HOME   Cart(0)   Quotation   About-Us Policy PDFs Standard-List
www.ChineseStandard.net Database: 189760 (18 Oct 2025)

YY/T 0119.5-2014 English PDF

US$259.00 ยท In stock
Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email.
YY/T 0119.5-2014: Spinal implants. Components used in the surgical fixation of the spinal skeletal system. Part 5: Test method for measure the static and fatigue bending strength of metallic screws
Status: Valid
Standard IDContents [version]USDSTEP2[PDF] delivered inStandard Title (Description)StatusPDF
YY/T 0119.5-2014English259 Add to Cart 3 days [Need to translate] Spinal implants. Components used in the surgical fixation of the spinal skeletal system. Part 5: Test method for measure the static and fatigue bending strength of metallic screws Valid YY/T 0119.5-2014

PDF similar to YY/T 0119.5-2014


Standard similar to YY/T 0119.5-2014

YY 0118   YY/T 0119.4   YY/T 0119.1   YY 0117.1   YY 0117.2   YY/T 0119.2   

Basic data

Standard ID YY/T 0119.5-2014 (YY/T0119.5-2014)
Description (Translated English) Spinal implants. Components used in the surgical fixation of the spinal skeletal system. Part 5: Test method for measure the static and fatigue bending strength of metallic screws
Sector / Industry Medical Device & Pharmaceutical Industry Standard (Recommended)
Classification of Chinese Standard C35
Classification of International Standard 11.040.40
Word Count Estimation 11,184
Date of Issue 6/17/2014
Date of Implementation 7/1/2015
Quoted Standard GB/T 16825.1; GB/T 19701.2; ASTM E122; ASTM E467; ASTM E1942; ASTM F1839
Regulation (derived from) China Food and Drug Administration in 2014 Bulletin No. 30
Issuing agency(ies) State Food and Drug Administration
Summary This Standard applies to internal spinal fixation implant anchoring element metal spine screws. This Standard specifies the test method for the determination of metal spine screws static and dynamic bending properties.

YY/T 0119.5-2014: Spinal implants. Components used in the surgical fixation of the spinal skeletal system. Part 5: Test method for measure the static and fatigue bending strength of metallic screws



---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Spinal implants.Components used in the surgical fixation of the spinal skeletal system.Part 5. Test method for measure the static and fatigue bending strength of metallic screws ICS 11.040.40 C35 Medical device industry standards People's Republic of China Spinal implants come within the spine system components Part 5. Metal screw spinal static and fatigue bending Strength Test Method system-Part 5. Testmethodformeasurethestaticandfatiguebendingstrength ofmetalicscrews Issued on. 2014-06-17 2015-07-01 implementation China Food and Drug Administration released

Foreword

YY/T 0119 "was within the spinal implant spinal fixation system components" is divided into five sections. --- Part 1. General requirements; --- Part 2. Metal screws spine; --- Part 3. Metal spine board; --- Part 4. Metal spinal rod; --- Part 5. Metal screw spinal static and fatigue bending strength Test method. This section YY/T Section 50119 of. This section drafted in accordance with GB/T 1.1-2009 given rules. Used in this section refer to redraft the law ASTMF2193-07 "standard spinal fixation system components and test methods" in Appendix A4 And X5 establishment. Please note that some of the content of this document may involve patents. Release mechanism of the present document does not assume responsibility for the identification of these patents. This part is proposed by the China Food and Drug Administration. This part of the National Surgical implants and orthopedic instruments Standardization Technical Committee materials and orthopedic implants Technical Committee (SAC / TC110/SC1) centralized. This section is drafted. Tianjin Medical Device Quality Supervision and Inspection Center, Shanghai Medical Technology Co., Ltd. minimally invasive orthopedics. The main drafters of this section. Zhang said, Jiang Xi, Zhang Chen, Zhang Lu, Wang Guohui, Xuezong Yu. Spinal implants come within the spine system components Part 5. Metal screw spinal static and fatigue bending Strength Test Method

1 Scope

YY/T 0119 provisions of this part of the test method for the determination of metal screws spine static and dynamic bending properties. This section applies to the spinal fixation implants in the spine metal screws. This section is intended to provide a means for different types of spinal screw the mechanical characterization methods. There is currently not enough pre-knowledge Use screws to know the results of specific design, it is not the purpose of this test method is used to define the level of performance of spinal screw.

2 Normative references

The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document. Inspection and GB/T 16825.1 of static uniaxial testing machines - Part 1 Rally and (or) pressure testing machine force measuring system calibration GB/T 19701.2 ultra-high molecular weight polyethylene - Part 2. plastic molding ASTME122 using specific tolerances, the sample size to assess the production batch or process characteristics mean calculation specifications (Practicefor calculatingsamplesizetoestimate, withaspecifiedtolerableerror, theaverageforacharacteristicofa lotorprocess) ASTME467 Axial Fatigue Test System for constant amplitude dynamic loads calibration specification (Standardpracticeforverification ofconstantamplitudedynamicforcesinanaxialfatiguetestingsystem) ASTME1942 cycle fatigue and fracture mechanics test data acquisition system evaluation guidelines (Guideforevaluatingdataac- quisitionsystemsusedincyclicfatigueandfracturemechanicstesting) ASTMF1839 orthopedic instruments and tools to test using standard materials - rigid polyurethane foam (Specificaionforrigidpolyure- thanefoamforuseasastandardmaterialfortestingorthopedicdevicesandinstruments)

3 Summary of Test Method

3.1 pairs of spinal screw quasi-static cantilever bending load is applied, so that the recording member in response to applied mechanical load corresponding. By Loads - Deflection curves were used to determine the data for deriving the spinal screw flexural properties. 3.2 pairs of sinusoidal cantilever sample is applied to a given predetermined frequency spinal screw bending loads. Until one of the following three conditions Fatigue test. samples Product failure, limit or terminate the trial reaches a predetermined number of cycles (termination limit). The test results information is used to draw the moment - cycle times Curve to characterize a given moment in the spinal screw type range applied general fatigue properties. Further, in the case where the result of the appropriate Information can be used to determine the spinal screw N at a predetermined number of cycles in the value of fatigue moment.

4 significance and application

4.1 part of the establishment of a unified method for static and dynamic bending test, and thus different types of spinal screw static and fatigue bending properties Properties were characterized and compared. This test method is used to determine the maximum bending moment in the spinal screw range conditions fatigue life, in addition, it may also be

Tips & Frequently Asked Questions:

Question 1: How long will the true-PDF of YY/T 0119.5-2014_English be delivered?

Answer: Upon your order, we will start to translate YY/T 0119.5-2014_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time.

Question 2: Can I share the purchased PDF of YY/T 0119.5-2014_English with my colleagues?

Answer: Yes. The purchased PDF of YY/T 0119.5-2014_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet.

Question 3: Does the price include tax/VAT?

Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countries

Question 4: Do you accept my currency other than USD?

Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.