US$149.00 · In stock Delivery: <= 2 days. True-PDF full-copy in English will be manually translated and delivered via email. GB/T 15893.1-2014: Industrial circulating cooling water -- Determination of turbidity -- Diffused radiation method Status: Valid GB/T 15893.1: Evolution and historical versions
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| GB/T 15893.1-2014 | English | 149 |
Add to Cart
|
2 days [Need to translate]
|
Industrial circulating cooling water -- Determination of turbidity -- Diffused radiation method
| Valid |
GB/T 15893.1-2014
|
| GB/T 15893.1-1995 | English | 199 |
Add to Cart
|
2 days [Need to translate]
|
Industrial circulating cooling water--Determination of turbidity--Diffused radiation method
| Obsolete |
GB/T 15893.1-1995
|
PDF similar to GB/T 15893.1-2014
Basic data | Standard ID | GB/T 15893.1-2014 (GB/T15893.1-2014) | | Description (Translated English) | Industrial circulating cooling water -- Determination of turbidity -- Diffused radiation method | | Sector / Industry | National Standard (Recommended) | | Classification of Chinese Standard | G76 | | Classification of International Standard | 13.060.60 | | Word Count Estimation | 5,584 | | Date of Issue | 7/8/2014 | | Date of Implementation | 12/1/2014 | | Older Standard (superseded by this standard) | GB/T 15893.1-1995 | | Quoted Standard | GB/T 6682-2008 | | Adopted Standard | ISO 7027-1999, NEQ | | Regulation (derived from) | 2014 National Standards Bulletin No. 18 | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China | | Summary | This standard specifies the method for measuring scattered light industrial circulating cooling water turbidity. This standard applies to industrial circulating cooling water turbidity measurement, measurement range 0 NTU ~ 50 NTU. |
GB/T 15893.1-2014: Industrial circulating cooling water -- Determination of turbidity -- Diffused radiation method ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Industrial circulating cooling water. Determination of turbidity. Diffused radiation method
ICS 13.060.60
G76
National Standards of People's Republic of China
Replacing GB/T 15893.1-1995
Industrial circulating cooling water turbidity
Light scattering method
(ISO 7027.1999, Waterquality-Determinationofturbidity, NEQ)
Issued on. 2014-07-08
2014-12-01 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China released
Foreword
This standard was drafted in accordance with the rules given in GB 1.1-2009.
This standard replaces GB/T 15893.1-1995 "Industrial circulating cooling water turbidity measuring scattered light method" and GB/T 15893.1-
Compared to 1995, the main technical changes are as follows.
--- Turbidity units FNU replaced by the NTU;
--- Increased formazin turbidity standard solution preparation list (see 4.3);
--- Modify the scattered light turbidity meter light source incident light angle and measure the receiver (see 5.1, version 5.1 in 1995).
The standard reference method using redrafted ISO 7027.1999 "Determination of water turbidity of the" establishment, and ISO 7027.1999's
Non-equivalent.
The standard proposed by China Petroleum and Chemical Industry Association.
This standard by the National Standardization Technical Committee chemical water treatment chemicals Technical Committee (SAC/TC63/SC5) centralized.
This standard is drafted by. Hao bright Chemical Research and Design Institute Co., Ltd. Hangzhou-China environmental protection equipment factory in Guangzhou City special pressure
Equipment Inspection Institute, Jiashan green environmentally friendly materials factory, Xiamen City, measuring Detection Technology Co., Ltd., China Petroleum & Chemical Corporation North
North of Beijing Yanshan Branch hospital, CNOOC Tianjin Chemical Research & Design Institute, Chongqing University.
The main drafters of this standard. Guoxi Min, Abolished Guotai Xi Yu Fen, Yu Minghua, Wan Feng, Wang Dong, Shao Hongqian, ZHENG Huai-li, Sun Tingting.
This standard replaces the standards previously issued as follows.
--- GB/T 15893.1-1995.
Industrial circulating cooling water turbidity
Light scattering method
1 Scope
This standard specifies the method for measuring scattered light industrial circulating cooling water turbidity.
This standard applies to the determination of industrial circulating cooling water turbidity measurement range 0NTU ~ 50NTU.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
Laboratory use specifications and test methods GB/T 6682-2008 Analysis
3 Method summary
This standard formazin polymer (Formazinpolymer) as turbidity standard solution, with the principle of scattered light turbidity meter measuring water
Sample turbidity.
4 Reagents and materials
4.1 standard reagents and water, unless otherwise specified, shall use analytical reagent and in line with GB/T 6682-2008 in two water
Provisions.
4.2 formazin turbidity standard stock solution. 400NTU. Weigh 10.00g ± 0.01g hexamethylenetetramine, dissolved in water and diluted to
100mL, this is the A solution; weighing 1.000g ± 0.001g of hydrazine sulfate, dissolved in water and diluted to 100mL, this is a solution B; pipette
A solution 5mL and 5mL solution B, mixing, placing 24h at 25 ℃ ± 3 ℃, then diluted with water to 100mL, that is formalin
Hydrazine turbidity standard stock solution. The solution at 25 ℃ ± 3 ℃ in the shade storage, stable 30d.
4.3 formazin turbidity standard solution. Table 1 using a pipette to take a certain volume of formazin turbidity standard stock solution was diluted with water to
100mL, to formulate the required turbidity formazin turbidity standard solution, stable period of 7d.
Table 1 formazin turbidity standard solution preparation (0NTU ~ 50NTU)
Take turbidity standard volume of stock solution/mL 0 1.25 2.50 3.75 5.00 7.50 10.00 12.50
Turbidity standard solution turbidity/NTU 0 5 10 15 20 30 40 50
5 instruments and equipment
5.1 scattered light turbidity meter
5.1.1 source. equipped with a monochromator tungsten, or monochromatic light emitting diodes, or monochrome lasers, visible light can also be used.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB/T 15893.1-2014_English be delivered?Answer: Upon your order, we will start to translate GB/T 15893.1-2014_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 2 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB/T 15893.1-2014_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 15893.1-2014_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version GB/T 15893.1-2014?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB/T 15893.1-2014 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|