HOME   Cart(1)   Quotation   About-Us Tax PDFs Standard-List
Powered by Google-Search & Google-Books www.ChineseStandard.net Database: 189759 (28 May 2023)

GB/T 18457-2015 (GBT 18457-2015)

Chinese standards (related to): 'GB/T 18457-2015'
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)See DetailStatusGoogle Book
GB/T 18457-2015English125 Add to Cart 0-9 seconds. Auto delivery. Stainless steel needle tubing for the manufacture of medical devices GB/T 18457-2015 Valid GBT 18457-2015
GB 18457-2001English70 Add to Cart 0-9 seconds. Auto delivery. Stainless steel needle tubing for manufacture of medical devices GB 18457-2001 Obsolete GB 18457-2001


   
BASIC DATA
Standard ID GB/T 18457-2015 (GB/T18457-2015)
Description (Translated English) Stainless steel needle tubing for the manufacture of medical devices
Sector / Industry National Standard (Recommended)
Classification of Chinese Standard C31
Classification of International Standard 11.040.20
Word Count Estimation 19,146
Date of Issue 2015-12-10
Date of Implementation 2017-01-01
Older Standard (superseded by this standard) GB 18457-2001
Quoted Standard GB/T 6682; GB/T 14233.1-2008; ISO/TS 15510-2003
Drafting Organization Zhejiang Kindly Medical Devices Co., Ltd.
Administrative Organization National medical syringe (needle) Standardization Technical Committee (SAC/TC 95)
Regulation (derived from) National Standard Announcement 2015 No.38
Proposing organization China Food and Drug Administration
Issuing agency(ies) Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China; Standardization Administration of China
Summary This standard specifies the size, surface and mechanical properties of normal wall, thin wall and ultra thin wall tube of 0.6mm ~ 2.1mm in metric size of 0.2mm ~ 3.4mm. This standard applies to the human skin, subcutaneous, muscle and intravenous injection needle and other medical equipment with hard straight stainless steel needle (hereinafter referred to as needle). This standard does not apply to pliable stainless steel syringes, which are different from the hard straight stainless steel syringes specified in this standard, but encourage the manufacturer and purchaser of the flexible needle to use the dimensions specified in this standard.


