JJF 1149-2006English759 Add to Cart 5 days Calibration Specification for Cardiac Defibrillators and Cardiac Defibrillators-monitors JJF 1149-2006 Obsolete JJF 1149-2006

BASIC DATA
Standard ID JJF 1149-2006 (JJF1149-2006)
Description (Translated English) Calibration Specification for Cardiac Defibrillators and Cardiac Defibrillators-monitors
Sector / Industry Metrology & Measurement Industry Standard
Classification of Chinese Standard A56
Classification of International Standard 17.220
Word Count Estimation 19,125
Date of Issue 2006-05-23
Date of Implementation 2006-08-23
Quoted Standard GB 9706.8; JJG 760
Drafting Organization Army medical Electromagnetics Research Station Measurement and Testing
Administrative Organization National Radio Measurement Technical Committee
Regulation (derived from) ?AQSIQ Announcement 2014 No.89
Summary This standard applies to newly manufactured, use and repair common after cardiac defibrillators (hereinafter referred to as defibrillators) and cardiac defibrillator monitors (hereinafter referred to as the defibrillator monitors) defibrillation part (including defibrillation affect some part of the ECG) calibration, does not apply to automatic external defibrillator, vivo implantable defibrillator calibration. Defibrillator Monitor ECG part of metrological characteristics and calibration reference JJG 760-2003 ECG test procedures appropriate terms.

JJF 1149-2006
Calibration Specification for Cardiac Defibrillators and Cardiac Defibrillators-monitors
National Metrology Technical Specification of the People's Republic
Cardiac defibrillator and cardiac defibrillator
Calibration specification
Released on.2006-05-23
Implementation of.2006-08-23
The General Administration of Quality Supervision, Inspection and Quarantine issued
Cardiac defibrillator and defibrillation
Monitor calibration specification
Defibrilators
Monitors
This specification was approved by the General Administration of Quality Supervision, Inspection and Quarantine on May 23,.2006, and
Implemented on August 23,.2006.
Focal Point. National Radio Measurement Technology Committee
The main drafting unit. the military medical electromagnetic measurement test research station
China Institute of Metrology
Participated in the drafting unit. Hangzhou Quality and Technology Supervision and Inspection Institute
This specification is interpreted by the National Radiometric Technical Committee.
The main drafters of this specification.
Li Yuxue (the military medical electromagnetic metrology test research station)
Jia Jiange (the military medical electromagnetic measurement test research station)
卞昕 (China Institute of Metrology)
Participate in the drafters.
Jiang Xueping (Hangzhou Quality and Technical Supervision and Inspection Institute)
table of Contents
1 range (1)
2 Citations (1)
3 Terms and definitions (1)
3.1 Defibrillator monitor (1)
3.2 Release energy (1)
3.3 Synchronous mode (1)
3.4 Delay time (1)
3.5 Internal discharge (1)
4 Overview (1)
5 Measurement characteristics (2)
5.1 Release energy (2)
5.2 Charging time (2)
5.3 Number of charge and discharge (2)
5.4 Energy loss rate (2)
5.5 Internal discharge (2)
5.6 Synchronous mode (2)
5.7 Recovery of ECG monitor after defibrillation (2)
5.8 Anti-interference ability of ECG monitor for charging or internal discharge (2)
6 Calibration conditions (2)
6.1 Environmental conditions (2)
6.2 Measurement standards and other equipment (3)
7 Calibration items and calibration methods (3)
7.1 Appearance and normality check (3)
7.2 Release energy (3)
7.3 Charging time (4)
7.4 Number of charge and discharge (4)
7.5 Energy loss rate (4)
7.6 Internal discharge (5)
7.7 Synchronous mode (5)
7.8 Recovery of ECG monitor after defibrillation (5)
7.9 Anti-interference ability of ECG monitor for charging or internal discharge (6)
8 Calibration result expression (8)
8.1 Calibration Record (8)
8.2 Processing of calibration results (8)
9 re-study interval (8)
Appendix A Assessment of Uncertainty in Energy Release (9)
Appendix B Calibration Record Format (10)
Appendix C Calibration Certificate Page Format (12)
Cardiac Defibrillator and Cardiac Defibrillator Monitor Calibration Specifications
1 Scope
This specification applies to new cardiac defibrillators (hereinafter referred to as defibrillators) and hearts that are newly manufactured, used, and repaired.
Defibrillation monitor (hereinafter referred to as defibrillation monitor) defibrillation part (including defibrillation part of the defibrillation part of the ECG monitoring department)
Calibration of the influence of the fraction; does not apply to the calibration of an automatic external defibrillator or an in vivo implantable defibrillator. Defibrillator monitor
The measurement characteristics and calibration of the ECG monitoring part can refer to the corresponding provisions of the JJG760-2003 ECG monitor verification procedure.
get on.
2 Citations
GB 9706.8-1995 Dedicated safety requirements for cardiac defibrillators and cardiac defibrillators
JJG760-2003 ECG monitor verification procedures
Use of this specification should be done with the current valid version of the above cited documents.
3 Terms and definitions
The following terms and definitions established in GB 9706.8-1995 apply to this specification.
3.1 defibrillation monitor cardiacdefibrilator-monitor
Refers to a defibrillator