Test methods for unsaturated polyester resins
| Valid |
GB/T 7193.1-1987 | English | 199 |
Add to Cart
|
2 days
|
Unsaturated polyester resin determination of viscosity
| Obsolete |
Excerpted PDFs (Download full copy in 9 seconds upon purchase)PDF Preview: GB/T 7193-2008
GB/T 7193-2008: Test methods for unsaturated polyester resins---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/GBT7193-2008
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 83.120
Q 23
Replacing GB/T 7193.1~7193.6-1987, partly replacing GB/T 8238-1987
Test methods for unsaturated polyester resins
(ISO 2554.1997, ISO 584.1982, NEQ)
Issued on: JUNE 30, 2008
Implemented on: APRIL 01, 2009
Issued by. General Administration of Quality Supervision, Inspection and
Quarantine;
Standardization Administration of the People’s Republic of China.
Table of Contents
Foreword... 3
1 Scope... 6
2 Normative references... 6
3 Terms and definitions... 6
4 Test methods... 7
5 Test report... 19
Appendix A (Informative) Drum or rotor matched speed... 20
Test methods for unsaturated polyester resins
1 Scope
This Standard specifies the test methods, scope, principles, test samples, instruments
and equipment, test procedures, test results and test reports for testing the properties of
liquid unsaturated polyester resins.
This Standard is applicable to the determination of absolute viscosity, hydroxyl value,
solid content, reactivity at 80 °C, heat stability at 80 °C, gel time at 25 °C and refractive
index of liquid unsaturated polyester resins.
2 Normative references
The following referenced documents are indispensable for the application of this
document. For dated references, all subsequent amendments (not including errata
content) or revisions do not apply to this standard. However, parties to agreements that
are based on this Standard are encouraged to study whether the latest versions of these
documents can be used. For undated references, the latest edition applies to this
Standard.
GB/T 2895, Determination for acid value of unsaturated polyester resins
GB/T 6682, Water for analytical laboratory use - Specification and test methods
3 Terms and definitions
The following terms and definitions are applicable to this Standard.
3.1
hydroxyl value
The number of milligrams of potassium hydroxide consumed to neutralize the acetic
acid produced by the acetylation reaction of 1 g of unsaturated polyester resins.
3.2
acid value
The number of milligrams of potassium hydroxide consumed to neutralize 1 g of test
sample under the test conditions.
3.3
total acid value
The number of milligrams of potassium hydroxide consumed to neutralize all carboxyl
groups, free acids and free anhydrides in the polyester.
3.4
solid content
The mass fraction of non-volatiles contained in unsaturated polyester resins under
specific test conditions.
3.5
heat stability at 80 °C
The time required for the liquid unsaturated polyester resins, at a temperature of 80 °C,
from the start of the test to the appearance of gelation.
3.6
gel time
The time FROM the addition of the initiator to the resin TO the resin viscosity reaches
50 Pa·s.
4 Test methods
4.1 Viscosity
4.1.1 Principle
The drum or rotor rotates in the sample at a fixed speed. Due to the viscosity of the
liquid, resistance is applied to the drum or rotor during rotation, generating torque,
which is measured by a certain method.
The measurement process is achieved by the change of the digital pointer caused by the
compression of the coil spring. The absolute viscosity is measured by the rotational
viscometer by multiplying the reading by a coefficient that depends on the speed and
the type of drum or rotor.
4.1.2 Test samples
4.1.2.1 Uniform, free of bubbles and impurities.
4.1.2.2 The quantity should be sufficient for viscometer measurement.
4.1.3 Instruments and apparatuses
4.1.3.1 Rotational viscometer. drum type or rotor type.
4.1.3.2 Constant temperature water bath. temperature control accuracy ±0.5 °C.
4.1.3.3 Thermometer. measuring range 0 ℃ ~ 50 ℃, minimum graduation value 0.2 ℃.
4.1.3.4 Container. shall meet the requirements of the viscometer.
4.1.3.5 Stopwatch.
4.1.4 Test procedures
4.1.4.1 Select the drum or rotor and speed of the rotational viscometer (see Appendix
A) so that the measured reading falls between 20% and 90%, preferably between 45%
and 90%, of the full scale value.
