Calibration Specification for Oscilloscope Voltage Probes
| Valid |
JJF 1437-2013
|
PDF similar to JJF 1437-2013
Basic data Standard ID | JJF 1437-2013 (JJF1437-2013) | Description (Translated English) | Calibration Specification for Oscilloscope Voltage Probes | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A56 | Classification of International Standard | 17.220 | Word Count Estimation | 25,255 | Quoted Standard | JJF 1057-1998; JJF 1059.1-2012; JJF 1071-2010; GB/T 15289-1994 | Regulation (derived from) | AQSIQ Notice No. 162 of 2013 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard specifies the oscilloscope voltage probe calibration items, methods and results processing for all types of calibration frequencies below 6GHz oscilloscope voltage probe. |
JJF 1437-2013: Calibration Specification for Oscilloscope Voltage Probes---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Oscilloscope Voltage Probes
People's Republic of China National Metrology Technical Specifications
Oscilloscope voltage probe calibration specifications
Issued on. 2013-11-28
2014-02-28 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Oscilloscope voltage probe calibration specifications
Focal point. the National Radio Technical Commission metering
Main drafting unit. Guangdong Institute of Metrology
Dongguan, Guangdong Province Institute of Metrology Branch
Participated in the drafting. Agilent Technologies (China) Co., Ltd.
This specification commissioned the National Radio Technical Commission responsible for the interpretation of measurement
The main drafters of this specification.
Zhang Nan Guangdong Institute of Metrology
Chang Zhi Fang Guangdong Provincial Institute of Metrology
Ye Junjiang Dongguan, Guangdong Province Institute of Metrology Branch
Drafters participate.
Tian Zhidong Agilent Technologies (China) Co., Ltd.
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2. References (1)
3 Terms and Definitions (1)
3.1 DC voltage attenuation ratio (1)
3.2 passive probes (1)
3.3 active probe (1)
3.4 differential probe (1)
4 Overview (1)
5 Metrological characteristics (2)
5.1 DC voltage attenuation ratio (2)
5.2 Frequency Response (2)
5.3 Rise time (2)
5.4 input resistance (2)
Calibration Condition 6 (2)
6.1 environmental conditions (2)
6.2 calibration standard metering, instrumentation (2)
7 calibration items and calibration methods (3)
7.1 appearance and working properly check (3)
7.2 DC voltage attenuation ratio (4)
7.3 Frequency Response (5)
7.4 Rise time (6)
7.5 input resistance (6)
8 calibration results expression (6)
9 Recalibration interval (7)
Appendix A calibration record format (8)
Appendix B calibration results format (10)
Appendix C major projects calibration Uncertainty Evaluation Examples (11)
Appendix D Using the oscilloscope's FFT function or frequency response using vector network analyzer calibration probe (16)
Appendix E differential probe common-mode rejection ratio (CMRR) (18)
Introduction
This specification based JJF 1071-2010 "National Calibration Specifications writing rules" establishment, where measurement uncertainty
Evaluation of in accordance JJF 1059.1-2012 "Evaluation and Expression of Uncertainty in Measurement" in.
The scope of this specification means for providing a voltage probe and oscilloscope are test points between the physical and electrical connection
Connect voltage probes, current probes and does not include the logic probe.
The specification for the first time to develop.
Oscilloscope voltage probe calibration specifications
1 Scope
This specification defines the oscilloscope voltage probe calibration items, processing methods and results for all kinds of frequencies below
Oscilloscope voltage probe calibration to 6GHz.
2. References
This specification refers to the following documents.
JJF 1057-1998 digital storage oscilloscope calibration specification
JJF 1059.1-2012 Evaluation and Expression of Uncertainty in Measurement
JJF 1071-2010 National Calibration Specifications writing rules
GB/T 15289-1994 digital storage oscilloscope General technical requirements and test methods
For dated references, only the dated edition applies to this specification; undated reference documents
Member, the latest edition (including any amendments) applies to this specification.
3 Terms and Definitions
3.1 DC voltage attenuation ratio DCattenuationratio
Oscilloscope voltage probe to measure the ratio of the input voltage and output voltage. It is defined as the voltage oscilloscope probe
Within the first measurement range, the input DC voltage and output DC voltage ratio.
3.2 passive probe passiveprobe
It does not contain the active device that requires no external power supply can work properly probe.
3.3 Active Probe activeprobe
Comprises an active device, you need to probe external power to function properly. In general, the input capacitance is small, the output
The resistance is small, it can reduce interference to the circuit under test to achieve a wide measurement bandwidth. Non-differential active probes
The maximum input voltage is generally small, and the need to guard against electrostatic discharge damage and other factors active probe, and Active Differential
Probes can be used for differential measurement of high voltage.
3.4 Differential Probe differentialprobe
There is a positive and negative input terminals and two separate ground active probe. Voltage output voltage between the two inputs
Proportional to the difference. The reference signal is used as a reference instead of the mutual observation ground level, and observed a significant dc
Bias or other common-mode component of the signal, such as a small power supply noise superimposed signal.
4 Overview
Oscilloscope probe voltage is measured using an oscilloscope important equipment can be different forms of transmission line
According to certain rules to convert the voltage signal to a voltage signal coaxially provided oscilloscope measurements. According to whether to include active devices
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1437-2013_English be delivered?Answer: Upon your order, we will start to translate JJF 1437-2013_English as soon as possible, and keep you informed of the progress. The lead time is typically 3 ~ 5 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1437-2013_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1437-2013_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|