Basic data
| Standard ID | JJF 1889-2020 (JJF1889-2020) |
| Description (Translated English) | Calibration Specification for Medical Hard Copy Cameras |
| Sector / Industry | Metrology & Measurement Industry Standard |
| Classification of Chinese Standard | A60 |
| Classification of International Standard | 17.180 |
| Word Count Estimation | 18,142 |
| Date of Issue | 2020-11-26 |
| Date of Implementation | 2021-05-26 |
| Quoted Standard | JJF 1032-2005; GB/T 17006.3-2000; GB/T 17006.6-2003; GB/T 34988-2017; HG/T 4127-2017; YY/T 0610-2007; IEC 61223-2-4-1994; IEC TR 60788-2004; SMPTE RP-133-1991 |
| Regulation (derived from) | Announcement No. 57 (2020) of the State Administration for Market Regulation |
| Issuing agency(ies) | State Administration for Market Regulation |
| Summary | This standard applies to the calibration of medical hardcopy cameras (also called "medical film printers"), including dry laser cameras and thermal film printers. |
JJF 1889-2020: Calibration Specification for Medical Hard Copy Cameras
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Medical Hard Copy Cameras
National Metrology Technical Specifications of the People's Republic of China
Calibration specification for medical hardcopy cameras
2020-11-26 release
2021-05-26 implementation
Issued by the State Administration for Market Regulation
Medical hard copy camera
Calibration specification
Focal point: National Medical Metrology Technical Committee
Main drafting unit: China Institute of Metrology
Shandong Institute of Metrology
Participated in the drafting unit: Hebei Provincial People's Hospital
Carestream (Shanghai) Medical Equipment Co:, Ltd:
Agfa Medical System Equipment (Shanghai) Co:, Ltd:
This specification entrusts the National Medical Metrology Technical Committee to interpret
The main drafters of this specification:
Zhang Pu (China Institute of Metrology)
Li Chengwei (China Institute of Metrology)
Ding Qiang (Shandong Institute of Metrology)
Participating drafters:
Baoyunfeng (Hebei Provincial People's Hospital)
Liu Zilong (China Institute of Metrology)
Li Guodong [Reco (Shanghai) Medical Equipment Co:, Ltd:]
Chen Bin [Agfa Medical System Equipment (Shanghai) Co:, Ltd:]
table of Contents
Introduction (Ⅱ)
1 scope (1)
2 Reference documents (1)
3 Terms and measurement units (1)
3:1 Medical dry film (1)
3:2 Hardcopy camera (1)
3:3 Grayscale(1)
3:4 Film density (2)
3:5 Grayscale reproducibility (2)
3:6 Basic film (2)
3:7 Test film (2)
3:8 Image distortion (2)
3:9 Film viewer (2)
4 Overview (2)
5 Measurement characteristics (2)
5:1 Grayscale reproducibility (2)
5:2 Image distortion (2)
6 Calibration conditions (2)
6:1 Environmental conditions (2)
6:2 Measurement standards and other equipment (3)
7 Calibration items and calibration methods (3)
7:1 Appearance and functional inspection (3)
7:2 Grayscale reproducibility (3)
7:3 Image distortion (3)
8 Calibration result expression (5)
8:1 Calibration record (5)
8:2 Processing of calibration results (5)
9 Recalibration interval (6)
Appendix A Format of Original Record (Recommended) for Calibration of Medical Hardcopy Cameras (7)
Appendix B Internal page of calibration certificate (recommended) format (8)
Appendix C Example of Uncertainty Evaluation of Gray Scale Reproducibility Calibration (10)
Appendix D Example of Uncertainty Evaluation of Image Distortion Calibration (12)
Introduction
JJF 1071-2010 "Rules for the Compilation of National Metrology and Calibration Specifications", JJF 1001-2011 "General Metrology Terms and
Definition" and JJF 1059:1-2012 "Evaluation and Expression of Measurement Uncertainty" together constitute the support for the formulation of this specification
The basic series of documents:
This specification is released for the first time:
Calibration specification for medical hardcopy cameras
1 Scope
This specification applies to the calibration of medical hard copy cameras (also called "medical film printers"), including dry lasers
Optical camera and thermal film printer:
2 Reference documents
This specification refers to the following documents:
JJF 1032-2005 Terms and Definitions of Optical Radiation Metrology
GB/T 17006:3-2000 Evaluation and routine tests in medical imaging departments Part 2-2: X-ray photography
Shadow cassette and film changer screen-film contact and screen-cassette assembly relative sensitivity stability test
GB/T 17006:6-2003 Evaluation and Routine Tests in Medical Imaging Department Part 2-4: Hard Copy Photos
Camera stability test
GB/T 34988-2017 Information Technology General Specification for Toner Cartridges for Monochrome Laser Printers
HG/T 4127-2017 medical dry film
YY/T 0610-2007 General technical conditions for medical imaging photo observation equipment
IEC 61223-2-4::1994 Evaluation and routine testing in medical imaging departments Part 2-4: Hard copy photography
Machine stability test (Evaluationandroutinetestinginmedicalimagingdepartments-Part
2-4: Constancytests-Hardcopycameras)
IEC TR60788:2004 Medical electrical equipment definition term glossary (Medicalelectrical
equipment-Glossaryofdefinedterms)
SMPTERP-133:1991 TV monitor and screenshot recording camera medical diagnostic imaging test mode specification
Fan (Specificationsformedicaldiagnosticimagingtestpatternfortelevisionmonitors)
For dated reference documents, only the dated version applies to this specification; for undated references
The latest version (including all amendments) is applicable to this specification:
3 Terms and units of measurement
3:1 Medical dryfilm medicaldryfilm
Contains medical dry laser film (consisting of polyester film coated with silver salt and protective layer) and medical dry thermal film
(Composed of a polyester film base, a silver-containing heat-sensitive layer and a protective layer), it can be used as a diagnostic basis without the use of a developer:
Medical imaging (CT, MRI, CR, DR, etc:) records:
Note: "Medical dry film" is referred to as "film" in this specification:
3:2 Hardcopy camera hardcopycamera
A device that uses the input signal provided by the imaging system to produce an indelible image on the film:
3:3 grayscale
The visibility of brightness differences between adjacent grayscale blocks:
...