GB/T 11539-2008: PDF in English (GBT 11539-2008) GB/T 11539-2008
NATIONAL STANDARD OF THE
PEOPLE’S REPUBLIC OF CHINA
ICS 71.100.60
Y 41
GB/T 11539-2008 / ISO 7359:1985
Replacing GB/T 11539-1989, GB/T 14455.9-1993
Fragrance/Flavor substances -
Analysis by gas chromatography on packed columns -
General method
(ISO 7359:1985, Essential oil - Analysis by gas chromatography on
packed columns - General method, IDT)
ISSUED ON: JULY 15, 2008
IMPLEMENTED ON: DECEMBER 01, 2008
Issued by: General Administration of Quality Supervision, Inspection and
Quarantine of the PRC;
Standardization Administration of the PRC.
Table of Contents
Foreword ... 3
Introduction ... 4
1 Scope ... 5
2 Normative references ... 5
3 Principle ... 5
4 Reagents and products ... 6
5 Apparatus ... 7
6 Preparation of test sample ... 8
7 Operating conditions ... 8
8 Column performance ... 8
9 Determination of retention indexes ... 13
10 Methods of determination ... 15
11 Expression of results ... 17
12 Test report... 19
Foreword
This Standard is identical to ISO 7359:1985 "Essential oil - Analysis by gas
chromatography on packed columns - General method". Compared with ISO
7359:1985, this Standard mainly changes the "essential oil" in the standard
name and the standard to "fragrance/flavor substances".
This Standard is the combination and revision of GB/T 11539-1989 "Isolate and
synthetic aromatics - Analysis by gas chromatography on packed columns -
General method" and GB/T 14455.9-1993 "Essential oils - Analysis by gas
chromatography on packed columns - General method".
This Standard was proposed by China National Light Industry Council.
This Standard shall be under the jurisdiction of National Standardization
Technical Committee on Flavor Fragrance & Cosmetic.
This Standard was drafted by Shanghai Research Institute of Fragrance &
Flavor Industry.
Main drafters of this Standard: Jin Qizhang, Xu Yi, Cao Yi.
The previous editions of the standard replaced by this Standard were released
as follows:
- GB/T 11539-1989;
- GB/T 14455.9-1993.
Fragrance/Flavor substances –
Analysis by gas chromatography on packed columns -
General method
1 Scope
This Standard specifies a general method for the analysis of fragrance/flavor
substances by gas chromatography on packed columns for the purpose of
determining the content of a specific constituent and/or searching for a
characteristic profile.
2 Normative references
The following documents contain provisions which, through reference in this
Standard, constitute provisions of this Standard. For the dated references, their
subsequent amendments (excluding corrections) or revisions do not apply to
this Standard. However, the parties who enter into agreement based on this
Standard are encouraged to investigate whether the latest editions of these
documents are applicable. For undated reference documents, the latest
editions apply to this Standard.
GB/T 11538 Essential oil - Analysis by gas chromatography on capillary
columns - General method (GB/T 11538-2006, ISO 7609:1985, IDT)
GB/T 14454.1 Fragrance/Flavor substances - Preparation of test samples
(GB/T 14454.1-2008, ISO 356:1996, MOD)
3 Principle
Analysis by gas chromatography under specified conditions of a small quantity
of fragrance/flavor substance on a column packed with appropriate material.
If required, identification of the different constituents from their retention indexes.
Quantitative determination of specific constituents by measurement of peak
areas.
9 Determination of retention indexes
If it is necessary to determine the retention indexes, prepare a mixture of the
test portion with normal alkanes including n-pentane. Choose the normal
alkanes in accordance with the range of retention indexes expected. After
stabilization of the temperature of the column, inject a suitable quantity of the
mixture and proceed with the analysis under the conditions specified in 10.1.
Chromatogram “B” is thus obtained.
9.1 Measurement of retention indexes
Compare chromatograms “A” (see 10.1) and “B” (see clause 9) and note on
chromatogram “B” the peaks corresponding to the normal alkanes.
On chromatogram “B”, make the following measurements.
9.1.1 Isothermal conditions
9.1.1.1 If a detector with thermal conductivity is used, calculate the difference
between the retention distances of the top of the peak under consideration and
the top of the air peak. Let be this difference, in millimetres.
Calculate the difference between the retention distances of the top of the air
peak and the top of the normal alkane peak appearing immediately before the
peak considered. Let be this difference, in millimetres.
Calculate the difference between the retention distances of the top of the air
peak and the top of the normal alkane peak appearing immediately after the
peak considered. Let be this difference, in millimetres.
9.1.1.2 If a flame ionization detector is used, calculate the difference between
the retention distances of the top of the peak considered and the top of the
methane peak. Let be this difference, in millimetres.
Calculate the difference between the retention distances of the top of the
methane peak and the top of the normal alkane peak appearing immediately
before the peak considered. Let be this difference, in millimetres.
Calculate the difference between the retention distances of the top of the
methane peak and the top of the normal alkane peak appearing immediately
after the peak considered. Let be this difference, in millimetres.
9.1.2 Procedure using linear temperature programming from injection
Measure the distance on the base line between the top of the peak the retention
...... Source: Above contents are excerpted from the PDF -- translated/reviewed by: www.chinesestandard.net / Wayne Zheng et al.
|