YY/T 1587-2018English145 Add to Cart 0-9 seconds. Auto-delivery. Medical endoscopes - Video endoscopes Valid

PDF Preview: YY/T 1587-2018                   
      

YY/T 1587-2018: PDF in English (YYT 1587-2018)

YY/T 1587-2018 PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE’S REPUBLIC OF CHINA ICS 11.040.99 C 40 Medical endoscopes - Video endoscopes ISSUED ON: JUNE 26, 2018 IMPLEMENTED ON: JULY 01, 2019 Issued by: National Medical Products Administration Table of Contents Foreword ... 3  1 Scope ... 4  2 Normative references ... 4  3 Terms and definitions ... 4  4 Requirements ... 8  5 Test methods ... 11  Appendix A (Normative) Test method of brightness response characteristics 14  Appendix B (Normative) Test method of signal-to-noise ratio ... 19  Appendix C (Normative) Test method of spatial frequency response ... 23  Appendix D (Normative) Test method of static image tolerance ... 29  Medical endoscopes - Video endoscopes 1 Scope This standard specifies the terms and definitions, requirements, test methods for medical video endoscopes. This standard applies to medical video endoscopes (hereinafter referred to as video endoscopes) for visible spectrum imaging in endoscopy and surgery. This standard does not apply to medical video endoscopes for special spectral and non-visible spectral imaging. 2 Normative references The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard. GB 9706.19 Medical electrical equipment - Part 2: Particular requirements for the safety of endoscopic equipment GB/T 14233.1-2008 Test methods for infusion transfusion injection equipment for medical use - Part 1: Chemical analysis methods GB/T 16886.1 Biological evaluation of medical devices - Part 1: Evaluation and testing YY 0068.1-2008 Medical endoscopes - Rigid endoscope - Part 1: Optical properties and test methods YY 0068.2 Medical endoscopes - Rigid endoscope - Part 2: Mechanical properties and test methods 3 Terms and definitions The following terms and definitions apply to this document. 3.1 Conversion function Where: S - Output signal; N - The root mean square of the noise signal. 3.4 Saturation value The output signal value which further increases the brightness of the object while the output signal remains constant. 3.5 Static image tolerance The ratio of the maximum brightness of the critical object plane to the minimum brightness of the critical object plane that the video endoscope can distinguish in a single exposure. 3.6 Spatial frequency response; SFR In a video endoscope, the function relationship between the ratio of the modulation of the output signal which is subjected to the calculated value of the reverse function of OECF to the modulation of the brightness of the object plane of the target, and the spatial frequency of the target. 3.7 Sine-based spatial frequency response; s-SFR The SFR when the target is a sine-wave modulation map. 3.8 Modulation transfer function The ratio of the maximum signal value minus the minimum signal value to the maximum signal value plus the minimum signal value. 3.9 Spectral neutral portion shall be consistent with the internal material. If it is necessary to coat the surface, the manufacturer shall provide the corresponding coating requirements and test methods. 4.1.2 Requirements for chemical composition 4.1.2.1 General The material used in the part in contact with the patient shall be clearly indicated in any form possible by the manufacturer. The metal material shall be marked by the designation and/or code as well as the chemical composition requirements of the material; and be verified by test. The Chinese and English abbreviations for non-metallic materials shall be clearly indicated in any possible form. 4.1.2.2 Requirements for dissolved precipitate The dissolved precipitates of the polymer material in contact with the patient are as follows: a) pH: As compared with the same batch of blank control solution, the difference in pH shall be not more than 2.0; b) Total content of soluble heavy metals: The total content of soluble heavy metals in the eluate does not exceed 5.0 μg/mL; c) Potassium permanganate reducing substance: The difference between consumption of the same batch of blank control solution of equivalent volume shall be not more than 2.0 mL. 4.1.3 Biocompatibility Materials in contact with patients shall be evaluated for biosafety according to the principles and requirements of GB/T 16886.1, to demonstrate good biocompatibility. The biological evaluation may consider the results of biological tests, wherein the selection of the test items is carried out according to the guidelines of GB/T 16886.1. For materials that have previously been proven to be applicable, if it can prove that the subsequent manufacturing process is not sufficient to create a biosafety hazard, it may not repeat the biological test. Note 1: If the material of the device under design has an arguable history of use in a specific application; or otherwise if the information about the well as the corresponding camera mode (if the video endoscope has multiple camera modes). The tolerance of the signal-to-noise ratio is -20%. The upper limit is not counted. 4.6 Spatial frequency response The manufacturer shall, in the accompanied data, give the nominal value of the angular frequency of the object space of the video endoscope corresponding to the SFR value of 50% and 30%, as well as the corresponding camera mode (if the video endoscope has multiple camera modes). The tolerance of the angular frequency of the object space corresponding to the SFR value of 50% and 30% is -20%. The upper limit is not counted. 4.7 Static image tolerance The manufacturer shall, in the accompanied data, give the nominal value of the static image tolerance of the video endoscope as well as the corresponding camera mode (if the video endoscope has multiple camera modes). The tolerance of the static image tolerance is -20%. The upper limit is not counted. 4.8 Mechanical properties The mechanical properties of rigid video endoscopes shall comply with the requirements of YY 0068.2. 