|
US$874.00 · In stock Delivery: <= 4 days. True-PDF full-copy in English will be manually translated and delivered via email. GB 9706.13-2008: Medical electrical equipment -- Part 2: Particular requirements for the safety of automatically-controlled brachytherapy after loading equipment Status: Obsolete GB 9706.13: Evolution and historical versions
| Standard ID | Contents [version] | USD | STEP2 | [PDF] delivered in | Standard Title (Description) | Status | PDF |
| GB 9706.13-2008 | English | 874 |
Add to Cart
|
4 days [Need to translate]
|
Medical electrical equipment -- Part 2: Particular requirements for the safety of automatically-controlled brachytherapy after loading equipment
| Obsolete |
GB 9706.13-2008
|
| GB 9706.13-1997 | English | 799 |
Add to Cart
|
3 days [Need to translate]
|
Medical electrical equipment--Part 2: Particular requirements for the safety of remote-controlled automatically-driven gamma-ray afterloading equipment
| Obsolete |
GB 9706.13-1997
|
PDF similar to GB 9706.13-2008
Basic data | Standard ID | GB 9706.13-2008 (GB9706.13-2008) | | Description (Translated English) | Medical electrical equipment -- Part 2: Particular requirements for the safety of automatically-controlled brachytherapy after loading equipment | | Sector / Industry | National Standard | | Classification of Chinese Standard | C43 | | Classification of International Standard | 11.040.60 | | Word Count Estimation | 22,259 | | Date of Issue | 2008-12-15 | | Date of Implementation | 2010-02-01 | | Older Standard (superseded by this standard) | GB 9706.13-1997 | | Adopted Standard | IEC 60601-2-17-2004, IDT | | Regulation (derived from) | Announcement of Newly Approved National Standards No. 21 of 2008 (No. 134 overall) | | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China | | Summary | This Chinese standard specifies the use dedicated afterloading brachytherapy technique for patients with type automatic control equipment safety requirements. This standard applies toutomatic controlled after-loading equipment requirements, this device and only use containing ��, �� and neutron sealed radioactive sources, sealed radioactive sources can be automatically sent from the source storage device inside the treatment applied to the source location and Back from the source storage device treatment position, designed to come into contact with patients, source drive mechanism should be pre-programmed by the control timer or timing device control, automatically complete the movement of radioactive sources. Control timer or timing device may be a programmable electronic subsystems PESS (a computer or microprocessor), it can be non-programmable electronic systems. The requirements of this standard equipment under the supervision of qualified personnel to use, regular maintenance |
GB 9706.13-2008: Medical electrical equipment -- Part 2: Particular requirements for the safety of automatically-controlled brachytherapy after loading equipment ---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Medical electrical equipment.Part 2. Particular requirements for the safety of automatically-controlled brachytherapy afterloading equipment
ICS 11.040.60
C43
National Standards of People's Republic of China
GB 9706.13-2008/IEC 60601-2-17.2004
Replacing GB 9706.13-1997
Medical Electrical Equipment
Part 2. Automatic Control brachytherapy
Safety requirements for equipment installed after
(IEC 60601-2-17.2004, IDT)
Published 2008-12-15
2010-02-01 implementation
Administration of Quality Supervision, Inspection and Quarantine of People's Republic of China
Standardization Administration of China issued
Table of Contents
Preface Ⅰ
Introduction Ⅱ
The first chapter outlines 1
1. Scope and object 1
2 Terms and Definitions 2
3 General requirements 4
General claim 4 Test 4
5 Category 4
6 to identify, mark and file 5
The second environmental conditions 7
10 environmental conditions 7
Title III of the danger of electric shock protection 7
Title IV of protection against mechanical hazards 7
The fifth chapter of unnecessary or excessive radiation hazard protection 7
29 X-ray radiation 7
30 α, β, γ neutron radiation and other particle radiation 8
Title VI of the risk protection ignite flammable anesthetic gases mixed 14
Title VII of the over-temperature and other security risk protection 14
The accuracy of the risk protection and output data of the eighth chapter work 14
Accuracy of operating data 14 50
Chapter 9. abnormal operating and fault states; Environmental Testing 15
52 failure and abnormal operation state 15
Tenth chapter 16 structural requirements
Annex L (normative) Normative References 17
Chinese Index 18
English 20 Index
GB 9706.13-2008/IEC 60601-2-17.2004
Foreword
All technical contents of this section is mandatory.
