US$390.00 · In stock · Download in 9 secondsGB/T 29628-2013: Guides for methods of measurement of the magnetic properties of permanent (magnetically hard) materials by pulsed field magnetometry Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See step-by-step procedureStatus: Valid
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivery | Name of Chinese Standard | Status |
GB/T 29628-2013 | English | 390 |
Add to Cart
|
0-9 seconds. Auto-delivery
|
Guides for methods of measurement of the magnetic properties of permanent (magnetically hard) materials by pulsed field magnetometry
| Valid |
Excerpted PDFs (Download full copy in 9 seconds upon purchase)PDF Preview: GB/T 29628-2013
GB/T 29628-2013: Guides for methods of measurement of the magnetic properties of permanent (magnetically hard) materials by pulsed field magnetometry ---This is an excerpt. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.), auto-downloaded/delivered in 9 seconds, can be purchased online: https://www.ChineseStandard.net/PDF.aspx/GBT29628-2013
GB
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 29.100.01
K 14
Guides for methods of measurement of the magnetic
properties of permanent (magnetically hard) materials by
pulsed field magnetometry
(IEC TR 62331.2005, Pulsed field magnetometry, NEQ)
Issued on: JULY 19, 2013
Implemented on: DECEMBER 02, 2013
Issued by. General Administration of Quality Supervision, Inspection and
Quarantine of the People's Republic of China;
Standardization Administration of the People's Republic of China.
Table of Contents
Foreword... 3
Introduction... 4
1 Scope... 5
2 Normative references... 5
3 Terms and definitions... 5
4 Measuring principle... 6
5 Specimen sizes... 7
6 Basic components... 7
7 Effects of temperature... 16
8 Calibration... 17
9 Reproducibility... 19
10 Test report... 19
Annex A (informative) Measurement comparison... 21
Bibliography... 22
Guides for methods of measurement of the magnetic
properties of permanent (magnetically hard) materials by
pulsed field magnetometry
1 Scope
This Standard specifies the method for measuring the magnetic properties of permanent
magnet (magnetically hard) materials using a permanent magnet pulse measuring
instrument.
This Standard is applicable to the measurement of magnetic properties of permanent
magnet materials with high intrinsic coercivity, such as rare earth cobalt and rare earth
iron boron.
The method specified in this Standard is an open circuit measurement method. The
measurement method for the magnetic properties of permanent magnet materials in a
closed magnetic circuit is specified in GB/T 3217.The comparison of measurement
curves between this Standard and GB/T 3217 for the same test object can be found in
Annex A.
2 Normative references
The following referenced documents are indispensable for the application of this
document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
GB/T 2900.60, Electrotechnical terminology -- Electromagnetism [IEC 60050(121)]
GB/T 3217, Method of test of the magnetic properties of permanent magnetic
(magnetically hard) materials (IEC 60404-5)
GB/T 9637, Electrotechnical terminology -- Magnetic materials and components
[IEC 60050(221)]
IEC 60050 (151), International electrotechnical vocabulary – Part 151.Electrical
and magnetic devices
3 Terms and definitions
For the purposes of this document, the terms and definitions defined in GB/T 2900.60,
5 Specimen sizes
All results in this Standard are based on cylindrical specimens with a size range from a
maximum diameter of 30 mm and a maximum length of 25 mm to a minimum diameter
of 5 mm and a minimum length of 5 mm. But the specimen is measured in an open
magnetic circuit, so there is no direct limitation on the size of the specimen. If the eddy
current effect can be compensated for and the design of the testing instrument meets the
requirements of a uniform magnetic field zone, then both large and small specimens
can be measured.
6 Basic components
6.1 Software and hardware conditions for testing
The permanent magnet pulse measurement method shall have the following software
and hardware conditions.
- Magnetic field generator; including.
• Capacitive discharge power supply;
Magnetized solenoid.
- Magnetic polarization strength and magnetic field strength testing coil;
- Data collection and digitization equipment; including.
• integrator;
• Digital devices.
- Data processing software with the following functions.
