US$145.00 · In stock · Download in 9 secondsYS/T 630-2016: Chemical analysis method of alumina - Determination of impurities in alumina - ICP-AES analytical method Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See step-by-step procedureStatus: Valid YS/T 630: Evolution and historical versions
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivery | Name of Chinese Standard | Status |
| YS/T 630-2016 | English | 145 |
Add to Cart
|
0-9 seconds. Auto-delivery
|
Chemical analysis method of alumina - Determination of impurities in alumina - ICP-AES analytical method
| Valid |
| YS/T 630-2007 | English | 319 |
Add to Cart
|
3 days
|
Chemical analysis methods of alumina. Determination of impurities in alumina. Inductively coupled plasma atomic emission spectrometric method
| Obsolete |
Excerpted PDFs (Download full copy in 9 seconds upon purchase)PDF Preview: YS/T 630-2016
YS/T 630-2016: Chemical analysis method of alumina - Determination of impurities in alumina - ICP-AES analytical method ---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/YST630-2016
YS
NONFERROUS METAL INDUSTRY STANDARD
OF THE PEOPLE'S REPUBLIC OF CHINA
ICS 77.120.99
H 12
Replacing YS/T 630-2007
Chemical analysis method of alumina - Determination of
impurities in alumina - ICP-AES analytical method
Issued on: JULY 11, 2016
Implemented on: JANUARY 1, 2017
Issued by. Ministry of Industry and Information Technology of PRC
Table of Contents
Foreword... 3
1 Scope... 4
2 Normative references... 4
3 Method summary... 5
4 Reagents and materials... 5
5 Instruments... 6
6 Samples... 6
7 Analysis steps... 6
8 Calculation of measurement results... 7
9 Precision... 8
10 Quality control and assurance... 10
Chemical analysis method of alumina - Determination of
impurities in alumina - ICP-AES analytical method
1 Scope
This standard specifies the method for determining the content of impurity elements
such as silicon dioxide, ferric oxide, sodium oxide, potassium oxide, copper oxide,
magnesium oxide, calcium oxide, boron trioxide, chromium trioxide, vanadium
pentoxide, zinc oxide, titanium dioxide, manganese oxide, gallium trioxide, lithium
oxide, and beryllium oxide in alumina.
This standard applies to the determination of the content of impurity elements such as
silicon dioxide, ferric oxide, sodium oxide, potassium oxide, copper oxide, magnesium
oxide, calcium oxide, boron trioxide, chromium trioxide, vanadium pentoxide, zinc
oxide, titanium dioxide, manganese oxide, gallium trioxide, lithium oxide, and
beryllium oxide in alumina. The measurement range is shown in Table 1.
Table 1 -- Measurement range
2 Normative references
The following documents are essential for the application of this document. For any
dated referenced document, only the dated version applies to this document. For any
undated referenced document, the latest version (including all amendments) applies to
this document.
GB/T 602 Chemical reagent - Preparations of standard solutions for impurity (ISO
6353-1.1982, NEQ)
Impurity element Measurement range (mass fraction)/%
Vanadium pentoxide
Impurity element Measurement range (mass fraction)/%
Silicon dioxide
Ferric oxide
Sodium oxide
Potassium oxide
Copper oxide
Magnesium oxide
Calcium oxide
Boron trioxide
Chromium trioxide
Zinc oxide
Titanium dioxide
Manganese oxide
Gallium trioxide
Lithium oxide
Beryllium oxide
3 Method summary
Under the action of a microwave digestion system, the test specimen is digested with
sulfuric acid at high temperature and high pressure, and the test solution is introduced
into argon plasma. Under the selected optimal operating conditions, the impurity
element content is determined by an inductively coupled plasma spectrometer, and the
matrix matching method is used to correct the influence of the matrix on the
determination.
4 Reagents and materials
4.1 Argon (with a purity of ≥99.99%).
4.2 Sulfuric acid (1+2). Guaranteed or UP grade.
4.3 Sulfuric acid (1+1). Guaranteed or UP grade.
4.4 High-purity aluminum (≥99.999%).
4.5 Deionized water or ultrapure water.
4.6 Aluminum matrix solution (20 mg/mL). Weigh 5.2873 g of high-purity aluminum
(4.4) into a 1000 mL beaker; add 400 mL of sulfuric acid (4.3), place the beaker on an
electric stove, and slowly heat until completely dissolved. Cool, transfer the solution to
a 500 mL volumetric flask, dilute to the mark with water, mix thoroughly, and store in
a polyethylene bottle. 1 mL of this solution contains 20 mg of aluminum oxide.
4.7 Standard storage solution
4.7.1 Prepare standard solutions of silicon, iron, sodium, potassium, copper, magnesium,
calcium, chromium, vanadium, zinc, titanium, manganese, gallium, lithium, beryllium,
and boron according to GB/T 602, with the concentrations of silicon, potassium,
vanadium, and titanium at 100 μg/mL and those of iron, sodium, copper, magnesium,
calcium, chromium, zinc, manganese, gallium, lithium, beryllium, and boron at 1000
μg/mL.