"Medical Electrical Equipment" safety standard series consists of two parts.
--- Part 1. General requirements for safety;
--- part 2. requirements for the safety.
This section requirements for safety, is part of the 13 GB 9706.
This section is equivalent to using IEC 60601-2-17.2004 "Medical Electrical Equipment Part 2. Automatic Control installation brachytherapy
Equipment requirements for the safety. "
For ease of use, made in this section, the following editorial changes. hereinafter "present standard" to "in this section."
This partial replacement GB 9706.13-1997 "Medical Electrical Equipment Part II. After a remote γ-ray apparatus driven automatic safety device
Particular requirements. "
This section compared with GB 9706.13-1997, was modified in the following areas.
--- standard name from "automatic remote loading driven after γ-ray equipment requirements for safety" to "automatic brachytherapy
Safety requirements for equipment installed after ";
--- This section is broader than the scope of GB 9706.13-1997, and applies to all installed after the use of technology for patient treatment close
Treatment automatic control devices, and includes only the β, γ and neutron radiation source after sealing apparatus installed;
--- parallel standard portion according to the present GB 9706.15-2008, IEC 60601-1-2 and IEC 60601-1-4 relatively GB 9706.13-
1997 by the NA becomes applicable.
Annex L of this portion is a normative, Annex AA is an information appendix.
This section proposed by the State Food and Drug Administration.
This part of the National Standardization Technical Committee medical radiation therapy, nuclear medicine and dosimetry equipment at the Technical Committee (SAC /
TC10/SC3) centralized.
This part of the drafting unit. China Nuclear Power Institute of equipment manufacturer, Beijing Medical Device Testing.
The main drafters of this section. Zhou Jianming, SONG Lian there, Han Yong, Jiao spring camp.
This section replaces the following previously issued as follows.
--- GB 9706.13-1997.
GB 9706.13-2008/IEC 60601-2-17.2004
Introduction
If the equipment fails to give close brachytherapy dose design or equipment does not meet the requirements of patient safety standards for medical electrical equipment
Requirements, the use of close brachytherapy device patients can be compromised. If the device itself fails to sufficiently shield the radiation source in the reservoir
Endogenous and (or) shielding design of the treatment room is not sufficient, then the device may cause damage to nearby equipment personnel.
This identifies the manufacturer specific standard should be installed after the automatic control equipment design and manufacturing brachytherapy compliance requirements, but unknown
DESCRIPTION OF THE PREFERRED fine performance requirements of the device, the purpose of determining the design for the basic safety related operating functions of the member. This specific standard
Set a limit to reduce device performance, when it exceeds the limit the scope of the present can be considered a fault condition, the radiation source can interlock return reservoir
Endogenous and then prevent the continued operation of the device.
GB 9706.13-2008/IEC 60601-2-17.2004
Medical Electrical Equipment
Part 2. Automatic Control brachytherapy
Safety requirements for equipment installed after
The first chapter outlines
Except for the following modifications, GB 9706.1-2007 "Medical Electrical Equipment Part 1. General requirements for safety" (hereinafter referred to as "secure universal
Requirements ") Chapter shall apply.
Scope 1 and object
In addition to these changes, "General requirements for safety" of this chapter apply.
1.1 Scope
supplement.
1.1.101 This standard specifies the specific safety requirements of automatic control devices for patients after loading with brachytherapy techniques.
1.1.102 This section specifies the requirements for automatic control of the loading device type, the device
--- and containing only the β, γ and neutron sealed radioactive sources;
--- can automatically sealed radioactive sources to the treatment site within the applicators to return from the reservoir and a reservoir source from the treatment site;
--- is designed to contact with a patient;
--- source should drive mechanism by a preset control timer or timing program controlling means, moving the radiation source is automatically completed. Total control
When the timing device, or may be a programmable electronic subsystem PESS (computer or microprocessor), it may not be electrically programmable
Subsystem.
