JJF 1173-2007_English: PDF (JJF1173-2007)
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Related Standard | Status | PDF |
JJF 1173-2007 | English | 839 |
Add to Cart
|
5 days
|
Calibration Specification of Measuring Receivers
|
JJF 1173-2007
| Obsolete |
JJF 1173-2007
|
Standard ID | JJF 1173-2007 (JJF1173-2007) | Description (Translated English) | Calibration Specification of Measuring Receivers | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A56 | Classification of International Standard | 17.220.01 | Word Count Estimation | 28,224 | Date of Issue | 2007-02-28 | Date of Implementation | 2007-05-28 | Quoted Standard | AQSIQ Announcement No. 41 of 2007 | Drafting Organization | China Aerospace Science and Industry Corporation Second Hospital 203 | Administrative Organization | National Radio Measurement Technical Committee | Regulation (derived from) | AQSIQ Announcement No. 41 of 2007 | Summary | This standard applies to new manufacturing, new acquisition, use and repair after 150kHz ~ 26. 5GHz frequency range of the calibration of the measuring receiver. |
JJF 1173-2007
Calibration Specification of Measuring Receivers
People's Republic of China National Metrology Technical Specifications
Calibration measuring receiver specification
Posted 2007-02-28
2007-05-28 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Calibration measuring receiver specification
Measuring
Receivers
This specification by the State Administration of Quality Supervision, Inspection and Quarantine, 2007 February 28 ratification and
From May 28, 2007 into effect.
Focal point. the National Radio Technical Commission metering
The main drafting units. China Aerospace Science and Industry Corporation II Hospital 2 ○ San Suo
Participated in the drafting unit. China Institute of Metrology
The specification of the National Radio Technical Commission responsible for the interpretation of measurement
The main drafters of this specification.
Jiayao Zong (China Aerospace Science and Industry Corporation II Hospital 2 ○ San Institute)
Wu Weiping (China Aerospace Science and Industry Corporation II Hospital 2 ○ San Institute)
Levin Italian (China Aerospace Science and Industry Corporation II Hospital 2 ○ San Institute)
Drafters participate.
Bian Xin (China Institute of Metrology)
table of Contents
1 Scope (1)
2. Referenced Documents (1)
3 Overview (1)
4 Metrological characteristics (1)
4.1 Frequency (1)
4.2 power (1)
4.3 tuning level (1)
4.4 FM (1)
PM 4.5 (2)
AM 4.6 (2)
Calibration Condition 5 (2)
5.1 environmental conditions (2)
5.2 Calibration Equipment (2)
6 calibration items and calibration methods (4)
6.1 normal check (4)
6.2 Power Calibration (4)
6.3 Calibration frequency range (5)
6.4 tuning level calibration (5)
6.5 Calibration FM (6)
6.6 Calibration phase modulation (10)
Calibration AM 6.7 (12)
7 Calibration Results The expression (14)
8 Recalibration interval (14)
Appendix A calibration record format (15)
Appendix B Measurement Uncertainty Evaluation (20)
Calibration measuring receiver specification
1 Scope
This standard applies to newly manufactured, 150kHz ~ 26.5GHz frequency range measured new purchase, use, and after repair
Calibration quantity receiver.
2 Citations
JJF 1071-2000 "National Calibration Specifications writing rules"
JJF 1059-1999 "Evaluation and Expression of Uncertainty in Measurement"
3 Overview
Measuring receiver is used to measure faint signal instruments, based on different uses, it can access a variety of types of measurements
Receiver. Broadly speaking, spectrum analyzers, modulation analyzers, Field intensity meter, interference meter, time domain reflectometry
Meter, network analyzers, and so are all the measuring receiver category. This specification relates to measuring receiver, mainly used in the letter
Generator and attenuator calibration measurements.
Measuring receiver with frequency, power, tuning levels (attenuation), amplitude modulation, frequency offset, phase offset, frequency modulation
And demodulation distortion measurement functions and other technical parameters.
This specification above metrological characteristics (specifications) for calibration, measurement and its uncertainty are given.
Currently in use have a measuring receiver HP8902A, HP8902S, N5530S and FSMR models. different
Model Integration not the case, the frequency range and performance indicators there are differences, but the measurement principle and a basic measurement method
Cause, therefore, in order to calibrate the measurement operability, this specification will be compiled based HP8902A, other models
Calibration of the measuring receiver can refer to the relevant provisions.
4 Metrological characteristics
4.1 Frequency
Frequency range. 150kHz ~ 26.5GHz;
Frequency accuracy. 1 × 10-6.
4.2 Power
Power range. 30dBm ~ -20dBm;
The maximum permissible error. ± (0.01 ~ 0.02) dB.
4.3 tuning level
Level range. 0dBm ~ -127dBm;
The maximum permissible error. ± (0.01 ~ 0.5) dB.
4.4 FM
Carrier frequency range. 15
|