US$299.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 551-2021: (Verification Regulations of Sulfur Dioxide Gas Detector) Status: Valid JJG 551: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJG 551-2021 | English | 299 |
Add to Cart
|
3 days [Need to translate]
|
(Verification Regulations of Sulfur Dioxide Gas Detector)
| Valid |
JJG 551-2021
|
JJG 551-2003 | English | 170 |
Add to Cart
|
0--9 seconds. Auto-delivery
|
Verification regulation of sulfur dioxide gas detectors
| Obsolete |
JJG 551-2003
|
JJG 551-1988 | English | 639 |
Add to Cart
|
5 days [Need to translate]
|
Verification Regulation of Sulfur Dioxide Analyzer
| Obsolete |
JJG 551-1988
|
PDF similar to JJG 551-2021
Standard similar to JJG 551-2021 JJG 631 JJG 635 JJG 680 JJG 499 JJG 548
Basic data Standard ID | JJG 551-2021 (JJG551-2021) | Description (Translated English) | (Verification Regulations of Sulfur Dioxide Gas Detector) | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A61 | Word Count Estimation | 15,153 | Issuing agency(ies) | State Administration for Market Regulation |
JJG 551-2021: (Verification Regulations of Sulfur Dioxide Gas Detector)---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
(Verification Regulations of Sulfur Dioxide Gas Detector)
People's Republic of China National Metrological Verification Regulations
Sulfur dioxide gas detector
Published on 2021-07-28
2022-01-28 Implementation
Released by the State Administration for Market Regulation
Sulfur dioxide gas detector verification regulations
Replacing JJG 551-2003
Focal Point. National Technical Committee of Environmental Chemical Metrology
Main drafting unit. China Institute of Testing Technology
Shanghai Institute of Metrology and Testing Technology
Participating in the drafting unit. Henan Institute of Metrology
Gansu Metrology Research Institute
This regulation is entrusted to the National Environmental Chemical Metrology Technical Committee to be responsible for the interpretation
The main drafters of this regulation.
Liu Qing (China Institute of Testing Technology)
Ye Hong (Shanghai Institute of Metrology and Testing Technology)
Zhang Yi (China Institute of Testing Technology)
Participating drafters.
Cai Jianhua (Shanghai Institute of Metrology and Testing Technology)
Li Bo (Henan Institute of Metrology)
Shi Liyu (Gansu Metrology Research Institute)
Qi Ji (China Institute of Testing Technology)
content
Introduction (Ⅱ)
1 Scope(1)
2 Overview(1)
3 Measurement performance requirements (1)
4 General technical requirements (1)
4.1 Appearance and structure(1)
4.2 Signs and identification(2)
4.3 Power-on inspection(2)
4.4 Alarm function check (2)
4.5 Insulation resistance(2)
5 Measuring Instrument Control(2)
5.1 Verification Conditions (2)
5.2 Verification items (3)
5.3 Verification method (3)
5.4 Processing of Verification Results (5)
5.5 Verification Period(5)
Appendix A Verification Original Record (6)
Appendix B Verification Certificate/Inner Page Format of Verification Result Notification (8)
Introduction
The main technical indicators of this regulation refer to GB 12358-2006 "Workplace Environmental Gas Detection and Alarm Instrument Communication
"Technical Requirements", GB/T 13971-2013 "Technical Conditions for Ultraviolet Gas Analyzers", GB/T 25923-2010
"Technical Conditions for Online Gas Analyzers", GB/T 25929-2010 "Technical Conditions for Infrared Gas Analyzers"
and other technical regulations. This regulation is the revision of JJG 551-2003 "Sulfur Dioxide Gas Detector". with JJG 551-
Compared with.2003, the main technical changes except editorial changes are as follows.
--- Modified the scope of application (see Chapter 1);
---Modified the measurement performance requirements, listed the sulfur dioxide gas analyzer and sulfur dioxide gas detection alarm respectively
The measurement performance requirements of the instrument (see Chapter 3);
--- Deleted the alarm setting error in the.2003 version of the regulations. For the sulfur dioxide gas detection alarm, add
Check the alarm function (see 4.4);
---Modified the expression of relative expanded uncertainty and inclusion factor of sulfur dioxide gas standard material, changed.2003
In the annual edition, the relative expanded uncertainty is not more than 2% (k=3), and is revised to "When calibrating the sulfur dioxide gas analyzer, nitrogen
The relative expanded uncertainty of sulfur dioxide gas standard material is not more than 1.5%, k=2;
When the alarm instrument is used, the relative expanded uncertainty of the standard substance of sulfur dioxide gas in nitrogen is not more than 2%, and k=2". "When using gas
When diluting the device, the relative expanded uncertainty of the standard gas after dilution shall meet the above requirements” (see 5.1.2.1);
--- Modify the accuracy level of the flowmeter in the.2003 version of the regulations to not be lower than level 3 to be not lower than the accuracy level
Class 4.0 (see 5.1.2.3);
---Delete the dielectric strength verification item, and delete the dielectric strength tester and the dielectric strength verification method.
expression;
--- Added the information and format of the inner page of the verification certificate/verification result notification (see Appendix B).
The previous versions of this regulation have been released as follows.
---JJG 551-2003;
---JJG 551-1988.
Sulfur dioxide gas detector verification regulations
1 Scope
This regulation is applicable to the initial verification, subsequent verification and in-use inspection of sulfur dioxide gas detectors. sulfur dioxide
Gas detectors include sulfur dioxide gas detection alarms and sulfur dioxide gas analyzers.
This regulation does not apply to sulfur dioxide gas analyzers and stationary pollution sources whose concentration is less than 1 μmol/mol
Sulfur dioxide gas analyzer for flue gas emission monitoring.
2 Overview
The sulfur dioxide gas detection and alarm instrument is mainly used to detect the concentration of sulfur dioxide gas in the workplace environment.
It is composed of detection unit, signal processing unit, display unit and alarm unit. The detection principle is mainly electrochemical method.
When the displayed value is greater than the alarm setting value, it has sound, light or vibration alarm. According to the sampling method, it can be divided into diffusion type and pump suction
According to the way of use, it can be divided into portable and fixed type; according to the way of work, it can be divided into continuous measurement and discontinuous
Measurement.
The sulfur dioxide gas analyzer is mainly used to detect the concentration of sulfur dioxide gas in the production process and workplace environment.
It is mainly composed of a detection unit, a signal processing unit and a display unit. The detection principles are mainly spectroscopy, thermal conductivity
law, etc. According to the sampling method, it can be divided into positive pressure conveying type and pump suction type; according to the working method, it can be divided into continuous measurement and discontinuous measurement.
sex measurement.
3 Measurement performance requirements
The metering performance requirements are shown in Table 1.
Table 1 Metrology performance requirements
Test item
Metrology performance requirements
Sulfur dioxide gas analyzer sulfur dioxide gas detection alarm
Indication error ±3%FS
±5%FS or ±10%
Satisfy one of them
Repeatability 1.5% 2%
Response time 90s 60s
drift
Zero drift ±1%FS ±2%FS
Range drift ±2%FS ±3%FS
4 General technical requirements
4.1 Appearance and structure
4.1.1 The instrument should not have external damage that affects its normal operation. The surface of the newly manufactured instrument should be smooth, smooth and painted.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 551-2021_English be delivered?Answer: Upon your order, we will start to translate JJG 551-2021_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 551-2021_English with my colleagues?Answer: Yes. The purchased PDF of JJG 551-2021_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJG 551-2021?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJG 551-2021 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|