4.9 Electrical safety It shall meet the requirements of GB 9706.19. 5 Test methods 5.1 Test of material requirements 5.1.1 Surface material Visual inspection. For products with a surface coating, follow the appropriate test methods as provided by the manufacturer. 5.1.2 Test of chemical composition 5.1.2.1 General Use the method which has an accuracy of better than or reaching to the tolerance or 1/3 of the limit value to carry out the test of the chemical When testing on the plane perpendicular to the axis, use the illuminometer to measure the illuminance at the center of the filed angle. Record it as E0. Calculate the arithmetic mean of the above four azimuth illuminances E1, E2, E3, E4, as well as the ratio of the arithmetic mean value to the measured value of the illuminance at the center of the field angle. The calculated ratio shall be divided by the relative effect of lambert edge light luminosity [see formula (2)], where w is the field angle corresponding to the position where 0.9wp intersects the field angle. Wp may be replaced by w. 5.4 Test method of brightness response characteristics Check the technical data as provided by the manufacturer. Use the method as specified in Appendix A to measure the brightness response characteristics of the video endoscope. 5.5 Test method of signal-to-noise ratio Check the technical data as provided by the manufacturer. Use the method as specified in Appendix B to measure the signal-to-noise ratio of the video endoscope. 5.6 Test method of spatial frequency response Check the contents of the accompanied data as provided by the manufacturer. Use the method as specified in Appendix C to measure the spatial frequency response of the video endoscope. 5.7 Test method of static image tolerance Check the contents of the accompanied data as provided by the manufacturer. Use the method as specified in Appendix D to measure the static image tolerance of the video endoscope. 5.8 Mechanical properties The mechanical properties of the rigid video endoscope shall be carried out according to the test methods as specified in YY 0068.2. 5.9 Electrical safety It shall be carried out according to the test method as specified in GB 9706.19. A.2 Steps A.2.1 Test conditions The temperature of the test environment is 23 °C ± 2 °C; the relative humidity is 50% ± 20%. The dark illuminance of the test environment is not more than 1 lx. The power supply voltage’s stability of the control light source shall be controlled within ±2%. The light source shall be fully preheated and stable. A.2.2 Test procedure A.2.2.1 White balance For the video endoscope which has the white balance function, use the test target of A.1.1 to carry out white balance under the test conditions. A.2.2.2 Viewing of video endoscope Adjust the camera distance to the desired position. Record it as the measuring working distance d0. A.2.2.3 Focusing If the video endoscope has an autofocus function, when shooting the tested target, it may slightly blur the focusing, to reduce the noise as generated by the texture of the block itself. The emphasis here is “slightly”. The boundaries between blocks must be kept distinct. A.2.2.4 Setting the brightness of background B on the test target Adjust the brightness of the background B on the test target, to achieve the brightness value L0 as specified by the manufacturer. During the whole course of brightness change of the small gray-scale block A, the brightness of the background B shall keep the overall gain of the tested video endoscope unchanged. A.2.2.5 Changing the brightness of the small gray-scale block A on the test target and collecting the analysis image Gradually change the brightness of the small gray-scale block A on the test target. Within the tolerance range, select not less than 10 different brightness levels which are basic in uniform distribution. Corresponding to each brightness level, measure the brightness value and record it as Li. Use the video endoscope to photograph the test target. Use the image collector to collect n images and store it, wherein n is not less than 8. For the collected image, select (M × N) pixels in the small gray-scale block-A area (recommended 32 × 32). Respectively read the output signals of the output signal matrix (M x N x 3) corresponding to each channel of red, green, blue of each image. For a certain brightness level Li, the average values , , of the signal of each channel of red, green, blue are calculated from the arithmetic mean of the R, G, B values of (M × N × n) pixels, respectively. A.2.2.6 Calculating the display brightness value and fitness For the obtained Li, corresponding to the , , data sets, use the elec- opto conversion function (reverse function of OECF) of the output brightness as given by the manufacturer to calculate the displayed brightness Lyi (yi represents , , , respectively). If the manufacturer gives a data list, the elec-opto conversion function of the output brightness may be obtained by the use of piecewise linear fitting. The result of Lyi corresponding to the brightness Li of each gray-scale of the actually measured target shall be represented in the form of Table and/or graph. Calculate the linear fit R2 of L - Ly (effective digits to 2 digits after the decimal point). The calculation formula is as shown in formula (A.1): Where: m - The number of gray-scales; - The average value of Li; - The average value of Lyi. The red, green, blue channels shall be calculated respectively. A.3 Expression of results Appendix B (Normative) Test method of signal-to-noise ratio B.1 Equipment B.1.1 Test target Same as A.1.1. B.1.2 Light source Same as A.1.2. B.1.3 Image collector Same as A.1.4. B.2 Steps B.2.1 Test conditions Same as A.2.1. B.2.2 Test procedure B.2.2.1 White balance Same as A.2.2.1. B.2.2.2 Viewing of video endoscope Same as A.2.2.2. B.2.2.3 Focusing Same as A.2.2.3. B.2.2.4 Setting the brightness of background B on the test target Same as A.2.2.4. B.2.2.5 Changing the brightness of the small