US$209.00 · In stock Delivery: <= 3 days. True-PDF full-copy in English will be manually translated and delivered via email. GB/T 9249-2008: Self-balance recorders and indicators for use in industrial-process measurement and control systems Status: Valid GB/T 9249: Evolution and historical versions
Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
GB/T 9249-2008 | English | 209 |
Add to Cart
|
3 days [Need to translate]
|
Self-balance recorders and indicators for use in industrial-process measurement and control systems
| Valid |
GB/T 9249-2008
|
GB/T 9249-1988 | English | 279 |
Add to Cart
|
3 days [Need to translate]
|
Self-balance recorders and indicators for use in industerial-process measurement and control systems
| Obsolete |
GB/T 9249-1988
|
PDF similar to GB/T 9249-2008
Basic data Standard ID | GB/T 9249-2008 (GB/T9249-2008) | Description (Translated English) | Self-balance recorders and indicators for use in industrial-process measurement and control systems | Sector / Industry | National Standard (Recommended) | Classification of Chinese Standard | N14 | Classification of International Standard | 25.040.40 | Word Count Estimation | 9,954 | Date of Issue | 2008-07-28 | Date of Implementation | 2009-02-01 | Older Standard (superseded by this standard) | GB/T 9249-1988 | Quoted Standard | GB/T 3386.1-2007; GB/T 18271.2; GB/T 18271.3; GB/T 22112-2008; GB/T 22162-2008; JB/T 9266-1999; JB/T 9329-1999 | Regulation (derived from) | Announcement of Newly Approved National Standards No. 12 of 2008 (No. 125 overall) | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China | Summary | This standard specifies the industrial process measurement and control systems automatically balance recorders and indicators (hereinafter referred to as instrumentation) of the basic parameters, requirements, test methods and inspection rules. This standard applies to industrial process measurement and control systems with an analog electrical signal electronic automatic potentiometer, balance electric bridge recorders and indicator (hereinafter referred to as the potential difference and bridge). Potentiometer input side of the nominal external resistance is not greater than 10000��, rated input side of the bridge outside the resistance is not greater than 1500�� This standard only applies to normal operating conditions using the instrument, the special operating conditions of the instrument using the additional technical requirements not within the standard range. This standard excludes devices with additional instrumentation additional requirements. |
GB/T 9249-2008: Self-balance recorders and indicators for use in industrial-process measurement and control systems ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Self-balance recorders and indicators for use in industrial-process measurement and conrol systems
ICS 25.040.40
N14
National Standards of People's Republic of China
Replacing GB/T 9249-1988
Industrial process measurement and control systems
Self - balance recorders and indicators
Posted 2008-07-28
2009-02-01 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China released
Foreword
The revised standard and replaces GB/T 9249-1988 "industrial process measurement and control systems automatically balance recorders and indicators."
This standard and GB/T 9249-1988 the main differences are as follows.
--- In accordance with the provisions of GB/T 1.1-2000 were editorial changes;
--- Update the normative references;
--- Table 1 Accuracy class increased by 0.2 of a file, 0.3 in brackets, indicating not recommended;
--- Increased supply voltage short interruptions, low supply voltage drop, electromagnetic radiation, electrostatic discharge immunity requirements;
, GB/T 18271.2-2000 and GB/T --- test methods according to the provisions of GB/T 3386.1-2007 18271.3-2000 of
carried out;
Title --- Chapter 7 to "storage."
The standard proposed by China Machinery Industry Federation.
The standard measurement and control of industrial processes by the National Standardization Technical Committee at its first Technical Committee (SAC/TC124/SC1)
Centralized.
This standard is drafted by. Shanghai Institute of Industrial Automation Instrumentation.
Participated in the drafting of this standard. Shanghai Automation Instrumentation System Inspection and Testing Institute.
The main drafters of this standard. Li Minghua, Tsai Wen-chi, Lo Ting.
This standard replaces the standards previously issued as follows.
--- GB/T 9249-1988.
Industrial process measurement and control systems
Self - balance recorders and indicators
1 Scope
This standard specifies the industrial process measurement and control systems automatically balance recorders and indicators (hereinafter referred to as instrumentation) of the basic parameters
Number of requirements, test methods and inspection rules.
This standard applies to industrial process measurement and control systems with electronic automatic potentiometer, balancing bridge electrical signal and analog recorder
Indicator (hereinafter referred to as the potentiometer and bridge). Potentiometer input side of the outer resistance rating of not more than 10000Ω; transmission bridge
Rated into the outer side of the resistance is not greater than 1500Ω.
This standard applies only under normal operating conditions additional technical requirements under the instrument instrument, special working conditions of use does not belong
Scope of this standard. This standard includes additional requirements nor the instrument with an additional device.
2 Normative references
The following documents contain provisions which, through reference in this standard and become the standard terms. For dated references, subsequent
Amendments (not including errata content) or revisions do not apply to this standard, however, encourage the parties to the agreement are based on research
Whether the latest versions of these documents. For undated reference documents, the latest versions apply to this standard.
GB/T 3386.1-2007 industrial process control systems with electric and pneumatic analogue chart recorders - Part 1. Performance Evaluation party
Method (IEC 60873-1.2003, IDT)
Tests under reference conditions. GB/T 18271.2 process measurement and control devices - General methods and procedures for evaluating performance - Part 2
(GB/T 18271.2-2000, idt IEC 61298-2.1995)
GB/T 18271.3 process measurement and assessment methods and procedures - Part 3 common control device performance. effects of influence quantities test
(GB/T 18271.3-2000, idt IEC 61298-3.1998)
GB/T 22112-2008 industrial automation, instrumentation and arrangement of terminals logo
GB/T 22162-2008 panel and rack - mounted industrial process measurement and control instrumentation disk and hole size (IEC 60668. 1980,
IDT)
JB/T 9266-1999 indicator Temperature measurement range
JB/T 9329-1999 instrumentation transportation, transportation, storage and basic environmental conditions and test methods
3 basic parameters
3.1 Environmental conditions
The meter shall be able to ambient temperature of 0 ℃ ~ 50 ℃, relative humidity of 10% to 90% of normal working conditions. Small instrument (Scale
Length not more than 140mm) ambient temperature range of 0 ℃ ~ 40 ℃.
3.2 Power
Instrumentation shall nominal voltage is 220V, 50Hz frequency nominal value of the AC power supply.
3.3 Dimensions
Meter dimensions should be consistent with GB/T 22162-2008 and other relevant standards.
3.4 Measuring range
Thermocouple and RTD temperature measurement instruments supporting its temperature measurement range should be consistent with the provisions of JB/T 9266-1999 of.
And supporting other instruments measuring instruments, the measurement range specified by the manufacturer or consultation with users.
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB/T 9249-2008_English be delivered?Answer: Upon your order, we will start to translate GB/T 9249-2008_English as soon as possible, and keep you informed of the progress. The lead time is typically 1 ~ 3 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB/T 9249-2008_English with my colleagues?Answer: Yes. The purchased PDF of GB/T 9249-2008_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version GB/T 9249-2008?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB/T 9249-2008 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|