US$439.00 ยท In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. JJG 581-2016: Lasers for Medicine Status: Valid JJG 581: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
JJG 581-2016 | English | 439 |
Add to Cart
|
4 days [Need to translate]
|
Lasers for Medicine
| Valid |
JJG 581-2016
|
JJG 581-1999 | English | 959 |
Add to Cart
|
5 days [Need to translate]
|
Verification regulation of laser for medicine
| Obsolete |
JJG 581-1999
|
JJG 581-1988 | English | RFQ |
ASK
|
3 days [Need to translate]
|
(Chinese Industry Standard)
| Obsolete |
JJG 581-1988
|
PDF similar to JJG 581-2016
Basic data Standard ID | JJG 581-2016 (JJG581-2016) | Description (Translated English) | Lasers for Medicine | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A60 | Word Count Estimation | 19,149 | Date of Issue | 2016-11-30 | Date of Implementation | 2017-05-30 | Older Standard (superseded by this standard) | JJG 581-1999 | Regulation (derived from) | State Administration of Quality Supervision, Inspection and Quarantine Notice No.119 of 2016 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine |
JJG 581-2016: Lasers for Medicine---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Lasers for Medicine
National Metrological Verification Regulation of the People's Republic of China
Medical laser source
2016-11-30 release
2017-05-30 implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
Medical laser source verification procedures
Replacing JJG 581-1999
Coordinate unit. National Committee for Optical Metrology
The main drafting unit. Beijing Institute of Metrology and Testing
China Institute of Metrology
Shanghai Institute of Metrology and Testing Technology
Participated in the drafting unit. Chinese People's Liberation Army Medical Laser Measurement Test Research Station
China Institute of Standardization
This procedure entrusts the National Committee for Optical Metrology to explain
The main drafters of this procedure.
Zhang jianliang (Beijing Institute of Metrology and Measurement Science)
Ma Chong (China Institute of Metrology)
Xia Ming (Shanghai Institute of Metrology and Testing Technology)
Participate in the drafters.
Yang Zaifu (People's Liberation Army Medical Laser Measurement and Research Station)
Cai Jianqi (China Institute of Standardization)
Sun Xianglin (Beijing Institute of Metrology and Measurement Science)
table of Contents
Introduction (II)
1 Scope (1)
2 References (1)
3 Terminology (1)
4 Overview (1)
5 Measurement performance requirements (1)
5.1 Indication of optical power measurement (1)
5.2 Repeatability (1)
5.3 Reproducibility (2)
5.4 Relative value of indication (2)
6 General technical requirements (2)
6.1 logo (2)
6.2 Appearance (2)
6.3 Display device (2)
6.4 Radiation warning signal (2)
6.5 Emergency switch (2)
6.6 Light guide system (2)
7 Control and verification of measuring instruments (2)
7.1 Test conditions (2)
7.2 Verification equipment (2)
7.3 Verification Project (3)
7.4 Verification method (3)
8 Test results and test cycle (6)
8.1 Processing of test results (6)
8.2 Test cycle (6)
Appendix A Certificate of Verification and Verification Results (End page) Format (7)
Appendix B Verification of the original record (inside page) Format (8)
Appendix C Commonly used laser wavelength table (9)
Appendix D Classification of laser sources (10)
Appendix E Examples of Medical Laser Source Uncertainty Analysis (12)
Introduction
This procedure is based on JJF1002-2010 "national measurement procedures for the preparation of rules", with reference to JJF1059.1-
"Measurement Uncertainty Assessment and Presentation", GB 9706.20-2000 "Medical electrical equipment Part 2. Diagnostics
And safety requirements for the treatment of laser equipment ", GB 11748-2005" carbon dioxide laser treatment machine ", GB 12257-
2000 "helium-neon laser treatment machine general technical conditions", IEC 60825-2007 "laser product safety", YY0307-
2011 "continuous wave doped neodymium yttrium aluminum garnet laser treatment machine", and according to the actual test of the problem, the JJG 581-1999
"Medical laser source" was partially revised.
--- to expand the scope of the test range, the detection range for the continuous laser source and pulsed laser source.
--- Adjusted references to references and terminology.
- According to the current development of technology, the original process outlined in the contents of a partial revision.
- Partial changes to the technical requirements based on the actual situation of the verification. deletion of the cooling system and leakage
Safety requirements (the test can not be effectively achieved), modified the light and light guide system technical requirements.
--- According to the medical laser source detection of the actual situation, in the measurement performance to increase the value of the relative error detection
project.
--- According to the actual situation of the test, the stability and reproducibility of the content to do a certain change.
- Leakage current detector, reading microscope and displacement movement mechanism are removed from the verification equipment (no actual detection
Method to achieve or through other equipment to complete), an increase of laser energy measurement device.
--- adjust the list of test items.
--- modify the instructions of the light test method, the light guide system to do the detection and merger.
--- Remove the light guide system reflectance, transmission ratio of the detection method (the actual work can not be on the medical laser source
Light guide system for demolition detection).
- A description of the supplementary notes and cautions has been added to the verification method.
--- on the results of the test results and verification of the contents of the cycle to do a certain change, the main increase "part of the qualified
The request ".
- Added an uncertainty analysis in the appendix, modified the template for verifying the original record and the certificate, deleted the
Measurement schematics, special equipment structure (the test can not be removed light guide system, the actual test of this project has been deleted),
warning sign.
Medical laser source verification procedures
1 Scope
This procedure applies to the medical laser source (not including medical excimer laser treatment machine) the first test, follow-up test
And the use of the inspection.
2 reference file
GB 9706.20-2000 Medical electrical equipment - Part 2. Diagnostic and treatment of laser equipment for safety
Claim
Carbon dioxide laser therapy machine
General specification for He - Ne laser therapy machine GB 12257-2000
IEC 60825-1..2007 Safety of laser products (Saftyoflaserproducts)
YY0307-2011 continuous wave doped neodymium yttrium aluminum garnet laser treatment machine
For dated references, only the dated edition applies to this rule; references that are not dated
, The latest version (including all amendments) applies to this procedure.
3 terminology
3.1 continuous laser source continuouswavelaser
A laser source that can be continuously emitted at a time greater than or equal to 0.25 s.
3.2 pulsed laser source pulselaser
A laser source that emits energy in the form of a single pulse or sequence pulse. The duration of a pulse is shorter than 0.25s.
4 Overview
The medical laser source is usually composed of a laser, a control section and an output section. The role of the laser part is to generate excitement
Light, different laser working materials will produce different wavelengths of laser (see Appendix C), can achieve different therapeutic purposes.
Laser control part of the main role is to adjust the medical laser source output laser energy, energy density, power, frequency, etc.
Parameters and safety control of medical laser sources. The laser output part of the role is to make the output laser light to a specified location, one
It consists of light guide system (including fiber). According to the different forms of output laser, can be divided into medical continuous laser source and medical
With pulse laser source two categories.
5 Measurement performance requirements
5.1 Indicates optical power measurement
Indicating that the light spot should be clearly visible, the power size, eye medical laser source does not exceed 1mW, other medical
Light source does not exceed 5mW.
5.2 Repeatability
5.2.1 For medical continuous laser sources, measuring the repeatability of the output laser power for 10 min of continuous exposure should be better than
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 581-2016_English be delivered?Answer: Upon your order, we will start to translate JJG 581-2016_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 581-2016_English with my colleagues?Answer: Yes. The purchased PDF of JJG 581-2016_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version JJG 581-2016?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version JJG 581-2016 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|