HOME   Cart(1)   Quotation   About-Us Tax PDFs Standard-List
Powered by Google-Search & Google-Books www.ChineseStandard.net Database: 189760 (2 Jul 2022)

JJF 1190-2008

JJF 1190-2008_English: PDF (JJF1190-2008)
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)Related StandardStatusPDF
JJF 1190-2008English459 Add to Cart 4 days Calibration specification for airborne particle counter JJF 1190-2008 Valid JJF 1190-2008

BASIC DATA
Standard ID JJF 1190-2008 (JJF1190-2008)
Description (Translated English) Calibration specification for airborne particle counter
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A61
Classification of International Standard 17.180.30
Word Count Estimation 20,283
Date of Issue 2008-03-24
Date of Implementation 2008-09-24
Older Standard (superseded by this standard) JJG 547-1988
Drafting Organization Shanghai Measurement and Testing Technology
Administrative Organization National Environmental stoichiometric Technical Committee
Regulation (derived from) AQSIQ Announcement No. 35 of 2008
Issuing agency(ies) State Administration of Quality Supervision, Inspection and Quarantine
Summary This standard specifies the communication light attenuator common metrological characteristics, calibration methods and equipment used in the calibration requirements, etc., for the first time the communication light attenuator calibration, calibration and use in subsequent testing, optical attenuator type evaluation requirements relating to metrological characteristics can be performed with reference to the specification, the other with light attenuation function instrument calibration can also be implemented by reference,

JJF 1190-2008
Calibration specification for airborne particle counter
People's Republic of China National Metrology Technical Specifications
Dust particle counter calibration specification
Posted 2008-03-24
2008-09-24 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Dust particle counter calibration specification
Instead of JJG547-1988
This specification by the State Administration of Quality Supervision, Inspection and Quarantine approved on March 24, 2008, and from
From September 24, 2008 into effect.
Focal point. the National Technical Committee on Environment stoichiometry
Main drafting unit. Shanghai Institute of Measurement and Testing
Participated in the drafting. Suzhou City, a hundred God Technology Co., Ltd.
This specification technical committee commissioned by the National stoichiometric environment responsible for interpreting
The main drafters of this specification.
Yu Peiying (Shanghai Institute of Measurement and Testing)
Ding Zhen Min (Shanghai Institute of Measurement and Testing)
Liu Yue (Shanghai Institute of Measurement and Testing)
Drafters participate.
Hu Syracuse (Suzhou City, a hundred God Technology Co., Ltd.)
table of Contents
1 Scope (1)
2 Terms (1)
3 Overview (1)
4 Metrological characteristics (1)
4.1 appearance requirements (1)
4.2 Insulation resistance (1)
4.3 Electric strength (1)
4.4 net time (2)
4.5 Flow error (2)
Timing error 4.6 (2)
4.7 Repeatability (2)
4.8 Error particle size distribution (2)
4.9 particle concentration indication error (2)
Calibration Condition 5 (2)
5.1 Calibration Environment (2)
5.2 Calibration with major equipment (2)
6 calibration items and calibration methods (2)
6.1 appearance requirements (2)
6.2 Insulation resistance (3)
6.3 Electric strength (3)
6.4 net time (3)
6.5 Flow error (3)
6.6 timing error (3)
6.7 Repeatability (3)
6.8 Error particle size distribution (4)
6.9 particle concentration indication error (5)
Expression 7 Calibration results (5)
8 recalibration intervals (6)
Appendix A monodisperse latex particle dilution and storage methods (7)
Appendix B dust particles causing measuring means (8)
Appendix C monodisperse particles of the aerosol concentration adjustment (9)
Appendix D particle concentration indication error of measurement uncertainty evaluation (10)
Appendix E of the particle size distribution measurement error Uncertainty Evaluation (12)
Appendix F original record format (14)
Appendix G calibration results within the page format (15)
Dust particle counter calibration specification
1 Scope
This standard applies to laser light scattering particle counter (hereinafter referred to as particle counter) calibration.
2 Terms
2.1 monodisperse particles (monodisperseparticle)
A group of polystyrene spherical particles, the particle diameter distribution of relative standard deviation ≤5%, see Appendix A.
2.2 diameter (particlesize)
Certain air of discrete particle diameter (short diameter), with the intensity of the scattered light corresponding to a monodisperse particle
The diameter of the sub-unit μm.
Particle diameter in the measured particle counter (0.1 to 10) μm range.
2.3 particle concentration (particleconcentration)
Within a predetermined sampling time and sampling flow, measured in the presence of air per unit volume of from less than a specified diameter
The number of scattered particles.
2.4 size distribution error (errorofparticlesize)
Particle counter for different size particles in aerosols percentage measurement error measurements.
3 Overview
Particle counters measure the particle size and particle concentration is clean air in discrete particles, to determine the clean room and clean
Cleanliness level of the instrument environment. Dust particle counter can be used for measuring the dust content, as well as clean bench and
Filter performance testing.
Particle Counter mainly by the gas system, an optical system, a data processing system components. It wor