This part of device requirements 1.1.103
--- under the supervision of qualified personnel;
--- regular maintenance;
--- maintenance on a regular basis by the user.
This section requires the use of sealed radioactive sources is not specified, the other requirements for sealed radioactive standards (see 6.8.3).
1.1.104 requires this section based on the following assumptions.
--- treatment plan to be effective, given the appropriate treatment parameters;
--- source intensity radioactive source equipment is known.
To ensure completion of the treatment parameters can be preset apparatus, this part of the special requirements.
--- chosen by the radiation source with respect to the selected combination may be applied to the source and positioned within the movable applicators;
--- irradiation at selected time, selected by a combination of radiation source radiation is administered;
--- equipment operators or other personnel around will not cause unnecessary harm.
1.2 Purpose
instead.
Purpose of this standard is to establish a dedicated apparatus after the automatic control device brachytherapy specifications and test requirements for the safety test. this
Special safety standards required only raised the total of functional requirements, without providing special technical methods of its implementation.
1.3 Specific Standard
supplement.
GB 9706.13-2008/IEC 60601-2-17.2004
Relations with the General Standard 1.3.101
Dedicated to understanding the present binding standard GB 9706.1-2007 "Medical Electrical Equipment Part 1. General requirements for safety" (hereinafter referred to as
A universal standard), this standard is a general standard specific additions and modifications, inspection and test simultaneously to follow the requirements of common standards.
The articles in this special standard, consistent with chapter numbers, bar and "common standards". For the "universal standards" and tied for standard text change
More, with the following words, he said.
"Instead of" means "General Standard" Chapter strip or entirely replaced by the standard-specific chapter or section.
"Complementary" refers to specific provisions of this Standard requires added "General Standard" in.
"Modify" means "General Standard" Chapter or provision of this specific standard provisions as indicated above has been modified.
For added "General Standard" in the article, figures, or tables are numbered starting from 101, supplemented by the letter Appendix AA, BB and so marked, make up
Charge of the project with aa), bb), etc. represent.
The term "in the present section" refers to a common reference standard must be present and the specific standard.
This is no specific standard at the respective chapter, chapter or article, the "General Standard" in the article, or chapter shall apply, without making a change.
It should be noted that any suitable part of the "General Standards", in this specific standard explanation has been given, although these parts
This may correspond to specific standard.
This standard requires special preference to the respective requirements of common standards.
1.3.102 relationships with other standards and documents
NOTE. See Appendix L (normative).
1.5 Collateral Standard
supplement.
1.5.101 GB 9706.15-1999
All of that parallel application of the criteria.
1.5.102 IEC 60601-1-2
All of that parallel application of the criteria.
1.5.103 GB 9706.12-1997
The collateral standard does not apply.
1.5.104 IEC 60601-1-4
All of that parallel application of the criteria.
2 Terms and Definitions
In addition to these changes, "General requirements for safety" of this chapter apply.
supplement.
Given in "General requirements for safety" and standard parallel and the following terms and definitions apply.
Note 1. The specific standard terminology index terms listed in alphabetical order and the original reference defined.
Note 2. In the present specific standard, the irradiating means that the patient receives the entire course of treatment in the treatment position and a treatment time, the radiation source does not belong to the transmission time
Healing period.
Note 3. In the specific standard, the term "operator" during their therapy control denotes brachytherapy device; term "user" means a close representation of the
Therapy equipment use and maintenance of the units and individuals responsible. The term "radiation therapy physician" and "radiation therapy experts" not to use this specific standard,
It represents the observation and personnel responsible for a given patient treatment prescribed dose.
2.1 parts and accessories and attachments
Additional definitions.
2.1.101
Use the following abbreviations in this section.
--- PESS programmable electronic subsystem;
--- SFC single fault condition;
GB 9706.13-2008/IEC 60601-2-17.2004
--- TCP treatment console.
2.1.102
Β source in the water along the perpendicular bisector 2mm absorbed dose rate at which the unit is Gy/s.
