GB/T 33647-2017 (GB/T33647-2017, GBT 33647-2017, GBT33647-2017) & related versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | See Detail | Status | Similar PDF |
GB/T 33647-2017 | English | 130 |
Add to Cart
|
0-9 seconds. Auto delivery.
|
Determination of silicon content in motor gasoline -- Inductively coupled plasma optical emission spectrometry (ICP-OES)
|
GB/T 33647-2017
| Valid |
GBT 33647-2017
|
Buy with any currencies (Euro, JPY, KRW...): GB/T 33647-2017 Preview this PDF: GB/T 33647-2017
GB/T 33647-2017: PDF in English (GBT 33647-2017) GB/T 33647-2017
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 75.160.20
E 30
Determination of Silicon Content in Motor Gasoline
– Inductively Coupled Plasma Optical
Emission Spectrometry (ICP-OES)
ISSUED ON. MAY 12, 2017
IMPLEMENTED ON. DECEMBER 1, 2017
Issued by. General Administration of Quality Supervision, Inspection and
Quarantine;
Standardization Administration of PRC.
3. No action is required - Full-copy of this standard will be automatically &
immediately delivered to your EMAIL address in 0~60 minutes.
Table of Contents
Foreword ... 3
Introduction ... 4
1 Scope ... 5
2 Normative References ... 5
3 Method Summary ... 6
4 Interference ... 6
5 Apparatus ... 6
6 Reagents and Materials ... 7
7 Sampling ... 8
8 Preparation of Standard Solution and Specimen Solution ... 8
9 Apparatus Preparation ... 8
10 Calibration ... 9
11 Test Procedures ... 9
12 Result Calculation ... 10
13 Quality Control ... 10
14 Precision ... 10
15 Test Report ... 11
Bibliography ... 12
Foreword
This Standard was drafted as per the rules specified in GB/T 1.1-2009.
This Standard was proposed by and under the jurisdiction of National
Technical Committee for Standardization of Petroleum Products and Lubricants
(SAC/TC 280).
Drafting organizations of this Standard. National Quality Supervision & Testing Center
for Petroleum and Petrochemical Products (Guangdong), National Petroleum Products
Quality Supervision and Inspection Center (Shenyang), Northeast Marketing Branch
of PetroChina Co., Ltd., Sinopec Research Institute of Petroleum Processing, and
PetroChina Petrochemical Research Institute.
Chief drafting staffs of this Standard. Wen Huan, Zhang Wenmei, Lv Huanming, Yang
Lihua, Gao Ping, Wang Ke, He Jing, and Liu Huiqin.
Determination of Silicon Content in Motor Gasoline
– Inductively Coupled Plasma Optical
Emission Spectrometry (ICP-OES)
1 Scope
This Standard specifies the test method of adopting the inductively coupled plasma
optical emission spectrometry (ICP-OES) to determine the silicon content in motor
gasoline.
This Standard is applicable to determine the silicon content of 1.0mg/kg ~ 50.0mg/kg
in the motor gasoline (the volume fraction of oxygen-containing compounds shall not
exceed 15%), such as motor gasoline containing methyl tert-butyl ether, motor
methanol gasoline (M15) and motor ethanol gasoline (E10). For the motor gasoline
sample with silicon content higher than 50.0mg/kg, it can be diluted by the isooctane
with higher proportion, then determine according to this Standard method; but its
precision has not been statistics yet.
When testing according to this Standard, some elements in the motor gasoline like
sulfur, lead, iron, manganese, phosphorus and chlorine shall not interfere the
determination results of the silicon content.
2 Normative References
The following documents are essential to the application of this document. For the
dated documents, only the versions with the dates indicated are applicable to this
document; for the undated documents, only the latest version (including all the
amendments) are applicable to this document.
GB/T 4756 Method for Manual Sampling of Petroleum Liquids
GB/T 17476 Standard Test Method for Determination of Additive Elements, Wear
Metals, and Contaminants in Used Lubricating Oils and Determination of Selected
Elements in Base Oils by Inductively Coupled Plasma Optical Emission
Spectrometry
NOTE. When using different types of atomizers, it may affect the determination results of the
silicon content.
5.3 Peristaltic pump (recommended)
In order to provide stable sample injection operation, it is recommended to use
peristaltic pump for sample injection, the pump speed range is of 0.5mL/min ~ 3mL/min.
5.4 Sample injection tube of peristaltic pump
The use of sample injection tube of peristaltic pump shall ensure when contact with
the organic solvent (diluent) for at least 6h, it is exempted from swelling, dissolving,
deforming, hardening, and the like appearance changes; it is recommended to use
synthetic rubber tube.
