Volatile organic compounds in furniture components -- Test method for the rapid determination in the field
| Valid |
BUY with any currencies (Euro, JPY, GBP, KRW etc.): GB/T 39939-2021 Related standards: GB/T 39939-2021
PDF Preview: GB/T 39939-2021
GB/T 39939-2021: PDF in English (GBT 39939-2021) GB/T 39939-2021
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 97.140
CCS Y 80
Volatile organic compounds in furniture components -
Test method for the rapid determination in the field
ISSUED ON: MARCH 09, 2021
IMPLEMENTED ON: OCTOBER 01, 2021
Issued by: State Administration for Market Regulation;
Standardization Administration of the PRC.
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative references ... 4
3 Terms and definitions ... 4
4 Principle ... 5
5 Reagents or materials ... 5
6 Equipment ... 6
7 Test procedure ... 8
8 Test data processing ... 9
Appendix A (Normative) Gas bag ... 11
Volatile organic compounds in furniture components -
Test method for the rapid determination in the field
1 Scope
This document specifies the terms and definitions, principle, reagents or
materials, equipment, test procedure, and test data processing of the test
method for the rapid determination in the field for volatile organic compounds
in furniture components.
This document is applicable to the rapid determination in the field of volatile
organic compounds in furniture components, whose maximum side length does
not exceed 800 mm and height does not exceed 300 mm.
2 Normative references
The contents of the following documents, through normative references in this
text, constitute indispensable provisions of this document. Among them, for
dated references, only the edition corresponding to that date applies to this
document. For undated references, the latest edition (including all amendments)
applies to this document.
GB/T 18204.2-2014 Examination methods for public places - Part 2:
Chemical pollutants
3 Terms and definitions
The following terms and definitions apply to this document.
3.1
Volatile organic compounds in furniture components
Organic compounds which are released from furniture components and can be
detected in the outlet gas of the test equipment.
3.2
Loading factor
The ratio OF the surface area of the furniture component TO the amount of
pressure is formed at the inflation interface. The vacuum pump can have its
own flowmeter or connected to an external flowmeter, which can quantify the
volume of gas filled. The flow accuracy error of the mini vacuum pump is within
±5%.
6.4 Area measuring instrument
Steel ruler or tape measure. The measurement accuracy shall not be less than
1 mm. If an area determination instrument is used, the accuracy shall not be
less than 100 mm2.
7 Test procedure
7.1 Preparation before test
Use the area measuring instrument (6.4) to measure the surface area of the
furniture component sample. According to the dimensions of the furniture
component, select the appropriate gas bag (6.2). Use a tie rod sealing strip to
seal the gas bag; fill the gas bag with the high-purity nitrogen (5.1) of 2/3 of the
total volume of the gas bag. Connect the rapid VOCs analyzer (6.1) to the gas
outlet of the gas bag; open the gas outlet valve; detect the VOCs concentration
of the air in the gas bag. The background concentration of target volatile organic
compounds in the gas bag shall not be higher than 0.005 mg/m3.
7.2 Air replacement and load retention in gas bag
Put the furniture component in the center of the gas bag (6.2); place one end
on the support element (5.2), to ensure that the furniture component does not
fit the gas bag as much as possible. Use a tie rod sealing strip to seal the gas
bag; connect the mini vacuum pump (6.3); vacuum the gas bag. After 5 min,
observe whether the gas bag leaks.
Fill the gas bag with about 1/2 volume of high-purity nitrogen (5.1); then
evacuate the gas bag; and, repeat this operation 3 times. Finally, fill with high-
purity nitrogen of 1/2 to 2/3 of the volume of the gas bag. According to formula
(1), calculate the loading factor of the sample. Make the loading factor of the
sample controlled at 5 m2/m3~10 m2/m3. Record the volume of nitrogen filled.
Close the gas valve; keep the gas bag sealed. Place the gas bag at room
temperature for 60 min; record the on-site temperature and atmospheric
pressure.
7.3 Collection and analysis
7.3.1 Sampling
Before sampling, the gas in the gas bag (6.2) shall be evenly mixed. Connect
the gas bag outlet to the rapid VOCs analyzer (6.1). The sampling pipeline shall
be firmly connected. The connecting pipe shall be as short as possible. It shall
use polytetrafluoroethylene or silica gel material.
7.3.2 Rapid formaldehyde analysis
After the load retention is over, within 30 s, open the outlet of the gas bag (6.2).
Collect the gas in the gas bag into the rapid formaldehyde analyzer (6.1.1). The
reading displayed on the rapid formaldehyde analyzer is the formaldehyde
concentration value. The measured value of the volume fraction shall, in
accordance with formula (15) in GB/T 18204.2-2014, 7.4.4.1, be converted into
mass concentration.
7.3.3 Rapid analysis for other aldoketone VOCs
After the load retention is over, within 30 s, open the outlet of the gas bag (6.2).
At a rate of 100 mL/min, collect the gas in the gas bag into the rapid analyzer
for aldoketone VOCs (6.1.2). The sampling time is 10 min. The reading
displayed on the rapid analyzer for other aldoketone VOCs is aldoketone
concentration value.
7.3.4 Rapid analysis for non-aldoketone VOCs
After the load retention is over, within 30 s, open the outlet of the gas bag (6.2).
At a rate of 100 mL/min, collect the gas in the gas bag into the rapid analyzer
for non-aldoketone VOCs (6.1.3). The sampling time is 10 min. The reading
displayed on the rapid analyzer for non-aldoketone VOCs is the concentration
value of non-aldoketone VOCs.
8 Test data processing
8.1 Loading factor of sample
The loading factor of the sample is calculated according to formula (1):
Where:
Lv - Loading factor of sample, in square meters per cubic meter (m2/m3);
d - The surface area of furniture component, in square meters (m2);
V - The volume of nitrogen filled into the gas bag, in cubic meters (m3).