2.1.103
Using one or more of a cavity for sealed radioactive sources,, superficial and radiotherapy in the tube between organizations.
2.1.104
Radiation therapy, the condition does not choose to run again after the irradiation treatment interruption and restart.
2.1.105
Radioactive sources or radioactive sources chains in selected residence time of the treatment site.
2.1.106
Reference air kerma rate γ source for brachytherapy, with at Gy 1m/s FIG.
Note. The representation allows the use of multiple (one thousandth, megabytes, etc.).
2.1.107
Radiation therapy, when the selection is completed and verified by the operating conditions and not by irradiation interrupt reach ready state, away from the ready state
The beginning of irradiation.
2.1.108
Stop irradiation before irradiation is terminated, this does not stop with the possibility of re-select operating conditions to continue treatment.
2.1.109
With or without applicators to one or more of a radioactive source into the body lumen (e.g., blood vessels, respiratory tract, gastrointestinal tract, etc.) of
Radiation Therapy.
2.1.110
Neutron source for the dose rate brachytherapy, with at Gy 1m/s FIG.
2.1.111
Acknowledged by the competent authorities have engaged in some kind of professional career.
2.1.112
Some arrangement for sealed radioactive sources after loading apparatus can be separated by a non-radioactive spacer, before each irradiation may be permanently
Or selectively combined together. Radiation source chain is typically selected to give a particular dose distribution.
2.1.113
In air after attenuation and scatter correction, the reference distance is air kerma rate of 1m.
Note. reference air kerma rate KR intensity ratio similar to the air release kinetic SK used in the USA, which is equal to the kinetic energy release rate ratio of air and free space along
In the center of the radiation source from the radiation source to the product of the perpendicular distance from the square of the calibration points.
GB 9706.13-2008/IEC 60601-2-17.2004
2.1.114
γ source intensity radiation source is used to specify the device, β source strength or neutron source strength.
2.1.115
Radiation therapy, or stop to prevent further irradiation (eg. Returns the ready state), and no reselection operation and reconfirmation
When the conditions are, no restart irradiation possibilities.
2.1.116
Transmitted by the radiation source is a reservoir to the treatment site or the process is transferred by a reservoir to the treatment site.
2.1.117
During the radiotherapy, the irradiation parameters of the patient described aspects, such as the energy radiation, absorbed dose, treatment time.
2.1.118
Treatment includes the sum of all the dwell times.
3 General requirements
"General requirements for safety" of this chapter apply.
General requirements Test 4
In addition to these changes, "General requirements for safety" of this chapter apply.
4.1 Test
supplement.
To complete the test safely be careful. For example. If possible, analog source. The specific standard in this type of test,
Routine testing can be used by the manufacturer during manufacture and installation and mounting assembly performed.
Category 5
instead.
Equipment and its application classification section must be labeled and (or) identification specified in Chapter 6. These include.
5.1 divided by type of protection against electric shock.
Apparatus within the scope of this class Ⅰ device should be.
5.2 points according to the degree of protection against electric shock.
Unless otherwise specified, the application part within the scope of this application should be the B-type portion.
5.3 according to the degree of protection against harmful inlet points (see GB 4208).
Unless otherwise specified, within the scope of this device should be no special protective equipment (IPX0).
5.4 recommended by the manufacturer of disinfection and sterilization methods points.
5.5 degree of security used by mixing with a gas mixture where gas or oxygen or nitrous oxide and anesthetic gas and air in flammable point.
Unless otherwise specified, within the scope of this apparatus should not be in a mixed gas and oxygen or an alkylene oxide or a flammable anesthetic gas and air
Apparatus for use in a mixed gas where nitrogen.
5.6 points by working the system.
Within the scope of this device should be suitable for continuous operation of the apparatus.
GB 9706.13-2008/IEC 60601-2-17.2004
6 to identify, mark and file
In addition to these changes, "General requirements for safety" of this chapter apply.
External marking device or apparatus components 6.1
z) a removable protector
instead.
When the device is not available for two or more alternative applications (ie. to use a particular function of a protection device removed), they should
When the device statement.