GB/T 18457-2015: PDF in English (GBT 18457-2015)
GB/T 18457-2015
GB
NATIONAL STANDARD
OF THE PEOPLE’S REPUBLIC OF CHINA
ICS 11.040.20
C 31
Replacing GB/T 18457-2001
Stainless-steel needle tubing for
the manufacture of medical devices
(ISO 9626.1991, MOD)
ISSUED ON. DECEMBER 10, 2015
IMPLEMENTED ON. JANUARY 01, 2017
Issued by. General Administration of Quality Supervision, Inspection and
Quarantine of the People's Republic of China;
Standardization Administration of the PRC.
Table of contents
Foreword ... 3 
1 Scope ... 5 
2 Normative references ... 5 
3 Materials ... 5 
4 Size ... 6 
5 Specification indication ... 7 
6 Surface ... 7 
7 Cleanliness ... 8 
8 pH ... 8 
9 Rigidity ... 8 
10 Toughness ... 9 
11 Corrosion resistance ... 10 
Appendix A (Informative) Stainless-steel designations ... 11 
Appendix B (Normative) Test solution preparation method ... 13 
Appendix C (Normative) Needle tubing pH test method ... 14 
Appendix D (Normative) Needle tubing rigidity test method ... 15 
Appendix E (Normative) Needle tubing toughness test method ... 17 
Appendix F (Normative) Test method for corrosion resistance of needle tubing
... 18 
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
This standard replaces GB 18457-2001 “Stainless-steel needle tubing for the
manufacture of medical devices”.
Compared with GB 18457-2001, the technical change mainly includes.
- ADD the technical requirements for three needle tubing sizes of 0.2 (33G),
0.23 (32G), 0.25 (31G);
- MDOFIY the distance between the fixed support point and the load acting
point for the ductile test of the needle tubing having the specifications of 1.4
mm ~ 3.4 mm in Table 3.
This standard, through the redrafting method, adopts ISO 9626.1991
“Stainless-steel needle tubing for the manufacture of medical devices” AND its
Amendment 1.2002.
The technical differences between this standard and ISO 9626.1991 & its
Amendment 1.2002 AND the causes are as follows.
- On the normative references, adjustment on technical differences is made
in this standard to adapt to the technical conditions of China, AND the
detailed adjustments are provided in Chapter 2 “Normative references”, as
below.
• USE GB/T 6682 which adopts international standard through modification
to replace ISO 3696;
• ADD reference to GB/T 14233.1.
- ADD the designations of austenitic stainless-steel in Chapter 3 AND in the
Appendix A, PROVIDE the equivalent austenitic stainless-steel
designations and chemical composition;
- ADD Appendix C (Normative). Needle tubing pH test method;
This standard has made the following editorial changes.
- ADD Appendix A (Informative) Stainless-steel designations;
- DELETE the Appendix A “Needle tubing pH judgment method” of ISO
9626.1991;
Stainless-steel needle tubing for
the manufacture of medical devices
1 Scope
This standard specifies the size, surface and mechanical properties of the tube
having normal wall and thin wall of nominal specification 0.2 mm ~ 3.4 mm AND
the ultra-thin wall of specification 0.6 mm ~ 2.1 mm.
This standard applies to the rigid straight stainless-steel needle tubing used for
human body skin, subcutaneous, muscle and vein injection (hereinafter referred
to as the needle tubing).
This standard does not apply to easy-to-bend stainless-steel needle tubing,
because its mechanical properties are different from the rigid stainless-steel
needle tube as specified in this standard; however, it is recommended for the
manufacturer and purchaser of the easy-to-bend stainless-steel needle tubing
to select the sizes as specified in this standard.
2 Normative references
The following documents are essential to the application of this document. For
the dated documents, only the versions with the dates indicated are applicable
to this document; for the undated documents, only the latest version (including
all the amendments) are applicable to this Standard.
GB/T 6682 Water for analytical laboratory use - Specification and test
methods (GB/T 6682-2008, ISO 3696.1987. MOD)
GB/T 14233.1-2008 Test methods for infusion transfusion injection
equipment for medical use - Part 1. Chemical analysis methods
ISO/TS 15510.2003 Stainless-steels - Chemical composition
3 Materials
The needle tubing shall be made by such austenitic stainless-steel as
X2CrNi18-9, X5CrNi18-9, X6CrNiNb18-10, X5CrNiMo17-12-2, X6CrNiMoTi17-
12-2, X6CrNiMoNb17-12-2 which comply with ISO/TS 15510.2003, OR other
austenitic stainless-steels equivalent to the aforementioned steels. In Appendix
Appendix B
(Normative)
Test solution preparation method
B.1 Principle
The needle is immersed in water to precipitate a soluble substance.
B.2 Instruments and reagents
B.2.1 Test water is the grade III water newly prepared in accordance with GB/T
6682.
B.2.2 Borosilicate glassware for laboratory use.
B.3 Test solution preparation procedure
B.3.1 WEIGH 3.0 g of needle tubing; CUT it into appropriate lengths; ADD 250
mL of water (B.2.1) into the glassware (B.2.2) of appropriate volume; IMMERSE
the needle tubing into it; MAINTAIN the water and the needle tubing at 37 °C ±
3 °C for 1h; TAKE the needle tubing out; ENSURE that all the water at the inner
and outer surface of the needle tubing are back to the glassware.
B.3.2 FOLLOW the procedures as described in B.3.1 without placing the needle
tubing into the glassware, in order to prepare the blank control solution.
a) MAKE the span equal to the values corresponding to the tested needle
tubing specifications in Table 2;
b) MAKE the end surface of the push rod at the center of the span;
c) MAKE the needle tubing be perpendicular to the two needle tubing
supports and push rod, and meanwhile MAKE the needle tubing centerline
coincide with the needle tubing support centerline.
D.3.2 Based on the force corresponding to the nominal specifications of this
needle tubing in Table 2, APPLY downwards the bending force through the push
rod to the needle tubing at the rate of 1 mm/min.
D.3.3 MEASURE and RECORD the needle tubing deflection at the force
application point D.2.2), accurate to 0.01 mm.
D.4 Test report
The test report contains at least the following information.
a) Specifications and description of the needle tubing;
b) Wall type of needle tubing. normal wall, thin wall, ultra-thin wall;
c) Measured deflection (unit in mm, accurate to 0.01 mm) and judgment
results;
d) Test date.
Appendix F
(Normative)
Test method for corrosion resistance of needle tubing
F.1 Principle
SOAK a portion of the needle tubing into the sodium chloride solution for a
predetermined period of time; COMPARE the soaked portion and the non-
soaked portion; and visually OBSERVE the corrosion trace.
F.2 Instruments and reagents
F.2.1 Sodium chloride solution. Distilled or deionized water in accordance with
the level 3 water requirements in GB/T 6682. PREPARE solution of c (NaCl) =
0.5 mol/L (analytical pure reagent).
F.2.2 Borosilicate glassware for laboratory use.
F.3 Test procedure
PLACE a needle tubing into the glassware (F.2.2) containing sodium chloride
solution (F.2.1) at 23 °C ± 2 °C, so that half of the length of the needle tubing is
immersed in the solution. MAINTAIN the solution and the needle tubing at 23 °C
± 2 °C for 7 h ± 5 min. TAKE the needle tubing out and USE the distilled water
or deionized water to rinse and dry it; USE normal vision or corrected visual
acuity to observe and compare the socked portion and the non-soaked portion,
so as to see whether there is corrosion trace due to immersion.
F.4 Test report
The test report contains at least the following information.
a) Specifications and description of the needle tubing;
b) Wall type of needle tubing. normal wall, thin wall, ultra-thin wall;
c) Corrosion of soaked portion of the needle;
d) Test date.
......