Volatile organic compounds from upholstered
furniture - On-site and rapid detection method
ISSUED ON: MARCH 09, 2021
IMPLEMENTED ON: OCTOBER 01, 2021
Issued by: State Administration for Market Regulation;
Standardization Administration of the PRC.
Table of Contents
Foreword ... 3
1 Scope ... 4
2 Normative references ... 4
3 Terms and definitions ... 4
4 Principle ... 5
5 Equipment ... 5
6 Reagents and materials ... 8
7 Test procedure ... 8
8 Test data ... 11
Volatile organic compounds from upholstered
furniture - On-site and rapid detection method
1 Scope
This document specifies the terms and definitions, principle, equipment,
reagents and materials, test procedure, and test data of the on-site and rapid
detection method for volatile organic compounds from upholstered furniture.
This document is applicable to on-site and rapid detection of volatile organic
compounds from upholstered furniture such as sofas, mattresses, and soft beds
with flat surfaces.
2 Normative references
The contents of the following documents, through normative references in this
text, constitute indispensable provisions of this document. Among them, for
dated references, only the edition corresponding to that date applies to this
document. For undated references, the latest edition (including all amendments)
applies to this document.
GB/T 18204.2-2014 Examination methods for public places - Part 2:
Chemical pollutants
GB/T 39763 Technical requirements for on-site and rapid collection
equipment of volatile organic compounds from furniture
3 Terms and definitions
The following terms and definitions apply to this document.
3.1
Volatile organic compounds from upholstered furniture
Volatile organic compounds released from the tested surface of upholstered
furniture and tested by on-site and rapid detection method.
3.2
Background clean
The background clean can be performed by either internal circulation (7.3.2) or
nitrogen purging mode (7.3.3).
7.3.2 Internal circulation
7.3.2.1 Adjust the three-way valve (position 6 in Figure 1) to the upper passage
position.
7.3.2.2 Set the flow rate of the gas pump (position 3 in Figure 1) to 600 mL/min.
Set the running time to 30 min.
7.3.2.3 Start the gas pump (position 3 in Figure 1). The air in the entire device
generates circulating power; so that the air in the collector, at a flow rate of 600
mL/min, passes through the mixing filter pipe (position 7 in Figure 1) and then
returns to the collector. The volatile organic compounds in the ambient
background air are absorbed by the mixing filter pipe (position 7 in Figure 1), to
form clean air circulation to replace the air in the collector.
7.3.3 Nitrogen purging
Connect the nitrogen source (position 13 in Figure 1) to the intake port (position
1 in Figure 1) on the left side of the mask body. Open the top exhaust port
(position 2 in Figure 1) to exhaust nitrogen. Turn on the nitrogen source, to
purge the air in the mark body at a flow rate of 1 L/min, for no less than 18 min;
replace the air in the mask body with nitrogen. After purging, immediately turn
off the nitrogen source and the top exhaust port.
7.4 Cyclic enrichment
7.4.1 After the background clean, adjust the three-way valve (position 6 in
Figure 1) to the lower passage position.
7.4.2 Start the thermostat (position 9 in Figure 1) to control the heating
temperature to (40±2)°C. Start the gas pump (position 3 in Figure 1). The
temperature of the air entering at the entrance (position 10 in Figure 1) of the
test mask body shall be (36±2)°C. When the air temperature at the entrance
does not meet the requirement of (36±2)°C, the thermostat (position 9 in Figure
1) shall be adjusted. The VOCs released by the sample are enriched in the
collector. The circulating gas flow is 600 mL/min. The enrichment time is 30 min.
7.5 Sampling and rapid VOCs analysis
7.5.1 General
The total amount of gas analyzed in each sampling shall not be greater than
80% of the volume of the mask body.
If the amount of gas collected for sampling analysis is greater than 10% of the
volume of the mask body, when the test is continued, 7.4.2 cyclic enrichment
shall be repeated before sampling.
Under the condition of meeting the detection requirements, the collection
parameters can be set according to the instructions for use of the instrument.
7.5.2 Rapid formaldehyde analysis
After the end of the cyclic enrichment, within 30 s, turn on the gas circuit switch
(position 5 in Figure 1); connect the rapid VOCs analyzer (position 15 in Figure
1) with the on-site and rapid collection equipment. Collect the gas in the mask
body for analysis. The reading displayed on the rapid formaldehyde analyzer is
the formaldehyde concentration value. The measured value of volume fraction
shall, according to formula (15) in 7.4.4.1 of GB/T 18204.2-2014, be converted
into mass concentration.
7.5.3 Rapid analysis for other aldoketone VOCs
After the end of the cyclic enrichment, within 30 s, turn on the gas circuit switch
(position 5 in Figure 1); connect the rapid VOCs analyzer (position 15 in Figure
1) with the on-site and rapid collection equipment. At a rate of 100 mL/min,
collect the gas in the mask body into the rapid analyzer for other aldoketone
VOCs. At the same time of collecting, at the left supply port (position 1 in Figure
1), at the same flow rate, replenish nitrogen (6.2) to the mask body, to maintain
the pressure balance in the system. When the sampling is completed, stop
replenishing gas. When replenishing gas, at the left supply port (position 1 in
Figure 1), it is possible to connect a supplementary gas bag (6.4) filled with
nitrogen. After sampling, the connection shall be disconnected immediately.
The sampling time and gas replenishing time are both 10 min.
After the analysis is completed, the reading displayed on the rapid VOCs
analyzer is the concentration value of other aldoketone VOCs.
7.5.4 Rapid analysis for non-aldoketone VOCs
After the end of the cyclic enrichment, within 30 s, turn on the gas circuit switch
(position 5 in Figure 1); connect the rapid VOCs analyzer (position 15 in Figure
1) with the on-site and rapid collection equipment. At a rate of 100 mL/min,
collect the gas in the mask body into the rapid analyzer for non-aldoketone
VOCs. At the same time of collecting, at the left supply port (position 1 in Figure
1), at the same flow rate, replenish nitrogen (6.2) to the mask body, to maintain
the pressure balance in the system. When the sampling is completed, stop
replenishing gas. When replenishing gas, at the left supply port (position 1 in
Figure 1), it is possible to connect a gas bag (6.4) filled with nitrogen. After
sampling, the connection shall be disconnected immediately. The sampling time