By examining a random file to verify compliance.
Supplementary items.
aa) the appropriate parts of the device should be easy to recognize in a clear, permanent marker fixed to shows.
1) the maximum total for each radionuclide source device design strength;
2) indicate potentially dangerous radioactive ionizing radiation symbol (see Annex L GB 18871-2002);
3) If a reservoir is used only in a restricted access a predetermined treatment chamber (see 30.1.1);
4) If an additional external supply (e.g. air).
bb) apparatus for multiple pieces of radiation sources, an additional means (e.g. PESS) should have a permanent and easy selection identifying each channel
The selected radioactive sources.
cc) should be permanent attachment device to indicate that the name is stored in a reservoir radionuclide source intensity and Calibration
date.
dd) each interchangeable applicators, applicators in the appropriate position on the package should be a permanent identification mark or separate.
ee) a significant asymmetry (e.g. bendable or partial protection), and can be inserted into a patient to different orientations of a rigid clinically applied
Source, a recommendation is used in the random file after the outer radiographic or other appropriate test methods shall be inserted orientation identification
Mark.
NOTE. Some claim administration activity brachytherapy source (e.g.. Ci of or Bq) should also be indicated.
6.7 lights and buttons
a) Indicator Color
instead.
When the indicator used in the treatment console or another console, a light color which meet the following requirements.
It calls for immediate emergency measures against accidental operating status. red;
Radiation source is in the treatment position. yellow;
Radioactive source in transmission. Yellow (flashing);
Ready state. Green;
The power switch is closed, but not yet ready state. white.
State of the device should be indicated, in addition can also be indicated by the color or shape or location indicator of the accompanying text.
In addition to the loading device operating near the patient, rather than in the treatment chamber for the radiation source is indicated on the console in the transmission
Or treating appropriate position indicator color is red. This requires the user to color observed in the treatment room the red indicator should immediately take urgent
Measures.
6.8 random file
6.8.2 Manual
d) cleaning, disinfection and sterilization in contact with the patient member
supplement.
Instructions for use should contain. When applicators for wet or steam sterilization, applicators lumen of residual liquid, then the residue
Possible warning of danger to the other components of the liquid left in the system. These warnings should include a residual liquid seal Radiation
Corrosion hazard and a source of damage to the drive mechanism.
GB 9706.13-2008/IEC 60601-2-17.2004
e) from the network power supply apparatus with additional power source and
instead.
Instructions shall comprising. a battery power supply and other additional internal periodic inspection, maintenance and operation of the charging method and the push
Recommended period.
Supplementary items.
aa) testing interlocking means
Instructions shall comprising. operating cycle and the recommended test methods for all interlocks.
Security bb) test connector
If the source device has a radiation source to a drive mechanism and a channel is connected to the source is applied, using the test specification shall contain even
Description connector safety operating procedures.
cc) the required procedures dedicated control timer fault
Instructions shall contain. Description of the special procedures required for the control failure timer (see 30.3.2).
dd) returns a failure radiation source is a reservoir device
Instructions shall contain. a case of failure to take when the radiation source is a reservoir return Recommended procedures and the
Recommended emergency equipment needs.
ee) delivered dose
Instructions shall comprising. a radiation source from the reservoir specified source transmits its way back to the treatment site, a predetermined bit 50.1.4
Air kerma (or absorbed dose) on the data set.
Manual should be recommended to determine the expected total delivered dose, if necessary, to allow the dose to be considered in the calculation.
Instructions shall contain Warning. If the source is sent repeatedly from the reservoir and retracting the radiation source during treatment, may be affected at
Accuracy party dose.
ff) limit the passage and applicators
Instructions should provide.
--- specific diameter of the normal operation of the minimum allowed bending radius of the applicators and a channel;
--- cloth can not meet the channel coherence state bits required 50.2.1 and applicators, and should deal with the clinical device
With an explanation of the relationship between (see 1.1.104);
--- characteristics of any obstructions impede passage of radioactive sources or radioactive sources chains and free movement of the applicators.
gg) checking pollution
Manual leakage of radioactive radiation source should be advised to make appropriate checks.
hh) asymmetric inspection of applicators
When appropriate (see 6.1), using recommended instructions should be asymmetric orientation applicators or other radiographic examination.
i) inspection and maintenance functions
User manual should be advised of the equipment, especially the source of the drive mechanism, regular function tests and maintenance to ensure equipment
Continued safe use.
