |
JJG 823-2014
|
| JJG 823-1993 | English | 439 |
Add to Cart
|
4 days [Need to translate]
|
Verification Regulation for Ion Chromatograph
| Obsolete |
JJG 823-1993
|
PDF similar to JJG 823-2014
Basic data | Standard ID | JJG 823-2014 (JJG823-2014) | | Description (Translated English) | Ion Chromatographs | | Sector / Industry | Metrology & Measurement Industry Standard | | Classification of Chinese Standard | A61 | | Classification of International Standard | 17.020 | | Word Count Estimation | 18,110 | | Date of Issue | 2/14/2014 | | Date of Implementation | 8/14/2014 | | Older Standard (superseded by this standard) | JJG 823-1993 | | Regulation (derived from) | AQSIQ Announcement No. 17 of 2014 | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | | Summary | This standard specifies equipped with a conductivity detector, UV-VIS detector and an electrochemical detector ion chromatograph initial verification, verification and subsequent use of checks. |
JJG 823-2014: Ion Chromatographs---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Ion Chromatographs
State verification procedures People's Republic of China
Ion Chromatography
Issued on. 2014-02-14
2014-08-14 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Ion chromatography test procedures
IonChromatograp
hs
Replacing JJG 823-1993
Focal point. the National Technical Committee stoichiometric physical
The main drafting units. China Institute of Metrology
Participated in the drafting unit. Shandong Institute of Metrology
Shaanxi Provincial Institute of Metrology
Instrument Co., Ltd. Qingdao Puren
Han Sheng Technology Co., Ltd. Qingdao chromatography
This procedure commissioned the National Technical Committee of the stoichiometric physics responsible for the interpretation
The main drafters of this procedure.
Shinai Jie (China Institute of Metrology)
Wang Qian (China Institute of Metrology)
Drafters participate.
Xu Aihua (Shandong Province Institute of Metrology)
Gao Xu Hui (Shaanxi Provincial Institute of Metrology)
Houqian Hui (Qingdao Puren Instrument Co., Ltd.)
Wang Qun (Qingdao Sheng Han Chromatography Technology Co., Ltd.)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2 Overview (1)
3 metering performance requirements (1)
3.1 Infusion System (1)
3.2 Column Compartment (2)
3.3 detector (2)
3.4 overall performance (2)
4 General Requirements (2)
4.1 appearance of the instrument (2)
4.2 Instrument circuit (2)
5 control of measuring instruments (3)
5.1 test conditions (3)
5.2 test items and test methods (3)
5.3 Verification Results Processing (7)
5.4 test cycle (8)
Appendix A test recording format (Reference) (9)
Appendix B test certificates/test results notice (within the page) format (12)
Introduction
JJG 823-2014 "Ion chromatography" is written JJF1001-2011 "Universal metering and fixed term
Justice ", JJF1002-2010" national verification procedures to write the rules "as the basis and foundation of JJG 823-1993
"Ion Chromatography" subject to revisions.
Compared with JJG 823-1993 "Ion chromatography", in addition to editorial changes, the main technical content are as follows
Variety.
--- Increased UV-Vis detector and an electrochemical detector test specifications. (See 1)
--- The original protocol column and instrument of integrated technical indicators, technical indicators to the instrument, and from this canceled
Sub-column related technical indicators, such as the separation column and the last peak time. (See section 3.3, 3.4)
--- For the original procedure was repeated linear range and correlation coefficient Technical indicators cancel the linear range, correlation coefficient reserved
Number of technical indicators.
--- Increasing the minimum detectable amount of the instrument to convert the relationship between the sample loop. (See section 5.2.2.4)
The previous versions of the release JJG 823 case.
--- JJG 823-1993.
Ion chromatography test procedures
1 Scope
This procedure is applied with a conductivity detector, UV-Vis detector and an electrochemical detector first ion chromatography
Second test, verification and subsequent use of checks.
2 Overview
Ion chromatography (hereinafter referred to as equipment) by infusion systems, injection systems, separation systems, detection systems and several
According to the analysis device and other components of the processing system, FIG. 1 is a schematic view of its composition. Because each component ions in the sample
Differences in distribution and adsorption properties of the stationary phase and mobile phase, mobile phase has different ions into the elution
Detector, according to the order to give the signal intensity of each ion to be measured. According to the retention time of the components and response
(Peak area or peak height) for qualitative and quantitative analysis.
Figure 1 a schematic view of the composition of ion chromatography
3 metering performance requirements
3.1 Infusion Systems
3.1.1 infusion line and interface are tight, within a predetermined pressure range allows no leakage.
3.1.2 flow setpoint error (SS) and flow stability (SR) shall comply with the requirements of Table 1.
Table 1 flow setpoint error SS and SR stability requirements
Flow setpoint/(mL/min) 0.2 ~ 0.5 0.5 ~ 1.0 greater than 1.0
Number of measurements/n 3 3 3
Collection time/min 20 ~ 10 10 ~ 5 5
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 823-2014_English be delivered?Answer: Upon your order, we will start to translate JJG 823-2014_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 823-2014_English with my colleagues?Answer: Yes. The purchased PDF of JJG 823-2014_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJG 823-2014?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJG 823-2014 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|