Verification Regulation of Medical Syringe Pump and Infusion Pump Analyzers
| Valid |
JJG 1098-2014
|
PDF similar to JJG 1098-2014
Basic data | Standard ID | JJG 1098-2014 (JJG1098-2014) | | Description (Translated English) | Verification Regulation of Medical Syringe Pump and Infusion Pump Analyzers | | Sector / Industry | Metrology & Measurement Industry Standard | | Classification of Chinese Standard | A50 | | Classification of International Standard | 17.020 | | Word Count Estimation | 20,241 | | Date of Issue | 8/1/2014 | | Date of Implementation | 11/1/2014 | | Quoted Standard | JJF 1259-2010; GB/T 6682-2008 | | Regulation (derived from) | AQSIQ Notice 2014 No. 89 | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | | Summary | This Standard applies to medical syringe pump and infusion pump detector initial verification, verification and subsequent use of checks. |
JJG 1098-2014: Verification Regulation of Medical Syringe Pump and Infusion Pump Analyzers---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Verification Regulation of Medical Syringe Pump and Infusion Pump Analyzers
State verification procedures People's Republic of China
Medical syringe pump and infusion pump tester
Issued on. 2014-08-01
2014-11-01 implementation
The State Administration of Quality Supervision, Inspection and Quarantine released
Medical syringe pump and infusion pump tester
Verification Regulation
Focal point. the National Technical Committee on Clinical Measurement
The main drafting units. China Testing Technology Research Institute
Supervision and Testing Nanjing hospital
Measurement and Testing Center in Wuxi City
Participated in the drafting unit. Dalian metrological verification tests
Xinjiang Institute of Metrology
The measurement procedure commissioned the National Technical Committee responsible for the interpretation of clinical medicine
The main drafters of this procedure.
Zhang from China (China Testing Technology Research Institute)
Hu Ning (Nanjing Supervision and Testing Institute)
SHI Min (Wuxi Metrology)
Drafters participate.
Gong Lan (China Testing Technology Research Institute)
Lei (Dalian metrological verification tests)
Li Feng (Xinjiang Institute of Measurement and Testing)
Lin Tao (China Testing Technology Research Institute)
table of Contents
Introduction (Ⅱ)
1 Scope (1)
2. References (1)
3 terminology and units of measurement (1)
Flow rate 3.1 (1)
3.2 cumulative flow (1)
3.3 occlusion pressure (1)
4 Overview (1)
5 metering performance requirements (1)
Flow rate 5.1 (1)
5.2 cumulative flow (2)
5.3 occlusion pressure (2)
6 General Requirements (2)
6.1 Appearance (2)
6.2 Identification (2)
6.3 Display (2)
7 control of measuring instruments (2)
7.1 verification Conditions (2)
7.2 Verification Project (3)
7.3 Verification Methods (3)
7.4 Verification Results Processing (6)
7.5 test cycle (6)
Appendix A test pattern original record format (7)
B test certificates/test results in the notice page format Shape appendix (10)
Appendix C water density table (13)
Introduction
JJF1001-2011 "common measurement terms and definitions", JJF1002-2010 "national verification procedures to prepare
Rules ", JJF1059.1-2012" Evaluation and Expression of Uncertainty in Measurement ", JJF1259-2010" medical syringe pump
And infusion pump calibration standard "constitute the support of the verification protocol development on the basis of a series of documents. Medical syringe pump
And measurement of the performance of infusion pumps detector using JJF1259-2010, including traffic congestion and pressure, the cumulative increase
Flow requirements.
This procedure was first formulated.
Medical syringe pump and infusion pump detector test procedures
1 Scope
The order applies to medical injection pump and infusion pump tester (hereinafter referred to as the detector) initial verification, follow-up inspection
And given the use of checks.
2. References
This procedure references the following documents.
JJF1259-2010 medical injection pump and infusion pump calibration specification
Laboratory use specifications and test methods GB/T 6682-2008 Analysis
For dated references, only the dated edition applies to this procedure; undated reference documents
Member, the latest edition (including any amendments) applies to this procedure.
3 terminology and units of measurement
3.1 Traffic flowrate
The volume of fluid flow per unit time of the pipe cross-section, in units of mL/h. [JJF1259-2010 Definition 3.1]
3.2 cumulative flow totalflow
Over a period of time the volume of fluid flowing through the pipe cross-section, measured in mL.
3.3 occlusion pressure occlusionpressure
In the pipe blocked, the pressure on the liquid wall, the unit is kPa.
4 Overview
Detector is used for a medical infusion pump and syringe pump flow and pressure parameters were measured, special measurement calibration
Instrument, measure the flow of medical infusion pump and syringe pump output through the built-in liquid flow and pressure sensors, tubes
Channel parameters such as pressure, mainly by the liquid sample, a pressure measurement module, a timer, a signal processing unit, the display unit
And the like.
5 metering performance requirements
5.1 Flow
Detector flow permissible errors in line with the provisions of 6.4 JJF1259-2010 in Table 1.
Table 1 Traffic
Flow range (mL/h) maximum permissible error Repeatability Resolution (mL/h)
[5,20) ± (2.0% reading a division value) 1.0% 0.01
[20,200] ± (1.0% reading a division value) of 0.5% or 0.1 0.01 *
(200,1000] ± (2.0% reading a division value) 0.1 1.0%
Note. "*" requires a minimum flow indication has four significant figures.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of JJG 1098-2014_English be delivered?Answer: Upon your order, we will start to translate JJG 1098-2014_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of JJG 1098-2014_English with my colleagues?Answer: Yes. The purchased PDF of JJG 1098-2014_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay.
|