4.1.4.2 Place the test sample in a container; adjust the temperature to approximately
25 °C; then place the container in a constant temperature water bath at 25 °C ± 0.5 °C
(or pour the sample into the measuring container of the viscometer), where the water
bath surface shall be slightly higher than the sample surface.
4.1.4.3 Immerse the viscometer drum or rotor vertically into the center of the test
sample. The immersion depth shall comply with the requirements of the viscometer.
Start timing at the same time.
4.1.4.4 During the entire measurement process, the test sample temperature shall be
controlled at 25 ℃ ± 0.5 ℃. When the drum or rotor has been immersed in the test
sample for 8 minutes, start the motor and read the reading after the drum or rotor rotates
for 2 minutes. After reading the value, turn off the motor; wait for 1 minute, then turn
on the motor; rotate for 1 minute and take the second reading.
4.1.4.5 Empty the container and repeat 4.1.4.2 ~ 4.1.4.4.
4.1.4.6 After measuring each test sample, the viscometer drum or rotor shall be cleaned
using the solvent.
4.1.5 Test results
4.1.5.1 Each sample shall be measured twice. The readings shall be calculated
according to the viscometer regulations and expressed as the arithmetic mean with three
significant figures.
4.1.5.2 The measurement result is in Pa·s.
4.2 Hydroxyl value
4.2.1 Principle
4.2.3.4 Water bath, temperature controlled at 50 ℃ ± 1 ℃.
4.2.3.5 Pipettes, 5 mL and 10 mL (for acetylation solution).
4.2.3.6 Potentiometric titrator, equipped with a calomel electrode/glass electrode
system and a titration stand.
4.2.3.7 Analytical balance, sensitivity 1 mg.
4.2.4 Procedures
4.2.4.1 Weigh approximately 5 mg of test sample containing equivalent hydroxyl
groups (1 g of sample = 280 equal parts/hydroxyl value) into a 250 mL conical flask,
accurate to 1 mg. If the approximate value of the hydroxyl value is unknown, perform
preliminary tests first.
4.2.4.2 Accurately add 10 mL of acetylation solution; place a magnetic stirring bar; plug
the bottle; use ethyl acetate to moisten the bottle mouth; use a magnetic stirrer to stir
the test sample to dissolve it. If the test sample cannot be completely dissolved after
heating, add another 5 mL or 10 mL of acetylation solution.
4.2.4.3 Place the conical flask in a water bath at 50 ℃ ± 1 ℃; immerse to a depth of
about 10 mm; keep it there for 45 minutes. (This time can be shortened, for example,
to 30 minutes or less, as long as the test results are the same.)
4.2.4.4 Remove the conical flask; cool it to room temperature; add 2 mL of distilled
water; use a magnetic stirrer to stir. When the solution is fully mixed, add 10 mL of
pyridine/water mixture and stir for 5 min.
4.2.4.5 Use 60 mL of n-butanol/toluene mixture to rinse the stopper and the inner wall
of the conical flask; add 5 drops of mixed indicator.
4.2.4.6 Use potassium hydroxide-methanol standard solution to titrate while stirring
continuously. When the color changes, add (1~2) drops of mixed indicator. The solution
changes from yellow to clear; record the milliliters of potassium hydroxide-methanol
standard solution consumed (V1); add one more drop of potassium hydroxide solution;
the solution color turns blue; if it does not turn blue, record the burette reading and add
one more drop of mixed indicator until blue appears.
4.2.4.7 The V1 value used in the calculation of the result is the volume of the potassium
hydroxide-methanol standard solution consumed before the drop that turns the solution
blue.
4.2.4.8 Carry out a blank test under the same conditions and record the amount of
potassium hydroxide-methanol standard solution consumed in milliliters (V0).
The test sample shall be weighed from not less than 100 mL of sample that is uniform
and free of mechanical impurities.
4.3.3 Instruments and apparatuses
4.3.3.1 Analytical balance. sensitivity 0.1 mg.
4.3.3.2 Electric constant temperature forced air drying oven. temperature control
accuracy ±2 °C.
4.3.3.3 Dryer. Use anhydrous calcium chloride or color-changing silica gel as the
desiccant.
