Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)StatusRelated Standard
JJG 26-2010EnglishRFQ ASK Days<=3 Verification Regulation of Expanded Stem Filling Value Tester JJG 26-2010
JJG 26-2010ChineseRFQ ASK <=1-day [PDF from Chinese Authority, or Standard Committee, or Publishing House]

   

BASIC DATA
Standard ID JJG 26-2010 (JJG26-2010)
Description (Translated English) Verification Regulation of Expanded Stem Filling Value Tester
Sector / Industry Metrology & Measurement Industry Standard

JJG 26-2010
Verification Regulation of Expanded Stem Filling Value Tester
Ministry of People's Republic of China Metrology Verification Regulations
JJG (Tobacco) 26-2010
Sweeping stem filling value tester verification procedure
Released on.2010-06-21
2010-07-01 implementation
National Tobacco Monopoly Bureau released
Sweeping stem filling value tester verification procedure
ValueTester
JJG (Tobacco) 26-2010
This regulation was approved by the State Tobacco Monopoly Bureau on June 21,.2010, and since July.2010
Implemented on the 01st.
Focal Point. National Tobacco Monopoly Bureau
Main drafting unit. China Tobacco Standardization Research Center
Participated in the drafting unit. Zhengzhou Jiade Electromechanical Technology Co., Ltd.
Hubei China Tobacco Industry Co., Ltd.
This procedure is entrusted to the National Tobacco Standardization Technical Committee for interpretation.
JJG (Tobacco) 26-2010
The main drafters of this procedure.
Nursery (China Tobacco Standardization Research Center)
Zeng Bo (Zhengzhou Jiade Electromechanical Technology Co., Ltd.)
Zhang Shenghua (Hubei China Tobacco Industry Co., Ltd.)
Participate in the drafters.
Zhang Wei (China Tobacco Standardization Research Center)
Yang Rongchao (China Tobacco Standardization Research Center)
Fan Li (China Tobacco Standardization Research Center)
Zhao Hang (China Tobacco Standardization Research Center)
Huang Weidong (Zhengzhou Jiade Electromechanical Technology Co., Ltd.)
Shen Jun (Hubei China Tobacco Industry Co., Ltd.)
Wang Qiong (Hubei China Tobacco Industry Co., Ltd.)
Mi Fangfang (Hubei China Tobacco Industry Co., Ltd.)
Xia Qiudong (Hubei China Tobacco Industry Co., Ltd.)
Tao Yongfeng (Hubei China Tobacco Industry Co., Ltd.)
JJG (Tobacco) 26-2010
table of Contents
1 range (1)
2 Citations (1)
3 Overview (1)
4 Metrology performance requirements (1)
4.1 Testing the atmosphere (1)
4.2 Performance requirements (1)
5 General technical requirements (2)
5.1 Visual inspection (2)
5.2 Working status check (2)
6 measuring instrument control (2)
6.1 Verification conditions (2)
6.2 Verification project (2)
6.3 Verification method (3)
6.4 Processing of verification results (4)
6.5 verification period (4)
Appendix A Sweeping stem fill value tester verification record format (5)
JJG (Tobacco) 26-2010
Sweeping stem filling value tester verification procedure
1 Scope
This test procedure applies to the first verification of the expanded stem cut-fill value meter for new manufacturing, during use and after repair.
Subsequent verification and in-use inspection.
2 Citations
This procedure refers to the following documents.
YC/T 163-2003 Determination of filling value of cigarette expansion stem
Use of this procedure should take care to use the current valid version of the above cited documents.
3 Overview
The expanded cut stem filling value measuring instrument is a special detecting instrument for detecting the filling ability of the cigarette expanding stem, and its work
The principle is to put a certain amount of expanded stems into the measuring barrel, and measure the expansion of the stems at a certain time and a certain pressure.
The volume that is kept below. The expander stem filling value meter mainly includes a measuring barrel, a force measuring probe, a length sensor and a time
Components such as controllers. When testing the filling value of the expanded stem, take a sample of the expanded cut stem of a certain mass and place the sample
Put it in the measuring bucket and apply a given pressure to the sample in the measuring bucket with the force applying probe for a certain period of time.
Record the height of the sample in the measuring barrel to obtain the volume of the sample in the measuring barrel at this time, and divide the volume by the mass.
The filling value to the expanded stem.
4 Metrological performance requirements
4.1 Testing the atmosphere
The temperature satisfies (22±2) °C, the relative humidity satisfies (60±5)%, and the atmospheric pressure exceeds (96±10) kPa.
It should be noted.
4.2 Performance requirements
4.2.1 Measuring bucket
The inner diameter should satisfy (60.0 ± 0.1) mm.
4.2.2 Forced probe
The pressure generated by the probe on the sample should meet (9.8 ± 0.1) N, and the diameter of the probe should meet (55.0 ±
0.1) mm.
4.2.3 Pressure rate
The speed at which the probe applies pressure to the specimen shall be (19.5 ± 0.5) mm/s.
4.2.4 Length Sensor
The measuring range should be no less than 50mm, the resolution should be 0.01mm, and the maximum tolerance should be ±0.03mm.