Path:
Home >
JJ >
Page5 > JJF 1641-2017
Price & Delivery
US$669.00 · In stock · Download in 9 secondsJJF 1641-2017: Program of Pattern Evaluation of Ultraviolet--Visible Spectrophotometers
Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See
step-by-step procedureStatus: Valid
| Std ID | Version | USD | Buy | Deliver [PDF] in | Title (Description) |
| JJF 1641-2017 | English | 669 |
Add to Cart
|
5 days [Need to translate]
|
Program of Pattern Evaluation of Ultraviolet--Visible Spectrophotometers
|
Click to Preview a similar PDF
Basic data
| Standard ID | JJF 1641-2017 (JJF1641-2017) |
| Description (Translated English) | Program of Pattern Evaluation of Ultraviolet--Visible Spectrophotometers |
| Sector / Industry | Metrology & Measurement Industry Standard |
| Classification of Chinese Standard | A60 |
| Classification of International Standard | 17.180 |
| Word Count Estimation | 29,297 |
| Date of Issue | 2017-09-26 |
| Date of Implementation | 2017-12-26 |
| Quoted Standard | JJG 178-2007; JJF 1001-2011; GB 4793.1-2007; GB/T 11606-2007 |
| Regulation (derived from) | General Administration of State Quality Inspection Announcement 2017 No. 78 |
| Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
| Summary | This standard applies to the new product type evaluation work in the wavelength range of 190 mm to 1 000 mm, with continuously adjustable wavelength spectrophotometer. |
JJF 1641-2017: Program of Pattern Evaluation of Ultraviolet--Visible Spectrophotometers
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
(UV - visible spectrophotometer type evaluation outline)
People's Republic of China National Measurement Specifications
UV-visible spectrophotometer type evaluation outline
Program of Pattern Evaluation
Ultraviolet-Visible Spectrophotometers
2017-09-26 Posted
2017-12-26 Implementation
General Administration of Quality Supervision, Inspection and Quarantine released
UV-visible spectrophotometer
Type evaluation outline
Program of Pattern Evaluation
Ultraviolet-Visible Spectrophotometers
The focal point. National Physical Chemistry Metering Technology Committee
The main drafting unit. China Institute of Metrology
Participate in the drafting unit. Beijing Tongzhou Weipu Technology Co., Ltd.
This specification entrusts the National Physical Chemistry Metrology Technical Committee responsible for the interpretation
The main drafters of this specification.
Li Lu Ming (China Institute of Metrology)
Participate in the drafters.
Sun Hongwei (Beijing Tongzhou Weipu Technology Co., Ltd.)
table of Contents
Introduction (Ⅲ)
1 Scope (1)
2 References (1)
3 Terms (1)
3.1 Spectrophotometer (1)
3.2 UV-visible spectrophotometer (1)
4 Overview (1)
5 Legal Management Requirements (1)
5.1 Unit of measure (1)
5.2 accuracy (maximum allowable error) requirements (1)
5.3 Marks (2)
6 measurement requirements (2)
6.1 Wavelength indication error (2)
6.2 wavelength repeatability (2)
6.3 transmittance indication error (2)
6.4 Transmission Repeatability (2)
6.5 Noise (2)
6.6 Drift (2)
6.7 Spectral bandwidth (2)
6.8 Baseline Flatness (2)
6.9 Stray light (2)
7 General technical requirements (3)
7.1 Appearance and preliminary examination (3)
7.2 Functional Requirements (3)
7.3 Low temperature test (3)
7.4 high temperature test (3)
7.5 Power supply voltage adaptability (3)
7.6 Crash test (3)
7.7 Drop test (3)
8 type evaluation item list (3)
9 to provide the number of prototypes and the use of the prototype (4)
9.1 provide the number of prototypes (4)
9.2 use of the prototype (4)
10 Test methods and conditions of test items and data processing and acceptance criteria (4)
10.1 Preparation before testing (4)
10.2 Appearance (5)
10.3 Wavelength indication error and wavelength repeatability (5)
10.4 transmittance error and repeatability (6)
10.5 Noise and Drift (7)
10.6 Spectral bandwidth (8)
10.7 Baseline Straightness (8)
10.8 Stray light (9)
10.9 Low temperature and high temperature adaptability (9)
10.10 Power supply voltage adaptability (10)
10.11 Crash test (10)
10.12 Drop Test (11)
11 Test items used in measuring instruments and equipment table (11)
Appendix A UV-visible spectrophotometer type evaluation original records (13)
Introduction
This type evaluation outline based on JJF 1016 "Guidelines for the preparation of evaluation criteria for measuring instruments", JJF 1015 "meter
Gage Type Evaluation and Type Approval General Specification "and JJF 1001" General Measurement Terms and Definitions ".
The technical indicators of this type evaluation outline reference JJG178-2007 "UV, visible, near-infrared spectrophotometry
Meter "and GB/T 11606-2007" analytical instruments environmental test methods "and other technical regulations.
This type of evaluation outline for the first time released.
UV-visible spectrophotometer type evaluation outline
1 Scope
This type of evaluation program for the wavelength range of 190nm ~ 1000nm, continuously adjustable wavelength of the spectrophotometer
New product type evaluation work.
2 reference documents
This specification cites the following documents.
JJG178-2007 "UV, visible, near-infrared spectrophotometer"
JJF 1001-2011 "Common measurement terms and definitions"
GB 4793.1-2007 "Safety requirements for electrical equipment for measurement, control and laboratory use"
GB/T 11606-2007 "analytical instruments environmental test methods"
For dated references, only the dated version applies to this specification; for undated references
The latest version (including all amendments) applies to this specification.
3 Terms
3.1 spectrophotometer spectrophotometer
Based on Lambert-Beer's law, the theory used for colorimetric analysis.
3.2 UV-visible spectrophotometer ultraviolet-visiblespectrophotometer
The wavelength range of 190nm ~ 1000nm, wavelength continuously adjustable spectrophotometer.
4 Overview
UV - Vis spectrophotometer (hereinafter referred to as the instrument) is based on the material of UV - visible light radiation of choice
Absorption and Lambert-Beer Laws Quantitative analysis of substances and qualitative identification of the instrument.
The mathematical expression of Lambert-Beer's law is.
A = -lg (I/Io) = - lgT = klc
The physical meaning is that when a bundle of parallel monochromatic light perpendicular to the measured substance solution, the solution absorbance A
And the concentration of light-absorbing material c and liquid layer thickness l is proportional to.
The instrument is mainly composed of light source, monochromator, sample chamber, detector, signal processing and display and storage system.
5 legal management requirements
5.1 unit of measurement
UV - visible spectrophotometer using legal units of measurement.
5.2 accuracy (maximum allowable error) requirements
The accuracy of the instrument shall be in accordance with JJG178-2007 "UV, VIS, NIR Spectrophotometer"
Provisions.
...