5.5 Specimen container
30mL ~ 120mL plastic bottle or glass bottle with threaded cap.
5.6 Balance
Accurate to 0.0001g.
6 Reagents and Materials
6.1 Isooctane. analytically pure.
6.2 Silicon organic standard material. prepare or purchase the silicon organic
standard solution with concentration no less than 500mg/kg; for instance, dilute the
hexamethyldisiloxane with purity greater than 99% with isooctane, then prepare the
1000mg/kg silicon organic standard solution.
NOTE. for the motor gasoline sample with silicon content greater than 100mg/kg, it is more
suitable to use hexamethyldisiloxane as standard material; while the base oil-based silicon
organic standard solution may result in higher test results.
6.3 Argon, oxygen. purity no less than 99.995% (volume fraction).
6.4 10% (mass fraction) nitric acid solution.
6.5 Quality control (QC) sample. select the motor gasoline with the similar silicon
content as to-be-tested motor gasoline as the quality control sample, it can also be
obtained by diluting the proper amount of silicon organic standard material with
isooctane. The quality control sample is recommended to seal in the glass bottle with
threaded cap, and store in the dark at less than 10°C environment.
11.2 The blank solution and specimen solution shall be measured for three repeated
times. The average blank response value shall be subtracted from the signal response
value of each specimen solution before determining the average response value of
silicon in the specimen solution.
12 Result Calculation
The silicon content X [in (mg/kg)] in the motor gasoline sample shall be calculated by
Formula (1).
Where.
Ic – average response value of silicon in specimen solution;
Ibk – average response value of blank solution;
S – standard curve slope (kg/mg);
Kg – mass dilution factor, namely, sample mass/ total mass of sample plus dilution
solvent (g/g).
13 Quality Control
13.1 Determine the quality control sample’s average value and control limit before
use the quality control sample.
13.2 It is recommended to use quality control sample to check once every day when
the sample is measured. When measuring large quantities of samples, increase the
inspection frequency of quality control sample.
13.3 If the obtained results exceed the control limit, in order to ensure the accuracy
of the test results, re-calibrate the instrument and re-measure the sample.
14 Precision
14.1 Overview
According to the method in GB/T 6683, the precision of this Standard was obtained
through the test results of 5 motor gasoline samples with silicon content of 1.0mg/kg ~
50.0kg/mg in 12 collaborative laboratories. The instruments used in the collaborative
laboratories are equipped with oxygenating device and atomization chamber cooling
......
Standard ID | GB/T 33647-2017 (GB/T33647-2017) | Description (Translated English) | Determination of silicon content in motor gasoline -- Inductively coupled plasma optical emission spectrometry (ICP-OES) | Sector / Industry | National Standard (Recommended) | Classification of Chinese Standard | E30 | Classification of International Standard | 75.160.20 | Word Count Estimation | 9,958 | Date of Issue | 2017-05-12 | Date of Implementation | 2017-12-01 | Quoted Standard | GB/T 4756; GB/T 17476 | Drafting Organization | National Petroleum and Petrochemical Products Quality Supervision and Inspection Center (Guangdong), the National Petroleum Product Quality Supervision and Inspection Center (Shenyang), China Petroleum and Natural Gas Co., Ltd. Northeast Sales Company, China Petroleum & Chemical Corporation Petrochemical Research Institute, China National Petroleum Corporation Limited petrochemical research institute | Administrative Organization | National Standard Committee on Petroleum Products and Lubricants (SAC/TC 280) | Proposing organization | National Standard Committee on Petroleum Products and Lubricants (SAC/TC 280) | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, China National Standardization Administration Committee | Summary | This standard specifies a test method for the determination of silicon content in automotive gasoline using an inductively coupled plasma emission spectrometer (ICP-OES). This standard applies to the determination of silica content of 1.0 mg/kg ~ 50.0 mg/kg of vehicle gasoline (oxygen content of the volume fraction of not more than 15%), such as methyl tert-butyl ether vehicle gasoline, vehicle methanol gasoline (M15) and vehicle ethanol gasoline (E10). For automotive gasoline samples with a silicon content above 50.0 mg/kg can be diluted by a higher proportion of isooctane and determined according to this standard method, but its precision is not yet counted. When used in this standard, certain elements such as sulfur, lead, iron, manganese, phosphorus and chlorine in vehicle gasoline do not interfere with the results of silicon measurement. |
|