When necessary, the instructions for use should be recommended test simulated radiation source.
j) transfer time limit radiation source
Instructions for use should contain a radioactive source from a reservoir to the treatment site transmits its way back to the maximum transmission time, if not
There are restrictions must be declared.
kk) Check the radiation source intensity
Manual measurement methods employed should be recommended nationally recommended strength measurement each radiation source and the newly received device with
Unit value used for comparison.
6.8.3 SHEET
a) Overview
GB 9706.13-2008/IEC 60601-2-17.2004
supplement.
Technical specification provides for.
1) whether the source device and the reservoir is limited to a restricted access in the treatment room (see 30.1.1);
2) radionuclide equipment design, the maximum radiation exposure radiation source and a reservoir source's intensity;
3) radiation source chain (if any) allowed a range of combinations;
4) whether to allow access to equipment used in the radiation source from other suppliers, if the predetermined dimensions and tolerances should be allowed cladding radioactive sources;
5) radioactive sources should be consistent with international standards and other special requirements;
6) whether to allow the equipment used to obtain from other suppliers applicators, if allowed to be related to a predetermined dimensions and tolerances, and the source is recommended for administration
Whether the positioning device and reliability requirements of the connection test;
7 radiation source positioning error) during the treatment, the residence time of transmission time error and error.
Technical specification should further provide.
8) In all the fault condition, special requirements of this standard can not guarantee that indicates the location of the radiation source in a reservoir outside the filter;
9) radiation monitoring is recommended to use a separate type adapted irradiation near the patient;
10) use of a separate monitor radiation, any radiation monitor adjacent the portion of the device should be able to give a visual indication and irradiated
To run correctly under the anticipated maximum radiation dose levels;
After 11) mains supply fails, the monitoring of radiation continues to run and give an indication of several hours irradiation (e.g.. backup battery).
Technical specification shall contain. a remote device connected to a detailed description of the interrupt (see 30.1.8).
Technical specification shall contain. Guidance on identification may be represented, and operation of the processing device or radioactive components and
Prevention.
Technical specification should be given one or more doses of representative clinical application distribution, it should consider the impact of the dose distribution of applicators.
The second environmental conditions
10 environmental c...
Tips & Frequently Asked Questions:Question 1: How long will the true-PDF of GB 9706.13-2008_English be delivered?Answer: Upon your order, we will start to translate GB 9706.13-2008_English as soon as possible, and keep you informed of the progress. The lead time is typically 2 ~ 4 working days. The lengthier the document the longer the lead time. Question 2: Can I share the purchased PDF of GB 9706.13-2008_English with my colleagues?Answer: Yes. The purchased PDF of GB 9706.13-2008_English will be deemed to be sold to your employer/organization who actually pays for it, including your colleagues and your employer's intranet. Question 3: Does the price include tax/VAT?Answer: Yes. Our tax invoice, downloaded/delivered in 9 seconds, includes all tax/VAT and complies with 100+ countries' tax regulations (tax exempted in 100+ countries) -- See Avoidance of Double Taxation Agreements (DTAs): List of DTAs signed between Singapore and 100+ countriesQuestion 4: Do you accept my currency other than USD?Answer: Yes. If you need your currency to be printed on the invoice, please write an email to [email protected]. In 2 working-hours, we will create a special link for you to pay in any currencies. Otherwise, follow the normal steps: Add to Cart -- Checkout -- Select your currency to pay. Question 5: Should I purchase the latest version GB 9706.13-2008?Answer: Yes. Unless special scenarios such as technical constraints or academic study, you should always prioritize to purchase the latest version GB 9706.13-2008 even if the enforcement date is in future. Complying with the latest version means that, by default, it also complies with all the earlier versions, technically.
|