Path:
Home >
GB/T >
Page169 > GB/T 28851-2012
Price & Delivery
US$399.00 · In stock · Download in 9 secondsGB/T 28851-2012: Specifications for refrigerated incubators
Delivery: 9 seconds. True-PDF full-copy in English & invoice will be downloaded + auto-delivered via email. See
step-by-step procedureStatus: Valid
| Std ID | Version | USD | Buy | Deliver [PDF] in | Title (Description) |
| GB/T 28851-2012 | English | 399 |
Add to Cart
|
3 days [Need to translate]
|
Specifications for refrigerated incubators
|
Click to Preview a similar PDF
Basic data
| Standard ID | GB/T 28851-2012 (GB/T28851-2012) |
| Description (Translated English) | Specifications for refrigerated incubators |
| Sector / Industry | National Standard (Recommended) |
| Classification of Chinese Standard | N61 |
| Classification of International Standard | 19.040 |
| Word Count Estimation | 17,172 |
| Quoted Standard | GB/T 191-2008; GB/T 2829-2002; GB 4793.6-2008; GB 4793.1-2007; GB/T 9969-2008; GB/T 10592-2008; GB/T 17248.3-1999; JB/T 9512 - 1999; GB/T 11020-2005 |
| Regulation (derived from) | National Standards Bulletin 2012 No. 28 |
| Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China |
| Summary | This standard specifies the conditions of use of the incubator, requirements, test methods, inspection rules, marking, packaging, transportation, storage and accompanying documents. This standard applies to the active refrigeration equipment installed in |
GB/T 28851-2012: Specifications for refrigerated incubators
---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Specifications for refrigerated incubators
ICS 19.040
N61
National Standards of People's Republic of China
Incubator technical conditions
Issued on. 2012-11-05
2013-02-15 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China released
Table of Contents
Introduction Ⅲ
1 Scope 1
2 Normative references 1
3 1 Conditions of Use
3.1 Environmental Conditions 1
3.2 Supply Condition 1
3.3 Load Condition 2
4 requires 2
4.1 appearance and structure 2
4.2 Technical Performance 2
4.3 Safety and environmental protection 3
Test Method 4 5
5.1 Main test instruments and equipment 4
5.2 Test conditions 4
5.3 various tests to determine the number and position 4
5.4 Test appearance and structure 4
5.5 Operating temperature range 5 Test
5.6 Temperature fluctuation test 6
5.7 Temperature Uniformity Test 6
5.8 Cooling Time Trial 6
5.9 Heating time Test 7
Safety and environmental protection Test 7 5.10
6 Inspection rules 8
6.1 Inspection classification 8
6.2 Test Project 8
6.3 factory inspection 9
6.4 Inspection stereotypes 9
6.5 Periodic Inspection 9
7 logo 10
7.1 General 10
7.2 Product nameplate 11
7.3 Operations with 11 flags
7.4 Safety signs 11
7.5 Package Marking 11
8 packaging, transportation and storage 11
8.1 Packaging 11
8.2 Transport 12
8.3 Storage 12
9 accompanying documents 12
9.1 Technical Document 12
9.2 Other documents 12
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules.
The standard proposed by China Machinery Industry Federation.
The standard by the China Machinery Industry Federation.
This standard is drafted by. Southern Jiangsu Wuxi Test Equipment Co., Ltd., Hangzhou snow carbon thermostat Technology Co., Ltd., Chongqing Sida test
Equipment Co., Ltd., Zhejiang Institute of Metrology, Machinery Industrial Instrumentation Technology and Economy Institute, Ministry of Electronics Industry and Information Technology
Fifth Institute of Hunan Institute of Metrology detection on 海爱斯佩克 Environmental Equipment Co., Ltd.
Participated in the drafting of this standard. Anhui Institute of Metrology, Hengyang Heng Instrument Electric Co., Ltd., Shenyang Academy of Instrumentation Science, China
Electric Apparatus Research Institute, Ningbo Dongfang Heating Equipment Co., Ltd., Shanghai Jing Hong Experimental Equipment Co., Ltd., Chongqing Yinhe Experimental Instrument Co.,
the company.
The main drafters of this standard. Xu Yueming, Ni Yiming, Chen Yunsheng, Zhou Lianqin, Jin Lihui, Zousu Yang, Xiong Zhi-ming, Feng Hua, Jin Meifeng,
Liu Xiangheng, Xu Qiuling, Zhouxiu Yuan, Wang Xiaofeng, Lim Chun Wing, Li Yaqin, Wang Jialong, only Shen Zhong, He Jian Gang, Wang Huabin.
Incubator technical conditions
1 Scope
This standard specifies the incubator conditions of use, requirements, test methods, inspection rules, marking, packaging, transportation, storage and accompanying
file.
This standard applies to devices equipped with active cooling, in order to save the sample incubation and incubator for the purpose (hereinafter referred to as biochemical
box).
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 191-2008 Packaging - Pictorial signs
GB/T 2829-2002 periodic inspection Sampling procedures and tables (suitable for process stability test)
GB 4793.1-2007 measurement, control and laboratory safety requirements for electrical equipment - Part 1. General requirements
GB 4793.6-2008 measurement, control and laboratory safety requirements for electrical equipment - Part 6. Laboratory material heating apparatus
Particular requirements for equipment
GB/T 9969-2008 General Industrial Product Manual
GB/T 10592-2008 high and low temperature test chambers
Listing flammability test method GB/T 11020-2005 solid non-metallic materials when exposed to flame sources
GB/T 17248.3-1999 Acoustics - Noise emitted by machinery and equipment and other operating positions specified position emission sound pressure levels measured
Volume method in situ
Determination of JB/T 9512-1999 climate chamber test equipment and sound power levels
3 Conditions of Use
3.1 Environmental conditions
Biochemical box should be under the following conditions.
a) indoor use;
b) a temperature of 5 ℃ ~ 35 ℃, no drastic changes in ambient temperature;
c) when the ambient temperature is below 31 ℃ maximum relative humidity of 80% and an ambient temperature of 35 ℃, the maximum relative humidity linear drop
To 67%;
d) altitude less than 2000m;
e) There should be around the biochemical effects of vibration box use, the magnetic field (with the exception of the geomagnetic field), and other hot and cold radiation;
f) Pollution Level 2, around without high concentrations of dust or corrosive gases;
g) airflow around without a strong, strong demand when the ambient air flow, air flow should not be blown directly biochemistry of the refrigeration system and air inlet door
around.
3.2 Supply conditions
Biochemical box power supply should meet the following conditions.
...