Calibration Specification for Visual Electrophysiological Instruments
|
Click to Preview a similar PDF
Basic data
| Standard ID | JJF 1543-2015 (JJF1543-2015) |
| Description (Translated English) | Calibration Specification for Visual Electrophysiological Instruments |
| Sector / Industry | Metrology & Measurement Industry Standard |
| Classification of Chinese Standard | A60 |
| Classification of International Standard | 17.180 |
| Word Count Estimation | 28,274 |
| Date of Issue | 2015-08-24 |
| Date of Implementation | 2015-11-24 |
| Quoted Standard | ISCEV(2003) |
| Regulation (derived from) | AQSIQ Announcement 2015 No.106 |
| Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
| Summary | This standard applies to the clinical use of visual electrophysiology instrument calibration. |
JJF 1543-2015: Calibration Specification for Visual Electrophysiological Instruments
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Visual Electrophysiological Instruments
People's Republic of China National Metrology Technical Specifications
Calibration Specification for visual electrophysiology
Issued on. 2015-08-24
2015-11-24 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Calibration Specification for visual electrophysiology
Focal point. the National Technical Committee on Medical Measurement
The main drafting units. China Institute of Metrology
Participated in the drafting. Special Medical Equipment Co., Chongqing, China
Shandong Province Institute of Metrology
Chinese Academy of Sciences Institute of Optical Technology
This specification entrusted to the National Technical Committee is responsible for interpreting medical measurement
The main drafters of this specification.
Set Xiang (China Institute of Metrology)
HE Yu (Chongqing, China Special Medical Equipment Co., Ltd.)
Zhang Ji Yan (China Institute of Metrology)
Drafters participate.
Sun Xin (Shandong Province Institute of Metrology)
Bai Wei (Chongqing States special medical equipment Limited)
Shi Guohua (Institute of Optoelectronic Technology Institute)
table of Contents
Introduction (Ⅲ)
1 Scope (1)
2. References (1)
3 terminology and units of measurement (1)
3.1 average luminance pattern (1)
3.2 Graphics contrast (1)
3.3 brightness uniformity (1)
3.4 flash intensity (1)
3.5 flash duration (1)
Latency 3.6 (1)
4 Overview (1)
5 Metrological characteristics (1)
5.1 flash intensity setting error (1)
5.2 flash duration (2)
5.3 flash stimulator background brightness setting error (2)
5.4 Graphics Contrast (2)
5.5 Graphics average brightness setting value error (2)
5.6 Graphics stimulator luminance uniformity (2)
5.7 shows the magnitude of the electrical signal error (2)
5.8 electrical signal frequency indication error (2)
5.9 latency indication error (2)
Calibration Condition 6 (2)
6.1 environmental conditions (2)
6.2 standards and other measuring equipment (2)
7 calibration items and calibration methods (3)
7.1 appearance and function check (3)
7.2 flash intensity setpoint error (3)
7.3 flash duration (3)
7.4 flash stimulator background brightness setting error (3)
7.5 Graphics Contrast (3)
7.6 Graphics average brightness setting error (4)
7.7 Graphics stimulator luminance uniformity (4)
7.8 shows the magnitude of the electrical signal error (4)
7.9 Indication error frequency signals (5)
7.10 latency indication error (5)
8 calibration results expression (6)
8.1 Calibration records (6)
8.2 Calibration process results (6)
9 Recalibration interval (6)
Appendix A visual electrophysiology instrument calibration original records (recommended) formatting styles (7)
Appendix B calibration certificate within the page (recommended) formatting styles (10)
Appendix C Measurement Uncertainty Example (14)
Introduction
This specification is for the measurement of technical regulations visual electrophysiology instrument calibration making. Preparation of this specification to JJF 1071-2010
"National Calibration Specifications writing rules", JJF 1001-2011 "common measurement terms and definitions"
JJF 1059.1-2012 "Evaluation and Expression of Uncertainty in Measurement" as the basis and foundation, with reference to the ISCEV (2003)
"Clinical Visual electrophysiological stimulation and recording parameters Calibration Guide" (Guidelinesforcalibrationofstimulusand
recordingparametersusedinclinicalelectrophysiologyofvision) regulations.
This specification is the first release.
Calibration Specification for visual electrophysiology
1 Scope
This instruction applies to visual electrophysiology instrument calibration clinical use.
2. References
This specification refers to the following documents.
ISCEV (2003) "clinical visual electrophysiological stimulation and recording parameters Calibration Guide" (Guidelinesforcali-
brationofstimulusandrecordingparametersusedinclinicalelectrophysiologyofvision)
For dated references, only the dated edition applies to this specification; undated reference documents
Member, the latest edition (including any amendments) applies to this specification.
3 terminology and units of measurement
3.1 Graphics average luminance meanluminanceofpattern
The average of the graphic form of the highest brightness and the lowest brightness, unit. candela per square meter, symbol. cd/m2.
3.2 Graphics contrast patterncontrast
Drawing light grid brightness and brightness contrast dark grid, expressed as a percentage.
3.3 brightness uniformity luminanceuniformity
Graphics stimulator full brightness uniformity of display when expressed as a percentage.
3.4 flash intensity flashintensity
Instantaneous flash of brightness integrated value time unit. seconds candela per square meter, symbol. cd · s/m2.
3.5 flash duration flashduration
Single flash duration, unit. milliseconds symbol. ms.
3.6 latency latency
Eye irritation from the received light signal to an electrical signal characteristic between the time delays, unit. milliseconds symbol. ms.
4 Overview
Visual electrophysiology instrument is one kind of electrophysiological signals measured by the human visual system after light stimuli to diagnose
Instrument of visual function.
Visual electrophysiology instrument to the human eye to light stimulation signal emitted by the stimulator while collecting electrode by the human visual system
Biological signals generated by the electrical signal amplification and processing after recording and display, providing for clinical visual function
Medical diagnosis. Visual electrophysiology instrument by the flash stimulation, graphics stimulator, electrode, signal acquisition, data processing
Operating system and the display system.
5 Metrological characteristics
5.1 flash intensity settings error
No more than ± 10%.
...