Calibration Specification for Partial Discharge Testers Based Pulse Current Method
| Valid |
JJF 1616-2017
|
PDF similar to JJF 1616-2017
Basic data Standard ID | JJF 1616-2017 (JJF1616-2017) | Description (Translated English) | Calibration Specification for Partial Discharge Testers Based Pulse Current Method | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A55 | Classification of International Standard | 17.220 | Word Count Estimation | 24,275 | Date of Issue | 2017-02-28 | Date of Implementation | 2017-05-28 | Quoted Standard | GB/T 7354-2003 | Regulation (derived from) | General Administration of Quality Supervision, Inspection and Quarantine No. 23 of 2017 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to the pulse current method, measuring the frequency range of 10 kHz to 500 kHz between the partial discharge tester calibration. |
JJF 1616-2017: Calibration Specification for Partial Discharge Testers Based Pulse Current Method ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Calibration Specification for Partial Discharge Testers Based Pulse Current Method
National Metrological Technical Code of the People's Republic of China
Standard Specification for Calibration of Partial Discharge Tester for Pulsed Current
2017-02-28 released
2017-05-28 Implementation
State Administration of Quality Supervision, Inspection and Quarantine issued
Pulse current method for partial discharge tester
Calibration specification
Unit. National Electromagnetic Metrology Technical Committee High Pressure Metering Subcommittee
Major drafting unit. National high voltage metering station
State Grid Corporation of Zhejiang Electric Power Company Electric Power Research Institute
Participated in the drafting unit. State Grid Jiangsu Electric Power Company Electric Power Research Institute
Electric Power Research Institute of Guangxi Power Grid Co., Ltd
State Grid Tianjin Electric Power Company Electric Power Research Institute
State Grid Electric Power Research Institute Wuhan Nairui limited liability company
This specification is commissioned by the National Electromagnetic Metrology Technical Committee of the High Pressure Metrology Subcommittee.
The main drafters of this specification.
Lei Min (National High Voltage Metering Station)
Zhan Hongyan (State Grid Zhejiang Electric Power Company Electric Power Research Institute)
Zhang jun (national high voltage metering station)
Participate in the drafters.
Zhou Zhicheng (State Grid Jiangsu Electric Power Company Electric Power Research Institute)
Yin Liqun (Guangxi Power Grid Co., Ltd. Electric Power Research Institute)
Sun Zhe (State Grid Tianjin Electric Power Company Electric Power Research Institute)
Liu Yi (State Grid Electric Power Research Institute Wuhan Nairui limited liability company)
table of Contents
Introduction (Ⅲ)
1 Scope (1)
2 References (1)
3 terms and units of measurement (1)
3.1 Pulse current method partial discharge tester (1)
3.2 Impulse response of partial discharge tester (1)
3.3 Calibration pulse generator (1)
3.4 apparent charge (1)
3.5 Pulse resolution time (1)
3.6 pulse repetition rate (1)
3.7 Measurement Sensitivity (1)
3.8 lower limit frequency f1 and upper frequency f2 (2)
3.9 Double pulse generator (2)
4 Overview (2)
5 Metrological characteristics (3)
5.1 Upper and lower cutoff frequency (3)
5.2 Apparent charge amplitude (3)
5.3 Shift Impulse Response Gain Value (3)
5.4 Positive and negative impulse response asymmetry (3)
5.5 Low repetition rate impulse response (3)
5.6 Pulse resolution time (3)
5.7 Measurement Sensitivity (3)
5.8 pulse repetition rate (3)
5.9 Number of pulses (3)
5.10 Pulse acquisition capability (3)
5.11 Calibration pulse generator (3)
5.12 safety performance (4)
6 Calibration conditions (4)
6.1 Environmental conditions (4)
6.2 Measurement standards and other equipment (4)
7 Calibration items and calibration methods (4)
7.1 Calibration items (4)
7.2 Calibration method (5)
8 Calibration results (9)
8.1 Calibration Certificate (9)
8.2 Uncertainty Rating (9)
9 re-school time interval (9)
Appendix A Cutoff Frequency and Apparent Charge Amplitude Measurement Uncertainty Evaluation Example (10)
Appendix B Calibrate Original Record Format (13)
Appendix C Calibration Certificate Inside Page Format (16)
Introduction
This specification is based on JJF 1071-2010 "national measurement calibration standard preparation rules", JJF 1001-2011 "pass
Metrological terms and definitions ", JJF 1059.1-2012" measurement uncertainty assessment and representation "compiled from.
This specification is the first release.
Standard Specification for Calibration of Partial Discharge Tester for Pulsed Current
1 Scope
This specification applies to the pulsed current method, measuring the frequency range of 10kHz ~ 500kHz between the partial discharge test
Calibration of the test instrument.
2 reference file
This specification refers to the following documents.
Partial discharge measurement GB/T 7354-2003
For dated references, only the dated edition applies to this specification; references that are not dated
, The latest version (including all modifications) applies to this specification.
3 terms and units of measurement
GB/T 7354-2003 Definitions and the following terms and definitions apply to this specification.
3.1 pulsed current method partial discharge tester pulsecurrentmethodbasedpartialdischargetester
Through the acquisition of pulse current signal to achieve electrical equipment, local discharge measurement of special equipment.
3.2 Impulse response of partial discharge tester pulseresponseofpartialdischargetester
When the measuring circuit of the partial discharge tester injects the calibration pulse signal, its corresponding pulse output is displayed.
3.3 Calibration pulse generator calibratedpulsegenerator
By the voltage amplitude of U0 pulse voltage generator and the injection capacitor C0 with the composition, can produce a known charge amount q0
Of the pulse current generator.
3.4 apparent charge amount apparent charge [GB/T 7354-2003, definition 3.3.1]
The apparent charge is equal to the specified test circuit, if in a very short period of time between the two sides of the sample
The amount of readings on the measuring instrument is the same as that of the PD circuit itself. The apparent charge is usually used in the library
(pC).
3.5 pulse resolution time pulseresolutiontime
[GB/T 7354-2003, definition 3.9.5]
The shortest time between successive pulses of the same duration and the same amount of charge and the same amount of charge
Interval, the amplitude of the impulse response in this time interval is not greater than 10% of the amplitude of a single pulse.
The pulse resolution time is generally inversely proportional to the bandwidth Δf of the measurement system and is also a measure of the system.
The ability to characterize.
3.6 pulse repetition rate pulserepetitionrate
[GB/T 7354-2003, definition 3.3.2]
The ratio of the total number of localized pulses recorded in the selected time interval to the time interval.
Note. Actually only consider the pulse above the specified amplitude or within the specified amplitude range.
3.7 Measurement sensitivity detectablesensitivity
|