Verification Regulation of Calibration Device for Electric Cardiac Monitor
| Valid |
JJG 1016-2006
|
JJG 1016-1990 | English | RFQ |
ASK
|
3 days [Need to translate]
|
Rules for Drafting Norms of Pattern Evaluation of Measuring Instruments
| Obsolete |
JJG 1016-1990
|
PDF similar to JJG 1016-2006
Basic data Standard ID | JJG 1016-2006 (JJG1016-2006) | Description (Translated English) | Verification Regulation of Calibration Device for Electric Cardiac Monitor | Sector / Industry | Metrology & Measurement Industry Standard | Classification of Chinese Standard | A56 | Classification of International Standard | 17.220 | Word Count Estimation | 21,210 | Date of Issue | 2006-12-08 | Date of Implementation | 2007-03-08 | Quoted Standard | AQSIQ Announcement No. 187 of 2006 | Regulation (derived from) | AQSIQ Announcement No. 187 2006 | Issuing agency(ies) | General Administration of Quality Supervision, Inspection and Quarantine | Summary | This standard applies to ECG Calibrator initial verification, testing and use of follow-up inspection, also applies to the test with ECG heart function, EEG Monitor Calibrator center electrode initial verification test function, Subsequent testing and use test. |
JJG 1016-2006: Verification Regulation of Calibration Device for Electric Cardiac Monitor---This is a DRAFT version for illustration, not a final translation. Full copy of true-PDF in English version (including equations, symbols, images, flow-chart, tables, and figures etc.) will be manually/carefully translated upon your order.
Verification Regulation of Calibration Device for Electric Cardiac Monitor
National Metrology Verification Regulations of the People's Republic
ECG monitor
Released on.2006-12-08
2007-03-08 implementation
The General Administration of Quality Supervision, Inspection and Quarantine issued
This verification procedure was approved by the General Administration of Quality Supervision, Inspection and Quarantine on December 8,.2006.
And will be implemented as of March 8,.2007.
Focal Point. National Radio Measurement Technology Committee
Drafting unit. Inner Mongolia Autonomous Region Metrology and Testing Institute
Participated in the drafting unit. China Institute of Metrology
People's Liberation Army Medical Electromagnetic Metrology Station
This procedure entrusts the National Radiometric Technical Committee to explain
The main drafters of this procedure.
Ning Wei (Inner Mongolia Autonomous Region Measurement and Testing Institute)
Lu Jinhua (Inner Mongolia Autonomous Region Measurement and Testing Institute)
He Zhao (China Institute of Metrology)
Participate in the drafters.
Zhou Geping (Inner Mongolia Autonomous Region Metrology and Testing Institute)
Liu Feng (Inner Mongolia Autonomous Region Measurement and Testing Institute)
Jia Jiange (People's Liberation Army Medical Electromagnetic Metering Station)
Li Yuxue (People's Liberation Army Medical Electromagnetic Metrology Station)
table of Contents
1 Scope (1)
2 Citations (1)
3 Overview (1)
4 Metrology performance requirements (1)
4.1 Square wave signal generator (1)
4.2 Sine wave signal generator (1)
4.3 Standard heart rate signal generator (1)
4.4 Polarization voltage (2)
4.5 Input loop current sampling resistor (2)
4.6 Common mode rejection ratio verification device (2)
5 General technical requirements (2)
6 measuring instrument control (2)
6.1 Verification conditions (2)
6.2 Verification project (3)
6.3 Verification method (4)
6.4 Processing of verification results (9)
6.5 verification period (9)
Appendix A Original Record Format (10)
Appendix B Verification certificate page format (14)
Appendix C Format of the results of the verification result notice (16)
ECG monitor calibration instrument verification procedures
1 Scope
This procedure applies to the first verification, subsequent verification and in-use inspection of the ECG monitor; also applies to
The first verification and follow-up of the verification function of the central electric monitor of the heart and EEG machine with ECG monitoring function
Verification and in-use inspection.
2 Citations
JJG 760-2003 "Electrocardiograph"
JJG 749-1997 "Heart and Brain Electrogram Machine Verifier"
When using this procedure, care should be taken to use the current valid version of the above cited documents.
3 Overview
ECG monitor is used in accordance with JJG 760-2003 "Electrocardiograph" national metrological verification procedures
Special technical instruments for the measurement and verification of various technical indicators of ECG monitors, hereinafter referred to as “calibration instruments”.
The calibration instrument generates a standard signal for verification by an analog-to-digital converter, and the peak-to-peak value of various standard signals can be input to
The digital control of the analog-to-digital converter remains consistent.
4 Metrological performance requirements
4.1 Square wave signal generator
4.1.1 Voltage (peak-to-peak).
Range. 0.5mV~2mV Maximum allowable error. ±1%
4.1.2 Cycle.
Range. 0.5s~10s Maximum allowable error. ±1%
4.1.3 Output impedance.
Less than 600Ω.
4.2 Sine wave signal generator
4.2.1 Frequency.
Range. 0.1Hz~100Hz Maximum allowable error. ±1%
4.2.2 Voltage (peak-to-peak).
Range. Maximum allowable error in the range of 0.5mV~2mV. ±1%
4.2.3 Sine wave distortion.
< 5%.
4.2.4 Output impedance.
Less than 600Ω.
4.3 Standard heart rate signal generator (see Figure 1 for waveform)
......
|