US$759.00 · In stock Delivery: <= 6 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1923-2021: Calibration Specification for Verification Equipment of Electrical Measuring Devices Status: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1923-2021 | English | 759 |
Add to Cart
|
6 days [Need to translate]
|
Calibration Specification for Verification Equipment of Electrical Measuring Devices
| Valid |
JJF 1923-2021
|
PDF similar to JJF 1923-2021
Basic data Standard ID | JJF 1923-2021 (JJF1923-2021) | Description (Translated English) | Calibration Specification for Verification Equipment of Electrical Measuring Devices | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Word Count Estimation | 35,311 | Issuing agency(ies) | State Administration for Market Regulation |
JJF 1923-2021: Calibration Specification for Verification Equipment of Electrical Measuring Devices ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
(Calibration specification for calibration device of electrical measuring instrument)
National Metrology Technical Specification of the People's Republic of China
Calibration specification for calibration devices for electrical measuring instruments
Published on 2021-10-18
2022-04-18 Implementation
Released by the State Administration for Market Regulation
Calibration specification for calibration devices for electrical measuring instruments
Focal Point. National Electromagnetic Metrology Technical Committee
Main drafting unit. Henan Institute of Metrology
State Grid Hubei Electric Power Company Electric Power Research Institute
China Electric Power Research Institute Co., Ltd.
Participating in the drafting unit. Changsha Tianheng Measurement and Control Technology Co., Ltd.
Fluke Test Instruments (Shanghai) Co., Ltd.
Shenzhen Kelu Precision Instrument Co., Ltd.
This specification entrusts the National Electromagnetic Metrology Technical Committee to be responsible for the interpretation
The main drafters of this specification.
Chen Qingping (Henan Institute of Metrology)
Ning Liang (Henan Institute of Metrology)
Shen Li (State Grid Hubei Electric Power Company Electric Power Research Institute)
Zhao Sha (China Electric Power Research Institute Co., Ltd.)
Participating drafters.
Deng Zhijun (Changsha Tianheng Measurement and Control Technology Co., Ltd.)
Yang Shengli [Fluke Test Instruments (Shanghai) Co., Ltd.]
Zhou Fan (Shenzhen Kelu Precision Instrument Co., Ltd.)
content
Introduction (Ⅱ)
1 Scope(1)
2 References (1)
3 Overview(1)
4 Metrology Characteristics(1)
4.1 Accuracy class(1)
4.2 Short-term stability(2)
4.3 Load Regulation(2)
4.4 Waveform Distortion(2)
4.5 Symmetry(2)
4.6 DC ripple(2)
5 Calibration Conditions(2)
5.1 Environmental conditions (2)
5.2 Measurement standards and other equipment(2)
6 Calibration items and calibration methods (2)
6.1 Calibration items (2)
6.2 Calibration method(3)
7 Expression of calibration results (13)
8 Recalibration interval(13)
Appendix A Example of AC Voltage Calibration Uncertainty Evaluation(14)
Appendix B Calibration Raw Record Format(17)
Appendix C Calibration Certificate Inside Page Format (24)
Introduction
This specification is based on JJF 1071-2010 "Rules for Compilation of National Metrology Calibration Specifications", JJF 1059.1-2012
"Measurement Uncertainty Evaluation and Representation" was compiled.
This specification is published for the first time.
Calibration specification for calibration devices for electrical measuring instruments
1 Scope
This specification applies to the calibration of the calibration device for electrical measuring instruments of class 0.05 and below (hereinafter referred to as the calibration device).
The calibration device has single-phase or three-phase AC voltage, current, frequency, power and phase output functions, and can have direct
Current voltage, DC current output function.
In this specification, the upper limit of AC voltage is 1000V, the upper limit of AC current is 100A, and the frequency range is 40Hz~
70Hz; the upper limit of DC voltage is 1000V, and the upper limit of DC current is 100A.
2 Reference documents
This specification references the following documents.
GB/T 15637-2012 General Specification for Digital Multimeter Calibrator
For dated references, only the dated version applies to this specification; for undated references
The latest version (including all amendments) applies to this specification.
3 Overview
The calibration device is composed of a signal output unit, a standard measurement unit and a control display unit, etc. It is generally used for calibration/
Calibrate single-phase or three-phase voltmeter, ammeter, power meter and other AC and DC measuring instruments, its structure can be divided into meter source
There are two types. integrated type and table source separated type.
4 Metrology characteristics
4.1 Accuracy class
The accuracy level and maximum allowable error of the calibration device are shown in Table 1.
Table 1 Accuracy class and maximum allowable error of calibration device
Features
maximum allowable error
Level 0.05 Level 0.1 Level 0.2
AC voltage, current ±0.05% ±0.1% ±0.2%
Active power ±0.05% ±0.1% ±0.2%
Reactive power ±0.1% ±0.2% ±0.5%
Phase/(°) ±0.1 ±0.2 ±0.5
Frequency/Hz ±0.01 ±0.02 ±0.05
DC voltage, current ±0.05% ±0.1% ±0.2%
Note. The maximum allowable error of other functions except phase and frequency refers to the guide error, which is expressed as a percentage.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1923-2021_English be delivered?Answer: Upon your order, we will start to translate JJF 1923-2021_English as soon as possible, and keep you informed of the progress. The lead time is typically 4 ~ 6 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1923-2021_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1923-2021_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|