4.3.3.4 Weighing bottle. 5 mL ~ 10 mL.
4.3.3.5 Petri dish or metal plate. 75 mm in diameter.
4.3.4 Test procedures
4.3.4.1 Mark a clean petri dish or metal plate; dry it in a drying oven at 150 ℃ ± 2 ℃
for 30 minutes. After removing it, place it in a sealed desiccator and cool it to room
temperature. Weigh it on a balance (m1) to the nearest 0.1 mg.
4.3.4.2 Shake the sample and pour it into a clean and dry weighing bottle.
4.3.4.3 Weigh 2 g ± 0.2 g of test sample (m2) using the reduction method, accurate to
0.1 mg; place it in a Petri dish or metal plate; carefully flatten the entire bottom.
4.3.4.4 Place the Petri dish or metal tray horizontally on the top layer of a drying oven
that has been pre-set to a constant temperature of 150 °C ± 2 °C and has air blowing.
4.3.4.5 After drying for 60 min ± 2 min, take out and immediately place in a desiccator
to cool; after 20 min ~ 30 min, weigh (m3) accurate to 0.1 mg.
4.3.5 Test results
4.3.5.1 Calculate the solid content according to Formula (2) and round to three
significant figures.
Where.
SC – solid content (mass fraction) of unsaturated polyester resins, %;
m3 – mass of the Petri dish and the residual sample, in grams (g);
m1 – mass of the Petri dish, in grams (g);
4.4.4.2 Immediately pour the mixture into the test tube to a depth of 7 cm ~ 8 cm; insert
the thermocouple into the center of the mixture; then place the test tube in a water bath
at 80 ℃ ± 0.5 ℃, so that the sample liquid level is lower than the water bath surface.
4.4.4.3 A recorder, if used, shall be turned on and run throughout the reaction process.
Alternatively, record the maximum temperature reached by the mixture and the time
taken to rise from 65 °C to the maximum temperature. If the maximum temperature
exceeds 90 °C, record the time taken to rise from 65 °C to 90 °C.
4.4.4.4 Repeat the test until the temperature difference between two consecutive
measurements is no more than 5 °C and the time difference is no more than 10%.
4.4.5 Test results
The single value and average value of the maximum temperature of the two tests; the
single value and average value of the time taken to increase the temperature from 65 °C
to the maximum temperature; if the maximum temperature exceeds 90 °C, the single
value and average value of the time taken to increase the temperature from 65 °C to
90 °C.
4.5 Heat stability at 80 °C
4.5.1 Instruments and apparatuses
4.5.1.1 Electric blast drying oven. temperature control accuracy ±2 °C.
4.5.1.2 Ground-mouth white wide-mouth bottle. 125 mL.
4.5.1.3 Balance. sensitivity 0.2 g.
4.5.2 Test procedures
4.5.2.1 Weigh 100 g ± 1 g of the test sample into a clean, dry, white wide-mouth bottle;
tightly stopper it with a glass stopper; mark it. Each two test samples constitute a group.
4.5.2.2 Place the white wide-mouth bottle containing the test sample into an electric
forced air drying oven maintained at a constant temperature of 80 °C ± 2 °C; start the
test and record the time.
4.5.2.3 Invert the white wide-mouth bottle every 2 hours to check for bubbles. When
the bubbles are no longer flowing and the sample becomes sticky and clumpy, the
sample has gelled. Record the time when gelling occurs.
4.5.3 Test results
The heat stability of the sample at 80 ℃ is measured in hours, and the shorter time value
in each group is used as the test result.
4.6 Gel time at 25 °C
4.6.1 Principle
At 25 °C and a certain formulation (resin, accelerator, initiator), the time it takes for the
resin viscosity to reach 50 Pa·s (usually considered the viscosity of the gel state).
4.6.2 Reagents
The type and amount of initiator and accelerator can be selected according to the type
and application of the resin.
It is recommended to use cyclohexanone peroxide dibutyl phthalate paste or methyl
ethyl ketone peroxide as the initiator, and cobalt naphthenate styrene solution or cobalt
naphthenate solution as the accelerator. The usage amount of initiator and accelerator
is 4%.
4.6.3 Instruments and apparatuses
4.6.3.1 Constant temperature water bath. temperature control accuracy ±0.5 °C.
4.6.3.2 Stopwatch.
4.6.3.3 150 mL straight cylindrical beaker.
