US$369.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. JJF 1429-2013: Calibration Specification for Infrared Mammary Gland Examining Equipments Status: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJF 1429-2013 | English | 369 |
Add to Cart
|
3 days [Need to translate]
|
Calibration Specification for Infrared Mammary Gland Examining Equipments
| Valid |
JJF 1429-2013
|
PDF similar to JJF 1429-2013
Basic data Standard ID | JJF 1429-2013 (JJF1429-2013) | Description (Translated English) | Calibration Specification for Infrared Mammary Gland Examining Equipments | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A53 | Classification of International Standard | 17.080 | Word Count Estimation | 16,156 | Quoted Standard | GB/T 19953-2005; YY 0324-2008; SJ/T 10203-1991; ISO 12233 | Regulation (derived from) | AQSIQ Announcement No. 122, 2013; | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to: wavelength (0. 8 ~ 1. 5) ��m, using infrared transillumination method infrared breast examination calibration. |
JJF 1429-2013: Calibration Specification for Infrared Mammary Gland Examining Equipments---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Infrared Mammary Gland Examining Equipments
People's Republic of China National Metrology Technical Specifications
Infrared breast tester calibration specification
Issued on. 2013-09-02
2013-12-02 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Infrared breast tester calibration specification
ments
Focal point. the National Technical Committee on Medical Measurement
The main drafting units. China Testing Technology Research Institute
Foshan City of Quality Supervision and Testing Center
Participated in the drafting unit. Heilongjiang Provincial Academy of verification test
Xinjiang Institute of Metrology
Measurement and Testing Institute in Neijiang
This specification entrusted to the National Technical Committee is responsible for interpreting medical measurement
The main drafters of this specification.
Zhang from China (China Testing Technology Research Institute)
Gong Lan (China Testing Technology Research Institute)
Li Zhiqiang (Foshan City of Quality Supervision and Testing Center)
Drafters participate.
Huang Guangyi (Heilongjiang Provincial Academy of verification test)
Feng Xuefeng (Xinjiang Institute of Measurement and Testing)
He Jian (Neijiang Metrology Institute)
Fan Jie (China Testing Technology Research Institute)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2. References (1)
3 terminology and units of measurement (1)
4 Overview (1)
5 Metrological characteristics (1)
5.1 infrared spectral range (1)
5.2 infrared light power (2)
5.3 Image Resolution (2)
Calibration Condition 6 (2)
6.1 environmental conditions (2)
6.2 standards and other measuring equipment (2)
7 Calibration methods and calibration items (2)
7.1 appearance and function check (2)
7.2 infrared spectral range (2)
7.3 infrared light power (2)
7.4 Image Resolution (3)
8 calibration results expression (3)
9 Recalibration interval (4)
Appendix A resolution test card (5)
Appendix B-infrared breast examination Calibration original records (recommended) formatting styles (7)
Appendix C calibration certificate within the page (recommended) formatting styles (8)
Appendix D infrared light source power measurements Uncertainty Evaluation Example (10)
Introduction
JJF 1001-2011 "common measurement terms and definitions", JJF 1059.1-2012 "Measurement Uncertainty in
Represents ", JJF 1071-2010" National Calibration Specifications writing rules ", YY0324-2008" breast infrared inspection
Check instrument ", ISO 12233" photographic electronic still camera resolution measurement "(Photography-
Electronicstil-picturecameras-Resolutionmeasurements) together constitute the support of the Specification
Work basic series of documents.
The specification for the first time to develop.
Infrared breast tester calibration specification
1 Scope
This instruction applies to a wavelength (0.8 ~ 1.5) μm, infrared transillumination of breast infrared inspection instrument calibration.
2. References
This specification refers to the following documents.
GB/T 19953-2005 digital camera resolution measurements
YY0324-2008 infrared breast tester
SJ/T 10203-91 CCD color television camera measurement veneer applications
ISO 12233 Photography - Electronic still picture cameras resolution measurements (Photography-Electronic
stil-picturecameras-Resolutionmeasurements)
For dated references, only the dated edition applies to this specification; undated reference documents
Member, the latest edition (including any amendments) applies to this specification.
3 terminology and units of measurement
The following terms and definitions apply to this specification.
3.1 infrared breast tester infraredexaminingequipmentsformammarygland
Using infrared transillumination breast examination (detection) equipment.
NOTE. Breast infrared instrument, infrared breast scanner, near-infrared breast tester are also infrared breast tester.
3.2 infrared light probe infraredlightsourceprobe
Provide infrared transillumination beam probe, the probe types are infrared emission tube probe (halogen lamps, light emitting diodes, laser
Light tubes, etc.), the light guide fiber probe.
3.3 Image resolution imageresolution
The number of lines infrared breast tester of horizontal resolution capabilities, with the unit can distinguish between the image height range of representation, usually
Television is also used bank said, Acronym TVL.
4 Overview
Infrared breast tester (hereinafter referred to as the tester) principle is. Depending on the breast tissue absorption of infrared light
Varying degrees of application in the wavelength range (0.8 ~ 1.5) μm infrared light transillumination breast tissue, through a dedicated infrared camera
Camera capture an image by computer processing, the infrared absorption breast tissue image is displayed on the screen. The main tester
Probe by the infrared light source, high-resolution camera, the processing terminal and other components.
5 Metrological characteristics
5.1 infrared spectral range
Distribution center wavelength range. (0.8 ~ 1.5) μm.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJF 1429-2013_English be delivered?Answer: Upon your order, we will start to translate JJF 1429-2013_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJF 1429-2013_English with my colleagues?Answer: Yes. The purchased PDF of JJF 1429-2013_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|