4.8 Mixed standard solution. Pipette 10.00 mL each of the iron and calcium standard
solutions (4.7), 50.00 mL of the sodium standard solution (4.7), 5.00 mL each of the
copper, magnesium, chromium, zinc, manganese, gallium, lithium, beryllium, and
boron standard solutions (4.7), and 50.00 mL each of the vanadium and titanium
standard solutions (4.7) into a 250 mL volumetric flask, dilute to the mark with water,
mix thoroughly, and transfer to a polyethylene bottle for storage. This solution contains
40 μg/mL each of iron and calcium, 200 μg/mL of sodium, and 20 μg/mL each of copper,
magnesium, chromium, vanadium, zinc, titanium, manganese, gallium, lithium,
beryllium and boron.
5 Instruments
5.1 Inductively Coupled Plasma-Atomic Emission Spectrometer (ICP-AES).
5.2 Dryer. New activated alumina is used as the desiccant.
5.3 Microwave digestion apparatus. with a rated temperature of 300 °C, and a digestion
vessel volume of 100 mL.
5.4 Polytetrafluoroethylene sealed sample dissolver (autoclave).
5.5 Electronic balance. The sensitivity is 0.0001 g.
5.6 Oven. 300 °C±10 °C.
6 Samples
6.1 The sample shall pass through a sieve with an aperture of 0.125 mm.
6.2 The sample shall be dried in an oven (5.6) at 300 ℃±10 ℃ for 2 hours and placed
in a desiccator (5.2) to cool to room temperature.
7 Analysis steps
7.1 Test specimen
Weigh 0.2 g of the sample (Chapter 6) to the nearest 0.0001 g.
7.2 Number of measurements
Two measurements shall be performed independently, and the average value shall be
taken.
7.3 Blank test
A blank test shall be performed along with the specimen test.
7.4 Preparation of analytical solution
7.4.1 Place the specimen (7.1) in a polytetrafluoroethylene digestion vessel, add 12.0
mL of sulfuric acid (4.2), and digest in a microwave digester (5.3) until the sample is
completely dissolved. Remove the digestion vessel, cool to room temperature, transfer
the solution to a 100 mL volumetric flask, dilute to the mark with water, and mix
thoroughly.
7.4.2 Place the specimen (7.1) in a polytetrafluoroethylene reaction cup, add 12.0 mL
of sulfuric acid (4.2), and cover tightly. Place the reaction cup in a sample dissolver and
cover tightly. Heat the dissolver to 240 °C±2 °C in an oven, maintain for 6 hours,
remove the dissolver, and allow it to cool naturally to room temperature. Remove the
reaction cup, transfer the solution to a 100 mL volumetric flask, dilute to the mark with
water, and mix thoroughly.
7.5 Preparation of standard series solutions
7.5.1 Pipette 0 mL, 1.00 mL, 3.00 mL, 5.00 mL, 8.00 mL, and 10.00 mL of the silicon
standard solution (4.7) into a set of 100 mL volumetric flasks. Add 10.00 mL of the
aluminum matrix solution (4.6) to each flask. Dilute to the mark with water, mix
thoroughly, and store in polyethylene bottles.
7.5.2 Pipette 0 mL, 1.00 mL, 3.00 mL, 5.00 mL, 8.00 mL, and 10.00 mL of potassium
standard solution (4.7) into a set of 100 mL volumetric flasks. Add 10.00 mL of
aluminum matrix solution (4.6) to each flask. Dilute to the mark with water, mix well,
and store in polyethylene bottles.
7.6 Determination
Under the selected instrument's optimal working conditions, measure the standard
series solutions (7.5) and the analytical solution (7.4).
8 Calculation of measurement results
The contents of the standard series solutions are directly input into the corresponding
computer software system. Based on the strength values of the standard series solutions
and the analytical solution, the computer calculates, corrects, and outputs the
concentration (μg/mL) of the element to be measured in the analytical solution. The
mass fraction of each element to be measured is calculated according to formula (1).
...... Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.
Tips & Frequently Asked QuestionsQuestion 1: How long will the true-PDF of English version of YS/T 630-2016 be delivered?Answer: The full copy PDF of English version of YS/T 630-2016 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice. Question 2: Can I share the purchased PDF of YS/T 630-2016_English with my colleagues?Answer: Yes. The purchased PDF of YS/T 630-2016_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. www.ChineseStandard.us -- YS/T 630-2016 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds. Question 5: Should I purchase the latest version YS/T 630-2016?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version YS/T 630-2016 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
How to buy and download a true PDF of English version of YS/T 630-2016?A step-by-step guide to download PDF of YS/T 630-2016_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD). Step 2: Search keyword "YS/T 630-2016". Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart. Step 4: Select payment option (Via payment agents Stripe or PayPal). Step 5: Customize Tax Invoice -- Fill up your email etc. Step 6: Click "Checkout". Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively. Step 8: Optional -- Go to download PDF. Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice. See screenshots for above steps: Steps 1~3 Steps 4~6 Step 7 Step 8 Step 9
|