gray-scale block A on the test target and collecting the analysis image Gradually change the brightness of the small gray-scale block A on the test target. Within the tolerance range, select not less than 10 different brightness levels which are basic in uniform distribution. Corresponding to each brightness level, use the video endoscope to photograph the test target. Use the image collector to collect n images and store it, wherein n is not less than 8. For the collected image, select (M × N) pixels in the small gray-scale block-A area (recommended 32 × 32). Respectively read the output signals of the output signal matrix (M x N x 3) corresponding to each channel of red, green, blue of each image. B.2.2.6 Calculating the brightness signal component based on the output signal of each channel of red, green, blue For a certain brightness level, the average brightness signal component is calculated from the arithmetic mean of the Y values of (M × N × n) pixels, where Y is obtained by weighting the output signals of each channel of red, green, blue. The weight of each channel is valued according to the coding method as given by the manufacturer. Note: The general standard code may be found in ISO 22028-1:2004. Example: If the coding method uses the method as specified in ITU-R BT.709, the calculation of Y value may be weighted as follows: Where: R, G, B - The output signal values of each channel of red, green, blue. B.2.2.7 Calculating noise (represented by standard deviation) According to the Y value obtained in B.2.2.6, calculate the output signal values of the color difference channels (R - Y) and (B - Y). The noise may be calculated based on the standard deviation of brightness component σ (Y), the standard deviation of color difference channel σ (R - Y), and σ (B - Y) according to the formula (B.2). Where: σ (Y) - The standard deviation of the brightness signal component Y; σ (R - Y) - The standard deviation of the brightness channel lack of red; σ (B - Y) - The standard deviation of the brightness channel lack of blue. the spectral distribution curve’s shape of the light source is similar to the spectral distribution curve’s shape as specified by the manufacturer. The allowable difference of color temperature tolerance is ±10%. If the manufacturer uses the light source which is provided by itself as the test light source, the model and characteristics of the light source shall be declared in the test report. C.1.3 Image collector Same as A.1.4. C.2 Steps C.2.1 Conditions of measurement The illumination condition of the measured target is that the brightness of the measured target shall cause the video endoscope to produce an acceptable output signal level, but not over-exposed. C.2.2 Test procedure C.2.2.1 White balance Same as A.2.2.1. C.2.2.2 Viewing of video endoscope Adjust the photographing distance to the desired position. Record it as the measurement working distance d0. Center the test target. C.2.2.3 Focusing of video endoscope If the video endoscope has an autofocus function, it shall use the video endoscope's autofocus system to focus on the measurement working distance d0. In case of manual focus, select the sharpest focus setting when the spatial frequency is about 1/4 of video endoscope’s Nyquist frequency. C.2.2.4 Setting of video endoscope The image compression function of the video endoscope may significantly affect the measurement of resolution. For some video endoscopes, it may select whether to enable the image compression function by pressing a button. All settings of the video endoscope which may affect the measurement results, including the shooting mode, test distance, etc., shall be reported together with the measurement results. C.2.2.5 Collecting images D.2.2.5 Changing the brightness of the small gray-scale block A on the test target and collecting the analysis image The range of change of brightness level of the small gray-scale block A on the test target shall be more than the tolerance range. Meanwhile at least 5 brightness levels are lower than the cut-off critical brightness value of the dark area of the video endoscope to be tested. Near the cut-off critical brightness of the dark area and the saturated critical brightness of bright area, the ratio of adjacent brightness levels shall be not more than 1.1 times. Gradually change the brightness of the small gray-scale block A on the test target. For each selected brightness level, measure and record the brightness value Li. Use the video endoscope to photograph the test target of the corresponding brightness. Use the image collector to collect and save n images, wherein n is not less than 8. For the collected image, select (M × N) pixels in the small gray-scale block-A area (recommended 32 × 32). Respectively read the output signals of the corresponding output signal (M × N × 3) matrix corresponding to each channel of red, green, blue in each image. D.2.2.6 Calculating the brightness signal component based on the output signal values of each channel of red, green, blue Same as B.2.2.6. D.2.2.7 Drawing the curve of brightness and the output value of corresponding brightness signal According to Li and as obtained in steps D.2.2.5 and D.2.2.6A, draw the curve of the brightness and the corresponding brightness signal component. D.2.2.8 Reading the brightness saturation threshold Lsat in highlight region Read the critical brightness value Lsat when the brightness signal component in the highlight region on the curve is close to the saturation value. Note: Any channel that reaches saturation is saturated. D.2.2.9 Calculating the brightness cutoff threshold Lmin in dark area Read the critical brightness Lmin when the brightness signal component in the dark area starts cutting off from the curve. Determination of the cutoff state: Use the average value of the brightness signal component corresponding to the 5 sets of lower brightness levels as ......
 
Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.

Similar standards: YY/T 1603   YY 1296   YY 1082   YY/T 1889