GB/T 510-1983: PDF in English (GBT 510-1983) GB 510-83
GB
NATIONAL STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
UDC 665.5. 536.423
Replacing GB 510-77
Petroleum products – Determination of
solidification point
ISSUED ON. MARCH 9, 1983
IMPLEMENTED ON. DECEMBER 1, 1983
Approved by. National Standard Bureau
Table of Contents
1 Method summary ... 3
2 Apparatus and materials ... 3
3 Reagents ... 4
4 Preparatory work ... 4
5 Test procedure ... 5
6 Accuracy... 6
7 Report ... 6
Petroleum Products –
Determination of solidification point
This Method applies to determination of solidification point of petroleum products.
The maximum temperature, at which lubricating oils and dark petroleum products cool
down under test conditions so that the liquid surface stops moving, is called
solidification point.
1 Method summary
In accordance with the determination method, a specimen is loaded into a specified
test tube and cooled to the expected temperature, and then the test tube is tilted by 45
degrees for 1 min before it is observed whether the liquid surface moves.
2 Apparatus and materials
2.1 Apparatus
2.1.1 Round-bottom test tube. height 160 ± 10 mm; inner diameter 20 ± 1 mm; there
is a annular mark on the outer wall, 30 mm from the tube bottom.
2.1.2 Round-bottom glass bushing. height 130 ± 10 mm; inner diameter 40 ± 2 mm.
2.1.3 Wide-mouth vacuum flask or cylindrical container holding coolant. height not
less than 160 mm; inner diameter not less than 120 mm; it can be made of ceramics,
glass, wood or sheet iron with insulation layer.
2.1.4 Mercurial thermometer. it meets the specifications of GB 514-75; it is used for
petroleum products whose solidification point is lower than -35°C.
2.1.5 Liquid thermometer. it shall be as specified in GB 514-75, Specification for
liquid thermometers for testing of petroleum products, and shall be used to determine
the petroleum products having a solidification point higher than -35°C.
2.1.6 Thermometer of any type. it is used to measure the temperature of coolant.
2.1.7 Stand. it has a device to hold bushing, coolant container and thermometer.
2.1.8 Water bath.
mesh or a little cotton in a funnel and then spread newly calcined coarse salt crystal
on the funnel. When specimens contain lots of water, they shall be filtered through the
salt layers of 2 to 3 funnels.
4.3 Inject specimen into a dry, clean test tube, to make the liquid level to reach the
annular mark. Use a cork to fix a thermometer in the middle of the test tube, making
the mercury bulb 8 to 10 mm from the tube bottom.
4.4 The test tube containing specimen and thermometer is immersed vertically in a
water bath 50 ± 1°C, until the temperature of the specimen reaches 50 ± 1°C.
5 Test procedure
5.1 Take out the test tube containing specimen and thermometer from the water bath,
wipe the outer wall, use a cork to fix the test tube in a bushing and make sure the outer
wall of the test tube and the inner wall of the bushing are equally distant at any spot.
The installed device shall be fixed vertically on a clamp of the stand, and allowed to
stand at room temperature until the specimen in the test tube cools down to 35 ± 5°C.
And then immerse the set of devices into a container containing coolant. The
temperature of coolant is 7 to 8°C lower than the expected solidification point of the
specimen. The depth of the test tube (outer bushing) immersed in the coolant shall not
be less than 70 mm.
When cooling the specimen, the temperature of coolant shall be accurate to ± 1°C.
When the temperature of the specimen reaches the expected solidification point, tilt
the device immersed in coolant to 45 degrees, and maintain this tilting state for 1 min,
but the specimen part of the device shall still be immersed in coolant.
Then, take the device out carefully from coolant, use industrial alcohol to wipe the outer
wall of the bushing rapidly, place the device vertically and observe through the bushing
whether the liquid level in the test tube is movable.
NOTE When determining a solidification point lower than 0°C, 1 to 2 mL of absolute ethyl
alcohol shall be injected to the bottom of the test tube before test.
5.2 When the liquid level is movable, take out the test tube from the bushing, preheat
the specimen in the test tube to 50 ± 1°C once again, then carry out the determination
at a temperature 4°C lower than the temperature of last test or other temperature even
lower, until some test temperature makes the liquid level stop moving.
NOTE When the test temperature is lower than - 20°C, place the test tube containing
specimen and thermometer in room temperature before redetermination, and wait until
the temperature of the specimen rises to – 20°C before immersing the test tube in
water bath for heating.