• "Zero signal" processing;
• Positioning of J (H) hysteresis loop;
• Self demagnetization correction;
• Low pass filtering;
• Eddy current correction.
6.2 Magnetic field generator
6.2.1 Overview
Under conditions where the energy is sufficient to saturate the specimen magnetization,
the repeatability of the capacitor charging voltage is relatively unimportant. The only
sensitivity to voltage repeatability is the correction of the "zero signal" (see 6.5.2). ±1%
voltage repeatability is relatively easy to achieve. Reaching this level shall not affect
the correction of the 'zero signal'. In general, the voltage repeatability can reach ±0.1%.
6.2.3 Magnetized solenoid
6.2.3.1 Peak magnetic field strength
The impedance of a magnetized solenoid can affect the strength of the magnetic field.
The peak magnetic field strength generated by the magnetized solenoid shall cause the
specimen to reach magnetization saturation in both positive and negative directions,
with a margin of+10%. The required magnetic field strength depends on the material,
orientation, and self-demagnetization factor of the specimen.
6.2.3.2 Dimensions
The magnetized solenoid shall have a certain volume to accommodate the specimen
and test coil system.
6.2.3.3 Magnetic field uniformity
The magnetic field uniformity within the entire specimen area shall be within ± 1%.
6.2.3.4 Magnetic field frequency
The magnetic field shall have a low time change rate dH/dt to avoid generating large
eddy currents in conductive samples and to facilitate the correction of eddy current
effects.
6.3 Magnetic polarization strength and magnetic field strength test coil
6.3.1 Overview
The permanent magnet pulse measurement method uses the values of the magnetic
polarization intensity of the specimen induced by the test coil and the magnetic field
intensity of the specimen. The test coil includes.
- Magnetic polarization strength testing coil (J coil);
- Magnetic field strength testing coil (H coil).
6.3.2 J coil
The J coil is composed of a test coil and a compensation coil connected in series and
reversed. The test coil and the compensation coil have equal ampere turns, but the
degree of coupling with the specimen is different, so the measured magnetic
polarization strength of the specimen does not include the magnetic field signal. The J
to become measurement results.
After the above processing, the J signal plots the synchronized H signal to obtain the J
(H) hysteresis loop. The B (H) hysteresis loop can be obtained by converting the J (H)
hysteresis loop.
For the measurement results of conductive permanent magnet specimens, the influence
of eddy currents needs to be corrected.
6.5.2 "Zero signal"
Perform a permanent magnet pulse measurement without loading the specimen. The
apparent magnetic polarization signal obtained by the system at this time is generally
referred to as the "zero signal". For a precisely designed and manufactured testing
system, the signal shall be smaller and have good repeatability compared to the signal
generated by the specimen.
The "zero signal" is caused by eddy currents within the conductive material of the
magnetized solenoid and other factors within the magnetization field region. The
amplitude and waveform of the signal shall vary under different magnetic field
strengths.
Before measuring the sample, the "zero signal" shall be measured in advance under the
same magnetic field conditions without the sample. The sample measurement results
can only proceed to the next step of data processing after deducting this "zero signal".
If the "zero signal" accounts for 1% of the entire measurement signal and the change in
the "zero signal" between two measurements is 10%, then the measurement error of
magnetic polarization intensity J caused by the "zero signal" is about 0.1%.
6.5.3 Positioning of J (H) hysteresis loop
Although the external magnetic field increases from zero and eventually returns to zero,
the initial magnetic polarization strength of the specimen is unknown (the initial
magnetic polarization strength signal is related to the magnetization state of the
specimen). Therefore, it is necessary to perform positioning processing on the return
line. If the specimen reaches magnetization saturation in both positive and negative
directions, the midpoint between the positive and negative residual magnetic
polarization values can be considered as J=0.
