US$459.00 · In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1217-2009: Calibratior Specification for Electrosurgical Generator Status: Valid JJF 1217: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1217-2009 | English | 459 |
Add to Cart
|
4 days [Need to translate]
|
Calibratior Specification for Electrosurgical Generator
| Valid |
JJF 1217-2009
|
JJF 1217-1990 | English | 279 |
Add to Cart
|
3 days [Need to translate]
|
Operating technical norm of temperature secondary standard in the range 90. 188 to 273. 15 K
| Obsolete |
JJF 1217-1990
|
PDF similar to JJF 1217-2009
Basic data Standard ID | JJF 1217-2009 (JJF1217-2009) | Description (Translated English) | Calibratior Specification for Electrosurgical Generator | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Classification of International Standard | 17.220 | Word Count Estimation | 20,287 | Date of Issue | 2009-05-18 | Date of Implementation | 2009-08-18 | Quoted Standard | JJF 1001-1998; JJF 1071-2000; JJF 1059-1999; GB 9706.1-2007; GB 9706.4-1999 | Regulation (derived from) | ?AQSIQ "on release JJG492-2009" cesium atomic frequency standard test procedures "and other five national metrology technical regulations Bulletin" 2009 No.54; | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to newly manufactured, use and repair, the work is quite a range 0. 3MHz ~ 5. OMHz rate unipolar, bipolar electrosurgical medical calibration is quite high. This specification does not apply to a unipolar mode the maximum power is less than 50W of John the calibration of medical high-frequency electric knife, does not apply to dental bipolar mode, gynecology and dermatology, and other special high-frequency electric knife calibration. |
JJF 1217-2009: Calibratior Specification for Electrosurgical Generator---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibratior Specification for Electrosurgical Generator
People's Republic of China National Metrology Technical Specifications
High-frequency electric knife calibration specification
Posted 2009-05-18
2009-08-18 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
High-frequency electric knife calibration specification
forElectrosurg
icalGenerator
The calibration specifications by the State Administration of Quality Supervision, Inspection and Quarantine on May 18, 2009 approved,
And it shall become effective August 18, 2009.
Focal point. the National Technical Committee on Electromagnetic measurement
Drafted by. Hangzhou quality technical supervision detection hospital
China Institute of Metrology
China Testing Technology Research Institute
Army Medical electromagnetics Metrology Research Station
Shanghai Medical Device detection
The provisions made by the responsible National Technical Committee Electromagnetic measurement interpretation
The main drafters of this specification.
Jiang Xueping Hangzhou quality technical supervision detection hospital
Shao Haiming China Institute of Metrology
Shao Lei Hangzhou quality technical supervision detection hospital
Drafters participate.
Yong China Testing Technology Research Institute
Jia Jiange army medical Electromagnetic Metrology Research Station
He Jun Shanghai Medical Device detection
table of Contents
1 Scope (1)
2. Referenced Documents (1)
3 Terms and Definitions (1)
4 Overview (2)
5 Metrological characteristics (2)
5.1 high-frequency leakage current (2)
5.2 output power setting error (2)
5.3 Maximum output power (2)
5.4 enclosure leakage current (2)
Calibration Condition 6 (2)
6.1 environmental conditions (2)
6.2 standards and other measuring equipment (2)
7 calibration items and calibration methods (3)
7.1 appearance and work of checking (3)
7.2 high-frequency leakage current calibration (3)
7.3 output power set value calibration error (5)
7.4 Calibration The maximum output power (6)
7.5 enclosure leakage current measurement (6)
8 calibration results processing and calibration cycle (7)
8.1 Calibration records (7)
8.2 Calibration process results (7)
9 Recalibration interval (7)
Appendix A high-frequency electric knife calibration original records (8)
Appendix B Calibration Certificate page format (12)
High-frequency electric knife calibration specification
1 Scope
This standard applies to newly manufactured after use and repair work in a single frequency range of 0.3MHz ~ 5.0MHz
Pole, bipolar electrosurgical medical calibration. This specification does not apply to a unipolar mode of operation is less than the maximum output power
Medical high-frequency electric knife calibration 50W, and does not apply to bipolar work mode dental, gynecology and dermatology professionals
A high frequency electric knife calibration.
2 Citations
JJF 1001-1998 common measurement terms and definitions
JJF 1071-2000 National Calibration Specifications writing rules
JJF 1059-1999 Evaluation and Expression of Uncertainty in Measurement
GB 9706.1-2007 Medical electrical equipment Part 1. General requirements for safety
GB 9706.4-1999 Medical electrical equipment Part 2. requirements for the safety of high frequency surgical equipment
When using this specification it should be noted that using the currently valid version of the documents cited.
3 Terms and Definitions
3.1 surgery electrode activeelectrode
A method for generating a physical effect in some electrosurgery required, such as cutting, coagulation electrode.
3.2 bipolar electrode bipolarelectrode
There are two components surgery electrodes on the same bracket, when stimulated, mainly in high-frequency current between the two electrodes
flow.
3.3 neutral electrode neutralelectrode
Also known as patient plate, apply muscle or dispersion plate electrode, the area of the patient's body connected to a relatively large electrodes to mention
For low current density of the high frequency current loop, avoid physical effects such as burns like harmful in human tissues.
3.4 rated output power ratedoutputpower
It can be input to the non-inductive resistance of 50Ω ~ 2000Ω between unipolar output circuit and a bipolar output between 10Ω ~ 1000Ω
The maximum power of the high frequency circuit.
3.5 enclosure leakage current enclosureleakagecurrent
From the operator during normal use or patient accessible housing or casing members (except Application Part), via an external
Conductive connection rather than protective conductor current into the ground or other parts of the housing.
3.6 high-frequency leakage current high-frequencyleakagecurrent
Electrosurgical two output electrodes to the current non-functional, no effect on the operation and can cause burns and patients
Environmental pollution.
3.7 Cutting cutting
With a high current density of the high frequency current through it to cut the biological tissue surgery electrode.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1217-2009_English be delivered?Answer: Upon your order, we will start to translate JJF 1217-2009_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1217-2009_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1217-2009_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJF 1217-2009?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJF 1217-2009 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|