4.6.3.4 Thermometer. minimum graduation value 0.1 °C.
4.6.3.5 Balance. maximum weighing capacity 200 g, reading accuracy 0.2 g.
4.6.3.6 Pipette. capacity 5 mL, minimum graduation value 0.05 mL.
4.6.3.7 Dropping bottle. 50 mL.
4.6.4 Test procedures
4.6.4.1 Adjust the water bath temperature to 25 °C ± 0.5 °C.
4.6.4.2 Use a beaker as a container; weigh 100 g of sample and initiator using a balance,
accurate to ±0.2 g. Place the beaker in a water bath (the sample liquid level is 2 cm
below the water surface) to maintain a constant temperature; stir carefully.
4.6.4.3 When the test sample temperature is 25 °C ± 0.5 °C, accurately add the
accelerator using a pipette. When the last drop is added, start the stopwatch and stir the
sample thoroughly.
4.6.4.4 Observe every 30 seconds; use a glass rod to test the flow of the sample, until a
wire drawing state appears; stop the stopwatch; record the time shown on the stopwatch,
which is the gel time.
4.6.5 Test results
4.7.2.1.1 Abbe refractometer or other refractometer capable of producing equivalent
results. The measuring range of the instrument is from 1.300 to 1.700, and the
measuring accuracy is not less than 0.000 3.A temperature control device needs to be
provided for the sample and the prism.
4.7.2.1.2 White light source or sodium lamp light source.
4.7.2.1.3 Thermostat, accurate to ±0.1 °C.
Note. Circulating water shall be distilled water or deionized water.
4.7.2.2 Materials
Contact liquid, α-bromonaphthalene.
4.7.3 Test procedures
4.7.3.1 Calibration instruments. Check the refractometer's sight regularly using the
standard block that comes with the instrument. If the reading differs from the value of
the standard block by more than the measurement accuracy, adjust the instrument
according to the instruction manual. When using an Abbe refractometer, it can be
calibrated with a standard block or double distilled water. The refractive index of water
is 1.333 0 at 20 °C and 1.332 0 at 30 °C. The temperature coefficient is -0.000 1/℃.
4.7.3.2 Place the refractometer in a well-lit location; connect it to a thermostat; adjust
the temperature of the refractometer prism to 20 °C. Keep the water circulating to
ensure the required temperature and maintain it within ±0.1 °C.
4.7.3.3 Before each test, clean the prism surface of the instrument according to the
instruction manual. If not otherwise specified, use anhydrous ethanol and lens cleaning
paper to clean the prism; immediately use dry lens cleaning paper to wipe the prism
clean. Wait 2 minutes for the temperature to stabilize before placing the sample on the
prism.
4.7.3.4 Use a round-ended glass rod or pipette to add the sample to the surface of the
refracting prism so that the prism fits perfectly. Wait 2 minutes for the temperature to
stabilize.
4.7.3.5 Adjust the light source so that the divergent light shines on the prism. Adjust the
instrument's knob until the field of view appears. If the instrument is not equipped with
a compensating prism, adjust the light source to make the field of view boundary clear
and the contrast between light and dark maximized.
4.7.3.6 If the instrument is equipped with a compensating prism, adjust the
compensating prism until the field of view appears with light and dark parts and the
color fringes disappear.
...... Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.
Tips & Frequently Asked QuestionsQuestion 1: How long will the true-PDF of English version of GB/T 7193-2008 be delivered?Answer: The full copy PDF of English version of GB/T 7193-2008 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice. Question 2: Can I share the purchased PDF of GB/T 7193-2008_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 7193-2008_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. www.ChineseStandard.us -- GB/T 7193-2008 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds. Question 5: Should I purchase the latest version GB/T 7193-2008?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB/T 7193-2008 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
How to buy and download a true PDF of English version of GB/T 7193-2008?A step-by-step guide to download PDF of GB/T 7193-2008_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD). Step 2: Search keyword "GB/T 7193-2008". Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart. Step 4: Select payment option (Via payment agents Stripe or PayPal). Step 5: Customize Tax Invoice -- Fill up your email etc. Step 6: Click "Checkout". Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively. Step 8: Optional -- Go to download PDF. Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice. See screenshots for above steps: Steps 1~3 Steps 4~6 Step 7 Step 8 Step 9
|