6.5.4 Self demagnetization correction
The permanent magnet pulse measurement method adopts open magnetic circuit
measurement. In order to obtain the intrinsic magnetic properties of the specimen, a
self-demagnetization factor shall be used to correct the test results. This is also a
common problem faced by all open circuit measurement methods. The self-
demagnetization factor depends on the geometric dimensions of the specimen and the
generating heat, the temperature of the specimen can increase. On the other hand, the
heating of the magnetized solenoid shall also increase the temperature of the specimen
inside it. The design of a magnetized solenoid shall consider both the maximum
magnetization field and the factors causing heat transfer to the specimen due to the
heating of the solenoid. The magnetized solenoid circulating cooling system can
effectively reduce the temperature rise of the specimen caused by the heat conduction
of the magnetized solenoid.
7.2 Temperature measurement
The temperature of the specimen can be determined by various methods. The method
of direct contact temperature measurement between thermocouple and specimen can be
used. Laser temperature measurement method can also be used. Laser is directed
towards the magnet through the bottom opening of the magnetized solenoid, enabling
non-contact temperature measurement.
8 Calibration
The calibration of permanent magnet pulse measuring instruments mainly focuses on
the calibration of the H channel and J channel.
8.1 H channel calibration
Place the H coil in a uniform, static magnetic field calibrated by a nuclear magnetic
resonance magnetometer. Quickly remove this coil to determine the overall accuracy of
the test coil and integrator.
8.2 J channel calibration
The pure nickel standard specimen has a determined theoretical saturation magnetic
polarization strength. The eddy currents generated internally are basically the same as
those in pure copper specimens. The total eddy currents in the magnetic saturation
region of pure nickel specimens are independent of their relative magnetic permeability.
Therefore, the saturation magnetization state of pure nickel standard specimens can be
used for calibration of the J channel.
Under identical magnetic fields, two pure nickel and copper specimens with identical
shapes and sizes are measured using a permanent magnet pulse measuring instrument.
Using an iterative method, by adjusting the eddy current scaling factor of the pure
copper standard specimen, the eddy current in the pure copper standard specimen is
subtracted from the measurement results of the pure nickel standard specimen until the
eddy current influence in the saturation region of the hysteresis loop of the pure nickel
standard specimen is completely removed. After the above processing, the theoretical
saturation magnetic polarization intensity of the pure nickel standard specimen can be
used for calibration of the J channel.
...... Source: Above contents are excerpted from the full-copy PDF -- translated/reviewed by: www.ChineseStandard.net / Wayne Zheng et al.
Tips & Frequently Asked QuestionsQuestion 1: How long will the true-PDF of English version of GB/T 29628-2013 be delivered?Answer: The full copy PDF of English version of GB/T 29628-2013 can be downloaded in 9 seconds, and it will also be emailed to you in 9 seconds (double mechanisms to ensure the delivery reliably), with PDF-invoice. Question 2: Can I share the purchased PDF of GB/T 29628-2013_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 29628-2013_English will be deemed to be sold to your employer/organization who actually paid for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. www.ChineseStandard.us -- GB/T 29628-2013 -- Click this link and select your country/currency to pay, the exact amount in your currency will be printed on the invoice. Full PDF will also be downloaded/emailed in 9 seconds.
How to buy and download a true PDF of English version of GB/T 29628-2013?A step-by-step guide to download PDF of GB/T 29628-2013_EnglishStep 1: Visit website https://www.ChineseStandard.net (Pay in USD), or https://www.ChineseStandard.us (Pay in any currencies such as Euro, KRW, JPY, AUD). Step 2: Search keyword "GB/T 29628-2013". Step 3: Click "Add to Cart". If multiple PDFs are required, repeat steps 2 and 3 to add up to 12 PDFs to cart. Step 4: Select payment option (Via payment agents Stripe or PayPal). Step 5: Customize Tax Invoice -- Fill up your email etc. Step 6: Click "Checkout". Step 7: Make payment by credit card, PayPal, Google Pay etc. After the payment is completed and in 9 seconds, you will receive 2 emails attached with the purchased PDFs and PDF-invoice, respectively. Step 8: Optional -- Go to download PDF. Step 9: Optional -- Click Open/Download PDF to download PDFs and invoice. See screenshots for above steps: Steps 1~3 Steps 4~6 Step 7 Step 8 Step 9
|