US$279.00 · In stock · Download in 9 secondsYY/T 1996-2025: Brain-computer interface medical device - Closed-loop implantable neurostimulator - Test method for sensing and response performance
Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See
step-by-step procedureStatus: Valid
| Std ID | Version | USD | Buy | Deliver [PDF] in | Title (Description) |
| YY/T 1996-2025 | English | 279 |
Add to Cart
|
3 days [Need to translate]
|
Brain-computer interface medical device - Closed-loop implantable neurostimulator - Test method for sensing and response performance
|
Click to Preview a similar PDF
Basic data
| Standard ID | YY/T 1996-2025 (YY/T1996-2025) |
| Description (Translated English) | Brain-computer interface medical device - Closed-loop implantable neurostimulator - Test method for sensing and response performance |
| Sector / Industry | Medical Device & Pharmaceutical Industry Standard (Recommended) |
| Classification of Chinese Standard | C31 |
| Classification of International Standard | 11.40.040 |
| Word Count Estimation | 14,171 |
| Date of Issue | 2025-09-25 |
| Date of Implementation | 2027-10-01 |
| Issuing agency(ies) | National Medical Products Administration |
YY/T 1996-2025: Brain-computer interface medical device - Closed-loop implantable neurostimulator - Test method for sensing and response performance
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
ICS 11.40.040CCS C 31
Pharmaceutical Industry Standards of the People's Republic of China
Released on 2025-09-25
Implemented on October 1, 2027
Medical devices using brain-computer interface technology
Implantable neurostimulator with closed-loop function
Sensing and Response Performance Testing Methods
Brain‑computer interface medical device-Closed‑loop implantable
National Medical Products Administration issued
Table of contents
Preface III
1.Scope 1
2 Normative References 1
3.Terms and Definitions 1
4.Test Method 2
4.1 Test Data Requirements 2
4.2 Perceived Performance Test 2
4.3 Closed-loop control stimulus-response test 4
4.4 EEG signal acquisition and processing algorithm test 5
Appendix A (Informative) Correspondence between different states of test signals and stimulus protocol modulation 7
Reference 9
Foreword
This document complies with the provisions of GB/T 1.1-2020 "Standardization Work Guidelines Part 1.Structure and Drafting Rules of Standardization Documents".
Drafting.
Please note that some content in this document may involve patents. The issuing organization of this document assumes no responsibility for identifying patents.
This document was proposed by the National Medical Products Administration.
This document is under the jurisdiction of the Active Implants Subcommittee of the National Technical Committee on Standardization of Surgical Implants and Orthopedic Devices.
This document was drafted by. Shanghai Institute for Medical Device Testing, Beijing Pinchi Medical Equipment Co., Ltd., and Nanhu Brain-Computer Interaction Research Institute.
Research Institute, Xuanwu Hospital of Capital Medical University, Tsinghua University, Hangzhou Jialiang Medical Technology Co., Ltd., Boruikang Technology (Shanghai) Co., Ltd.
Jingyu Medical Technology (Suzhou) Co., Ltd., Shenzhen Yinghe Brain Science Co., Ltd., and Changzhou Ruishenan Medical Devices Co., Ltd.
The main drafters of this document are. Hu Sheng, Liu Xun, Yu Hongyi, Chen Hao, Hao Hongwei, Wang Yueming, Xiong Botao, Zhao Guoguang, Shan Yongzhi, Wang Weiming, and Yu Huiling.
Cao Peng, Zeng Yan, Xu Honglai, Huang Xiaoshan, Zhu Weiran, Deng Wei, Min Xiaoyi, Xin Sheng, An Shuqing.
Medical devices using brain-computer interface technology
Implantable neurostimulator with closed-loop function
Sensing and Response Performance Testing Methods
1 Scope
This document describes an implantable neurostimulator with closed-loop functionality that utilizes brain-computer interface technology, based on the perception and response of neural signals.
Performance testing methods.
This document applies to implantable neurostimulators with closed-loop functionality that utilize brain-computer interface technology.
Note. For implantable neurostimulators that do not have closed-loop functionality, their ability to sense neural signals (if any) shall be tested in accordance with the relevant clauses of this document.
method.
2 Normative references
The contents of the following documents, through normative references within the text, constitute essential provisions of this document. Dated citations are listed below.
For references to documents, only the version corresponding to that date applies to this document; for undated references, the latest version (including all amendments) applies.
document.
YY/T 1987 Terminology for Medical Devices Employing Brain-Computer Interface Technology
3 Terms and Definitions
The terms and definitions defined in YY/T 1987 and the following terms and definitions apply to this document.
3.1
Brain-computer interface implanted neurostimulator
Brain-computer interface technology is used to implant medical devices that provide electrical stimulation to the central nervous system.
3.2
Algorithm performance
The efficiency and effectiveness of an algorithm when performing a specific task.
Note. This typically includes the time required for the algorithm to execute, the memory or storage space required for the algorithm to execute, and the correctness of the algorithm in solving the problem.
Rate or accuracy, stability and reliability of the algorithm when handling abnormal or noisy data, performance of the algorithm when handling datasets of different sizes, and computational efficiency.
The speed of response and processing capabilities of the method in real-time applications.
3.3
stimulation threshold
The critical value of electrical stimulation intensity that can control symptoms.
Note. Generally includes a lower threshold and an upper threshold. The lower threshold refers to the minimum electrical stimulation intensity that can control symptoms, while the upper threshold refers to the intensity that can control symptoms while stimulating related accessory pathways.
The maximum safe electrical stimulation intensity that minimizes the effect.
3.4
electrical stimulation strategy
Electrical stimulation parameters and methods are developed and applied during treatment to optimize therapeutic effects and patient comfort.
...