US$509.00 · In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 755-2015: Ultraviolet Irradiance Working Standard Apparatus Status: Valid JJG 755: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJG 755-2015 | English | 509 |
Add to Cart
|
4 days [Need to translate]
|
Ultraviolet Irradiance Working Standard Apparatus
| Valid |
JJG 755-2015
|
JJG 755-1991 | English | 199 |
Add to Cart
|
2 days [Need to translate]
|
Verification Reguiation of Working Standard Equipment for UV Irradiance
| Obsolete |
JJG 755-1991
|
PDF similar to JJG 755-2015
Basic data Standard ID | JJG 755-2015 (JJG755-2015) | Description (Translated English) | Ultraviolet Irradiance Working Standard Apparatus | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A60 | Classification of International Standard | 17.180 | Word Count Estimation | 22,258 | Date of Issue | 2015-12-07 | Date of Implementation | 2016-06-07 | Older Standard (superseded by this standard) | JJG 755-1991 | Quoted Standard | JJG 384 | Regulation (derived from) | ?AQSIQ 2015 Nian (No.148) | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard is applicable to the first verification, follow-up verification and inspection during operation of the UV-irradiance reference device. |
JJG 755-2015: Ultraviolet Irradiance Working Standard Apparatus---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Ultraviolet Irradiance Working Standard Apparatus
National Metrological Verification Regulation of the People's Republic of China
UV radiation illumination work reference device
2015-12-07 released
2016-06-07 implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
UV radiation illumination work reference device
Verification procedures
WorkingStandard Appliances
Replacing JJG 755-1991
Coordinate unit. National Committee for Optical Metrology
The main drafting unit. China Institute of Metrology
Beijing Normal University
Participated in the drafting unit. Beijing Institute of Metrology and Testing
This procedure entrusts the National Committee for Optical Metrology to explain
The main drafters of this procedure.
Generation color red (China Institute of Metrology)
Wu Zhifeng (China Institute of Metrology)
Zhang Baozhou (Beijing Normal University)
Participate in the drafters.
Wang Yanfei (China Institute of Metrology)
Zhang qingxian (Beijing Institute of Metrology and Testing Science)
table of Contents
Introduction (II)
1 Scope (1)
2 References (1)
3 Overview (1)
4 Measurement performance requirements (1)
4.1 Spectral response and band division of the device (1)
4.2 Annual change rate of the device (2)
4.3 Operating Baseline Ultraviolet Radiation Illuminance Zero Error (Full Scale FS) (2)
4.4 working reference UV radiation illuminator long wave response error (2)
4.5 Working Baseline Ultraviolet radiation Illuminance meter cosine characteristics (directional response) (2)
4.6 Working standard Non-linearity error of ultraviolet radiation illuminometer (2)
4.7 Work reference UV radiation illuminometer shift error (2)
4.8 working reference UV radiation illuminance fatigue error (2)
4.9 working reference UV radiation illuminator maximum allowable relative error (2)
5 General technical requirements (2)
6 Metering device control (3)
6.1 Test conditions (3)
6.2 Test items (4)
6.3 Test method (5)
6.4 Processing of test results (10)
6.5 Test cycle (10)
Appendix A Certificate of Verification and Verification Results Page Format (11)
Appendix B Verification of the original record (12)
Appendix C Working Basics UV Radiation Illuminance Meter Measurement Uncertainty Evaluation Example (UV-A Band) (14)
Introduction
JJF1001 "General Metrology Terms and Definitions", JJF1032 "Terminology and Definitions of Optical Radiometry"
JJF1059.1 "Measurement Uncertainty Assessment and Representation" and JJF1002 "National Metrological Verification Rules"
Constitute the basic norms that support the revision of this Code.
Compared with JJG 755-1991, in addition to editorial changes, the main technical changes are as follows.
--- Remove the "SUV-1" in the new procedure, change the UV radiation illuminometer to "working reference UV radiation illuminometer"
(See 3);
- on the basis of preserving the four bands of the original procedure (the UVB band of the original procedure is named in this procedure
UV-310, UVA2 band named UV-365, UVC band named UV-254), the new International Lighting Commission
The three bands proposed by CIE are. UV-A, UV-B and UV-C (see 4.1);
--- Modification of the composition of the UV spectrum radiation illumination comparison device structure diagram (see 6.1.1.3
image 3);
--- change the UV radiation intensity of the work of the reference device annual change rate, the working standard UV radiation illuminator zero
Value error, long wave response error, cosine characteristic (directional response) error, nonlinear error, shift error, fatigue
Error, relative indication error and other technical indicators (see 4.2 ~ 4.9) and the test method (see 6.3.3 ~
6.3.10);
- An example of measurement uncertainty assessment has been modified (see Appendix C).
The previous release of this order is.
--- JJG 755-1991.
Verification procedure for the reference device of ultraviolet radiation
1 Scope
This procedure applies to the first test of the UV radiation exposure reference device, subsequent verification and inspection.
2 reference file
This procedure refers to the following documents.
JJG 384 Spectral Radiance Illumination Standard Lamp
For dated references, only the dated edition applies to this specification; references that are not dated
, The latest version (including all modifications) applies to this specification.
3 Overview
International Council on Illumination CIE will be ultraviolet radiation is divided into UV-A (315nm ~ 400nm), UV-B
(280nm ~ 315nm) and UV-C (100nm ~ 280nm) three bands. Due to 100nm ~.200nm
UV radiation is strongly absorbed in the air, so for the UV-C band, this procedure only consider.200nm ~ 280nm
Wavelength range.
UV Radiation Illumination Work Benchmark Each band consists of three working reference UV radiation illuminometers. jobs
The standard UV radiation illuminance meter is the transmission of UV-A, UV-B, UV-C, UV-A1, UV-365, UV-
UV-254 and other wavelengths of ultraviolet radiation standard value of the standard instruments, mainly by the detector, amplifier circuit and display instruments
composition. Where the detector is typically composed of a photodetector, a filter (band-pass glass or interference filter), and a diffuser group
to make. Figure 1 is a schematic diagram of the general composition of the operating reference UV radiation illuminometer detector, whose magnitude is measured by spectral
Radiometric comparison of alternative methods Trace to spectral radiance Illumination national benchmarks.
Figure 1 Schematic diagram of the general composition of the working reference ultraviolet radiation illuminance detector
4 Measurement performance requirements
4.1 Spectral response and band division of the device
UV-A band. 315nm ~ 400nm, peak wavelength λp = 365nm ± 5nm
UV-B band. 280nm ~ 315nm, peak wavelength λp = 297nm ± 5nm
UV-C band..200nm ~ 280nm, peak wavelength λp = 254nm ± 5nm
UV-A1 band. 320nm ~ 390nm, peak wavelength λp = 365nm ± 5nm
UV-365 band. peak wavelength λp = 365nm ± 2nm, peak half height width Δλ ≤ 10nm
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 755-2015_English be delivered?Answer: Upon your order, we will start to translate JJG 755-2015_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 755-2015_English with my colleagues?Answer: Yes. The purchased PDF of JJG 755-2015_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJG 755-2015?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJG 755-2015 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|